Siri Flogstad Svensson1, Jorunn Fraser-Green2, Omar Darwish3, José Rodriguez de Arcos3, Tryggve Holck Storås1, Sverre Holm4, Ralph Sinkus3,5, and Kyrre Eeg Emblem1
1Department for Diagnostic Physics, Oslo University hospital, Oslo, Norway, 2The Intervention Center, Oslo University hospital, Oslo, Norway, 3Division of Imaging Sciences & Biomedical Engineering, King's College, London, United Kingdom, 4Department of Informatics, University of Oslo, Oslo, Norway, 5INSERM U1148, LVTS, University Paris Diderot, Paris, France
Synopsis
We assessed the test-retest reliability of MR Elastography, and compared
the estimates of the shear storage modulus G’ (tissue elasticity) from two different reconstruction methods; a localized divergence-free finite-element approach and
a local curl approach. The median coefficient of variation
for two scans of the brain was 2.8 %, whereas the intraclass correlation
coefficient between scan 1 and scan 2 was 0.80. G’ values estimated using the local curl method
were 19% lower than with the divergence-free method.
MRE of the human brain appears robust, while absolute values are dependent on
the reconstruction method and should be used with care.
Introduction
Many
diseases are known to alter the biomechanical properties in tissue. Magnetic
resonance elastography (MRE) may be used to quantify these properties noninvasively by its shear module. However, for one to
rely on potential biomarkers from MR elastography, the technique needs to be
robust. That is, its test-retest reliability should be high. Moreover, the
output from MR elastography is determined by the reconstruction method applied.
In our study, we compare the estimates from two reconstruction methods; (I) a localized divergence free finite element
reconstruction approach1 and (II) a local curl
reconstruction approach2.Methods
The study was
performed in 15 healthy volunteers, and approved by the Regional Ethics Committee. The exams were
performed on a 3T Philips Ingenia MRI scanner. A T1-weighted series acquisition
was used as an anatomical reference scan, and diffusion tensor imaging was also
performed. Two MRE scans were acquired using
the Gravitational Transducer3, at 50 Hz. Image acquisition was
performed using a multi-slice gradient-echo acquisition4, with a
Hadamard encoding scheme5. Other scan parameters were: 15 slices, 3.1 mm isotropic resolution, TE=12
ms, TR =384 ms, and a 72 × 70 matrix size. Seven acquisitions over the wave
cycle were performed, and
each MRE acquisition lasted 5.5 minutes.
The two MRE acquisitions were
performed during the same scanning session, with approximately 15 minutes of
other MR sequences between the two MRE scans.
Two reconstruction
methods were used: A localized
divergence free finite element reconstruction1 and a
local curl reconstruction2. Further image analysis was done in Matlab (version R2018a, MathWorks, Natick, MA, USA). Regions-of-interest in the volunteers were
defined from an atlas of
brain regions6 and SPM12 (version 7487, Wellcome Trust Centre for Neuroimaging,
London) was used to warp these ROIs to
the native space of the MRE images.
To avoid
artefacts from MRE reconstruction at the edge of the brain, the outermost
voxels were removed from the cortical gray matter regions. This was done using
an eroded mask from each MRE magnitude image. Apparent diffusion coefficient
(ADC) maps from the diffusion acquisition was used to reduce partial volume effects.
ADC maps were coregistered and resliced to MRE space, and a mask with a cutoff
of ADC< 1.2 ×10-3 mm2/s, was
used to exclude voxels with a high content of cerebrospinal fluid7.
The analysis was performed in the full brain, as well as in white and
gray matter ROIs. The ROIs are illustrated in figure 1.
Intraclass
Correlation Coefficients (ICC) with 95% confidence intervals were calculated
using SPSS 23 (SPSS Inc, Chicago, IL) based on an absolute-agreement, 2-way
mixed-effects model. Moreover, coefficient of variations (CoV) were calculated
in Matlab by acquiring the ratio between the standard deviation and the mean of
the two measurements.Results
The median CoV for the
measured shear storage modulus G’ (tissue elasticity) for the entire brain,
using the divergence-free reconstruction method, was 2.8 % (range 0.2-13 %).
