Dapeng Liu1,2, Wenbo Li1,2, Feng Xu1,2, Dan Zhu3, Taehoon Shin4,5, and Qin Qin1,2
1Department of Radiology, Johns Hopkins University School of Medicine, Baltimore, MD, United States, 2F.M. Kirby Research Center for Functional Brain Imaging, Kennedy Krieger Institute, Baltimore, MD, United States, 3Department of Biomedical Engineering, Johns Hopkins University School of Medicine, Baltimore, MD, United States, 4Division of Mechanical and Biomedical Engineering, Ewha Womans University, Seoul, Korea, Republic of, 5Department of Medicine, Case Western Reserve University, Cleveland, OH, United States
Synopsis
In Fourier-transform based velocity-selective inversion (FT-VSI) arterial
spin labeling (ASL), either velocity-insensitive or velocity-compensated
waveforms could be applied for control modules. Under poor B0/B1 conditions with
inefficient refocusing, the scheme with velocity-insensitive control is
susceptible to strong false signal but maintained high labeling efficiency,
while the scheme with velocity-compensated control has poor velocity-selective labeling
profiles. This study overcome these problems by proposing a dynamic phase-cycling
scheme based on velocity-insensitive waveform and improve the robustness to B0/B1 field inhomogeneities for VSI ASL. Simulation, phantom and brain
ASL scans were conducted for demonstration.
Introduction
Fourier-transform
based velocity-selective inversion (FT-VSI) prepared arterial spin labeling
(ASL) has demonstrated improved SNR for brain perfusion mapping1.
Although paired and phase-cycled refocusing pulses improved its robustness to
B0 and B1 inhomogeneities, severe B0/B1 offsets are known to cause imperfect
refocusing leading to stripe artifacts in velocity-selective MR angiography2,3.
In ASL, the periodic banding of high spatial-frequency is averaged out due to
low spatial resolution. Instead, net DC-bias subtraction errors for perfusion
quantification was newly identified in the current study, especially when applying the control modules with gradients turned off as often adopted in
VSASL. Here a new dynamic phase-cycling scheme is proposed to mitigate the
DC-bias for VSI-ASL and is evaluated in simulations, phantom and human brain
scans at 3T.Methods
Numerical
simulation was conducted using Matlab. A 64ms FT-VSI pulse train was composed
of nine excitation pulses and eight 8ms velocity-encoding steps, each
containing paired and phase-cycled composite refocusing pulses4 and
four triangular gradient lobes (30mT/m, 0.6ms duration, 0.3ms ramp, 2cm/s
cut-off velocity). Both Mz-velocity and Mz-position
responses were evaluated for three schemes: 1) using velocity-insensitive
control with gradients turned off; 2) using velocity-compensated control with
uni-polar gradient lobes3-5) using velocity-insensitive
control while phase-cycling the refocusing pulses of every label/control pairs
by 90° through every four dynamics (dynamic 1: +0°; dynamic 2: +90°; dynamic 3:
+180° and dynamic 4: +270°) in addition to the MLEV-16 phase-cycling applied within
the VS pulse trains. The label/control subtraction results of phase-cycled dynamics are
averaged to yield the final image.
Experiments on
an oil phantom and five healthy volunteers (two females, 32 ± 5 yo) were
performed on a 3T Siemens Prisma scanner. VSI-ASL sequences with schemes 2 and
3 were implemented as described above. B1+ scale were manually adjusted to 0.8,
1.0 and 1.2. A six-segment 3D gradient- and spin-echo (GRASE) readout was
applied with: imaging volume=220x220x96mm3, resolution=3.4x3.4x4.0mm3,
reconstructed to 1.7x1.7x4.0mm3. Post-labeling delays (PLD) was 0.1s
for phantom scans and 1.5s for brains. Four dynamics were acquired in around
2min for phantom and 3.3min for brain scans. For phantom scans, the
subtraction results between label/control pairs were divided by M0. For brain scans
cerebral blood flow (CBF) of whole-brain gray matter (GM) were calculated.
