Huiming Dong1,2, Rizwan Ahmad2, and Arunark Kolipaka1,2
1Department of Radiology, The Ohio State University Wexner Medical Center, Columbus, OH, United States, 2Department of Biomedical Engineering, The Ohio State University, Columbus, OH, United States
Synopsis
Lung stiffness is a potential biomarker for
multiple lung diseases. MR elastography (MRE) allows non-invasive measurement
of lung stiffness. However, it is challenging to estimate stiffness using
direct Helmholtz inversion due to low signal-to-noise ratio (SNR) from lung MRE.
In this work, a compressed-sensing-based Helmholtz inversion is proposed where
noise is reduced via Laplacian of Gaussian (LoG) and Morozov’s discrepancy
principle, while the sparsity of stiffness map is explored in a wavelet domain.
Results demonstrated that the proposed inversion yielded robust stiffness
estimation and successfully detected higher stiffness at total lung capacity
(TLC) compared to residual volume (RV).
Introduction
Direct inversion (DI) of the Helmholtz equation
is an established inversion to estimate tissue stiffness using MR elastography
(MRE) data [1, 2]. Under the presence of noise and low wave penetration,
correctly estimating stiffness can be challenging because the Laplacian
operation within the Helmholtz equation can amplify the noise while obtaining
spatial derivatives from the measured displacement, leading to biased stiffness
estimates.
This becomes more problematic in the case of
estimating shear stiffness of the lung in which the low tissue density and
short T2* result in considerably low signal-to-noise ratio (SNR) during lung
MRE. It is appreciated that the normal function of the lung is closely
associated with its gross topographically heterogeneous mechanical properties
which can only be invasively quantified using current diagnostic tools [3],
making non-invasive in vivo lung MRE a valuable yet challenging technique [4-6].
To reduce the effect of noise, a large-size
kernel (e.g., 13x13 or larger) is usually used to estimate the Laplacian of the
displacement field. This strategy offers improved derivative estimation by
using the information from a wide coverage of pixels, making the Laplacian
operation less sensitive to local fluctuations resulted by noise. Despite its
practical significance, a large kernel will potentially reduce the achievable
spatial resolution of the stiffness map. Additionally, large kernel sizes
cannot resolve the problem of robustly/stably estimating stiffness because the
low SNR in the lung MRE data is the main concern.
In this work, a compressed-sensing-based Helmholtz
inversion using unconstrained optimization has been proposed where noise is
reduced from the measured data using Laplacian of Gaussian (LoG) and Morozov’s
discrepancy principle [7], while the sparsity of stiffness map in wavelet
domain is explored.Method
Theory
The optimization problem is described by the
following cost function:
$$$J(μ,u)=\underset{μ, u}{argmin}\space||c·LoG(u)·μ-u||\small{{2 \atop 2 }}+λ_1\||u-u_0\||\small{{2 \atop 2 }}+λ_2\||ψ(μ)||_1$$$
where c = -1/(ρω2), ω is the applied mechanical frequency, LoG is the Laplacian of Gaussian operator, µ is the stiffness map of interest, u is the pursued tissue displacement, u0 is the measured tissue displacement prior to
denoising, ψ is a non-decimated wavelet transform and λ1, λ2 are regularization coefficients. The first
term guarantees data fidelity (i.e., the Helmholtz equation). The second term allows the
pursued displacement u to deviate from the measured displacement u0,
reducing undesirable noise. The third term ensures the sparsity of stiffness
map in the wavelet domain. In this work, the Gaussian operator was chosen to
have a full width at half maximum (FWHM) of 13 pixels for LoG.
Figure 1 demonstrates
the details of the proposed optimization technique. The regularization
coefficient λ1 is automatically controlled via Morozov’s discrepancy
principle during each iteration to allow different amount of denoising by
comparing the measured noise (which is defined as noise ratio x u0) and the
difference between u and u0,
whereas λ2 is specially tuned for lung MRE. Balanced fast iterative/shrinkage
thresholding algorithm (bFISTA) and the least-square method (LSQR) were used in
optimizing the cost function [8,9].
Lung MRE
Lung MRE was performed on 5 healthy subjects
using a modified breath-hold SE-EPI MRE sequence reported in a previous
feasibility study [10]. Imaging parameters included: TE/TR=11.6/400 ms;
FOV=400x400x10 mm3; reconstruction matrix=256x256; mechanical frequency=50
Hz; MEG frequency=250 Hz; three-directional motion encoding; 4 MRE phase
offsets. Lung density of each volunteer was also measured at RV and TLC as
described in [10].