For the white- and gray-matter ROIs, the median CoV was 3.1 % (range 0.0-21.2
%). The ICC between scan one and two for G’ in the entire brain was 0.80,
with an ICC median of 0.77 for the ROIs. Table 1 shows the CoV and the ICC for
all ROIs, and for both reconstruction methods, which showed similar test-retest
reliability. Figure 2 shows the G’ from
the two scans.
Figure 3 shows the average
G’ values for each ROI, and using both reconstruction methods. The local curl reconstruction showed on average 19% lower
(p=0.0008, Wilcoxon signed-rank test) G’ values compared to that of the finite
element reconstruction.Discussion
An earlier study8 using a
different MR elastography hardware and reconstruction method studied the reliability
of their technique in three scans of brains of ten healthy volunteers. They
report a median CoV for global brain stiffness of 0.67% (maximum 1.11%), and of 1.98% (max 4.47%)
in the lobes of the brain, deep GM/WM
and cerebellum regions.
Murphy et
al8 evaluated
the test-retest reliability values for the shear stiffness instead of the shear
storage modulus, as in our study, due to the shear stiffness being more
resistant to noise. Because the shear modulus offers information of both
elasticity (G’) and viscosity (G’’), which are clinically relevant for several
pathologies, assessing the test-retest reliability of this variable is important.
The ICC
values are higher for the cortical gray matter regions than for the deep gray
and white matter regions. The ROIs analysed here differ in size, with the
cortical gray matter ROIs containing more voxels than the deep gray matter
ROIs. The smaller ROIs will be more sensitive to a correct warping of the brain
atlas regions to each subject’s image space, which could help explain the
differences between regions. Putamen and thalamus are two regions with lower ICC
than the rest of the brain, suggesting the measurements in these regions may be
less robust.
We observed
lower G’ values using the local curl method, compared to the divergence-free
FEM method, as reported before1. This warrants caution when using absolute
values from MRE data.Conclusion
MRE of the
human brain appears robust, while absolute values are dependent on the reconstruction
method and should still be used with care.Acknowledgements
No acknowledgement found.References
1) Fovargue
D, Kozerke S, Sinkus R, et al. Robust
MR elastography stiffness quantification using a localized divergence free
finite element reconstruction. Medical Image
Analysis. 2018;44:126–14
2) Sinkus R, Tanter M, Xydeas T, et al. Viscoelastic shear properties of in vivo breast
lesions measured by MR elastography. Magnetic Resonance Imaging. 2005;23:
159–165
2)
Runge J, Hoelzl SH,
Sudakova J, et al. A novel magnetic
resonance elastography transducer concept based on a rotational eccentric mass:
preliminary experiences with the gravitational
transducer. Phys. Med. Biol. 2019; 64(4)
3) Guenthner C, Sethi S, Troelstra
M, et al. Ristretto MRE: A
generalized multi-shot GRE-MRE sequence. NMR in Biomedicine. 2019; e4049.
4) Guenthner C, Runge J, Sinkus R
et al. Analysis and
improvement of motion encoding in magnetic resonance elastography. NMR in Biomedicine. 2018;31(5):e3908
5) Tzourio-Mayer N, Landeau B,
Papathanassiou, et al. Automated anatomical labeling of activations in SPM using a macroscopic
anatomical parcellation of the MNI MRI single-subject brain. Neuroimage. 2002;15(1):273-89.
6) Bykowski JL, Latoru LL, Warach S. More Accurate
Identification of Reversible Ischemic Injury in Human Stroke by Cerebrospinal
Fluid Suppressed Diffusion-Weighted Imaging. Stroke. 2004;35:1100-1106.
7) Murphy MC, Huston J, Jack Jr C, et al. Measuring the Characteristic Topography of Brain
Stiffness with Magnetic Resonance Elastography. PLoS One. 2013; 8(12): e81668.