Student’s paired t-test was used to evaluate signal difference.Results
Simulated
Mz-velocity and Mz-position responses for both velocity-insensitive and velocity-compensated
controls (schemes 1 and 2) with varied B1+ were shown in Fig.1. The subtracted
signal of scheme 1 showed well-maintained velocity-selective profiles, except
labeling efficiency scaled by B1. The velocity-selective profile of scheme 2 is
more susceptible to B1+ imperfection (Fig.1a). However, Mz-position responses
showed less stripes and DC-bias (average of stripes) in scheme 2 as the
velocity-compensated control largely cancels this effect (Fig.1b), as observed
experimentally before2,4,5.
In a specific
B0/B1 condition (B0=150Hz and B1+=0.7) as example, the subtraction of label and
control of scheme 2 yielded much reduced banding and DC bias (Fig.2a) than
scheme 1. Scheme 3 with averaging through dynamic phase-cycling further
suppressed artifacts (Fig.2b). When DC bias is plotted as functions of B0/B1 (Fig.2c,d),
similar performance are shown. The false perfusion signal was considerably
canceled with scheme 3 within a wide range of B0/B1.
The normalized
subtraction results of phantom scans at three B1+ scales were shown with all
four dynamics and their averages in Fig.3. For scheme 2, a strong positive
false signal were present with B1+ of 0.8. Some negative false signal can also
be seen at the edge of the phantom with B1+ of 1.2, probably due to a combined
B1 and B0 imperfections. For scheme 3, although each dynamic generated strong
false signal, they were mostly canceled after averaging.
Similar effects
were observed for brain CBF mapping (Fig.4). Note that for scheme 2, B1+ of 0.8
showed a stronger signal than 1.2, although with the same theoretical labeling
efficiency. This asymmetry was likely caused by different false signal with B1+ of 0.8 and 1.2 in scheme 2 (Fig.3a). In contrast, scheme 3 after averaging the
dynamic phase-cycles showed consistent CBF results.
The whole-brain GM CBF was displayed for all
five subjects (fig.5). The asymmetry of B1+ of 0.8 and 1.2 was significant in
scheme 2 (p=0.004), but not in scheme 3 (p=0.431). Interestingly, with scheme
2, B1+ of 1.2 showed significantly lower perfusion signal (p=0.012) than the
results with scheme 3, which could not be explained by false signal alone. This
was attributed to the loss of labeling efficiency in scheme 2 due to poor B1+ (Fig.1a).
In principle, B1+ of 0.8 in scheme 2 should have similar effect, but the lower
perfusion signal was likely counteracted by the strong positive false signal
observed in phantom experiments (Fig.3a). Conclusion
In this VSI ASL study, we proposed dynamic
phase-cycling scheme for the FT-VS label and control
modules while using
velocity-insensitive controls in order to achieve high labeling efficiency in
the presence of a wide range of B0/B1 field inhomogeneities. Simulation,
phantom scans, and brain perfusion measurements all demonstrated considerable
improvement with our proposed method.Acknowledgements
No acknowledgement found.References
1.
Qin Q, van Zijl PCM. Velocity-selective-inversion prepared
arterial spin labeling. Magn. Reson. Med. 2016; 76:1136–1148
2. Shin T, Qin Q, Park JY, Crawford RS, Rajagopalan S. Identification and reduction of image
artifacts in non-contrast-enhanced velocity-selective peripheral angiography at
3T. Magn Reson Med. 2016; 76(2):466-77
3. Shin T, Qin Q.
Characterization and suppression of stripe artifact in velocity-selective magnetization-prepared unenhanced MR angiography.
Magn Reson Med. 2018; 80(5):1997-2005
4.
Dapeng Liu, Wenbo Li, Doris D. Lin, Peter van Zijl, and Qin Qin.
Optimization of Velocity-Selective-Inversion Arterial Spin Labeling (VSI-ASL)
with 3D Whole-Brain Coverage. In
Proceedings of the 27th Annual Meeting of ISMRM, Montreal, Canada, 2019
5. Qin Q, Qu Y, Li W, Liu D, Shin T, Zhao Y, Lin DD, van Zijl
PCM, Wen Z. Cerebral blood volume mapping using Fourier-transform–based
velocity-selective saturation pulse trains. Magn. Reson. Med. 2019; 81:3544–3554