Image
Analysis
Lung shear stiffness at RV and total lung
capacity TLC was estimated using the proposed inversion algorithm and the
standard DI with 17x17 Romano filter in MRElab (Mayo Clinic, Rochester, MN).
For both methods, a 4th-order Butterworth bandpass directional
filter was applied in eight directions with cutoff values of 4–40 waves/FOV for
RV and 2–40 waves/FOV for TLC to remove the longitudinal and reflected waves as
described previously [10].Results and Discusssion
Figure 2 compared
the measured displacement and the displacement generated from the proposed
inversion technique for each direction. The reconstructed displacement contains
less noise compared to the measured displacement.
Figure 3 demonstrates
the lung stiffness map in the same volunteer after lung density correction at
RV and TLC for both optimization method and the standard DI method without any
post-processing (i.e., no median filter, boundary erosion, etc.).
Table I summarizes
the lung stiffness of each individuals. It is important to notice that the
standard DI yielded unrealistic stiffness values with large standard deviation
due to the noise presented in the data despite the applied 17x17 Romano filter.Table II summarizes the lung densities of each individuals.
The proposed optimization technique demonstrated
significantly higher lung stiffness at TLC than RV (paired t-test, P=0.0104). The mean optimization-derived
stiffness for all volunteers is 1.04±0.12 and 1.55±0.29 kPa at RV and TLC, respectively.
No significant difference was observed between DI-derived RV and TLC stiffness (paired
t-test, P=0.9668). The mean
DI-derived stiffness for all volunteers is 3.56±5.83 and 3.71±2.82 kPa at RV and TLC,
respectively.Conclusion
In this study, the proposed optimization
technique demonstrated more robust/stable stiffness estimates in the lung compared
to the standard DI method, successfully demonstrating higher shear stiffness at
TLC than that at RV.Acknowledgements
The
authors acknowledge grant sponsor: NIH–NHLBI (grant number:
NIH-R01HL124096).References
1. Papazoglou S, Hamhaber U, Braun J, and Sack
I. Algebraic Helmholtz inversion in planar magnetic resonance elastography.
Physics in medicine and biology 2008;53(12):3147-3158.
2. Oliphant TE, Manduca A, Ehman RL, and
Greenleaf JF. Complex‐valued stiffness reconstruction for magnetic resonance
elastography by algebraic inversion of the differential equation. Magnetic
resonance in Medicine 2001;45(2): 299-310.
3. Faffe DS, Zin WA. Lung Parenchymal Mechanics
in Health and Disease. Physiol Rev. 2009;89(3):759-775.
doi:10.1152/physrev.00019.2007.
4. Mariappan YK, Glaser KJ, Hubmayr RD, Manduca
A, Ehman RL, McGee KP. MR elastography of human lung parenchyma: Technical
development, theoretical modeling and in vivo validation. J Magn Reson Imaging.
2011;33(6):1351-1361. doi:10.1002/jmri.22550.
5. McGee KP, Mariappan YK, Hubmayr RD, et al.
Magnetic resonance assessment of parenchymal elasticity in normal and
edematous, ventilator-injured lung. J Appl Physiol. 2012;113(4):666-676.
doi:10.1152/japplphysiol.01628.2011.
6. Marinelli JP, Levin DL, Vassallo R, et al.
Quantitative assessment of lung stiffness in patients with interstitial lung
disease using MR elastography. J Magn Reson Imaging. 2017;46(2):365-374.
doi:10.1002/jmri.25579.
7. Scherzer O. The use of Morozov’s discrepancy
principle for Tikhonov regularization for solving nonlinear ill-posed problems.
Computing. 1993;51:45–60.
8. Ting ST, Ahmad R, Jin N, Craft J, Serafim da
Silveira J, Xue H, Simonetti OP. Fast implementation for compressive recovery
of highly accelerated cardiac cine MRI using the balanced sparse model.
Magnetic Resonance in Medicine 2017;77(4):1505-1515.
9. Paige CC and Saunders MA, LSQR: An Algorithm
for Sparse Linear Equations And Sparse Least Squares, ACM Trans. Math. Soft. 1982;8:43-71.
10. Fakhouri F, Dong H, Kolipaka A. Magnetic
resonance elastography of the lungs: A repeatability and reproducibility study.
NMR Biomed. 2019. doi:10.1002/nbm.4102.