Qihao Zhang1, Liangdong Zhou2, John Morgan3, Thanh D Nguyen4, Pascal Spincemaille3, and Yi Wang2
1Cornell University, Ithaca, NY, NY, United States, 2Weill Cornell Medicine, New York, NY, United States, 3Radiology, Weill Cornell Medicine, New York, NY, United States, 4Weill Cornell Medicine College, New York, NY, United States
Synopsis
Tissue flow vector field is generated by solving the inverse
problem of a voxelized transport equation using multi-delay ASL data and an optimization
method. The accuracy of the flow estimation is validated on a vasculature
model. For calculating flow vector field in vivo, time resolved 3D (4D) tracer
concentration data is acquired using a multi-delay pseudo continuous ASL
sequence. The background suppression pulse is interleaved into labeling pulse
to generate short post label delay. The output flow magnitude map shows a more
anterior-poster-uniform CBF than the traditional Kety’s method.
Introduction
Perfusion imaging evaluates the blood flow through the tissue
capillary bed1,which is an important biomarker of tissue vitality. Current quantification of perfusion is based on Kety’s equation,which requires
a global arterial input function (AIF), making perfusion
quantification highly dependent on the selection of the arterial region, its
size, shape, and location. Quantitative transport mapping (QTM) eliminates the
AIF problem by computing a velocity and flow field based on the transport equation of
mass flux2,3. QTM is applied to time resolved contrast enhanced data to map cerebral blood flow (CBF). Here we use multi-delay ASL data and compare CBF
obtained using QTM and the traditional Kety method.Methods
Algorithm:
Given multi-delay ASL data $$$c(\xi,t_i)$$$ for voxels $$$\xi$$$ at times $$$t_i$$$, QTM
solves the following transport inverse problem:$$\boldsymbol{q}=argmin_{\boldsymbol{q}}\sum_{i=1}^{N}||c_{t}(\xi,t_{i})+\nabla\cdot \frac{c_{t}(\xi,t_{i})\boldsymbol{q}(\xi)}{V\cdot CBV(\xi)} ||_{2}^{2}+\lambda||\boldsymbol{q}||_{2}^{2} \qquad (1)$$
with $$$c_{t}(\xi,t_{i})=\frac{1}{\Delta t}[c(\xi,t_{i+1})-c(\xi,t_{i})]$$$ the temporal derivative of c. $$$\boldsymbol{q}(\xi)=[q^{x}(\xi),q^{y}(\xi),q^{z}(\xi)]$$$ is the flow defined on the three surfaces of
the voxel at positive x,y,z directions, $$$CBV(\xi)$$$ is the cerebral blood volume fraction within the voxel, V the voxel volume, and $$$\nabla\cdot \boldsymbol{f}(\boldsymbol{\xi})=\sum_{i=x,y,z}f^{i}(\xi_{i})-f^{i}(\xi_{i-1})$$$ the divergence operator. An L2 regularization
is used for denoising. Eq.1 is solved
using a conjugate gradient method. $$$\boldsymbol{q}$$$ is assumed to be a steady flow with $$$\nabla\cdot \boldsymbol{q}=0$$$, which renders the
optimization of the discretized form of Eq.1 more robust. Eq.1
contains 3 unknowns for each voxel, so at least 4 time points (3 temporal
derivatives) are needed to solve Eq.1.
Numerical validation:
For validating QTM, a vessel tree network in a double-cone shape containing
2560 vessel segments with a 1cm mid cross-section radius and 6 cm length
(Fig.1a) was generated using the branching rule7. The inlet velocity
was set to 2cm/s and plug flow was assumed for each vessel segment in order to
simplify the forward problem of generating 4D tracer imaging data of sufficient
spatiotemporal resolution using the mass flux equation8 (Fig.1b, c
and d). The concentration data was simulated using finite element modeling
(FEM) of the transport equation with a boundary condition of a concentration
profile at the inlet being $$$c(t)=te^{-\frac{t}{5}}$$$. The FEM data was then down-sampled to the ASL voxel size and temporal frame duration and QTM was applied. $$$\lambda=1$$$ was chosen empirically to minimize simulation
error and used for both simulation and ASL data.
In vivo imaging experiment.
A background
suppression pulse was interleaved into the labeling pulse to realize short pulse
label delay 3D acquisition4. Two methods were used to acquire
multi-delay ASL: the first was to fix labeling time (1500ms) and vary post
label delay (PLD) (25ms-2525ms with 500ms interval), and the second was to fix PLD
(25ms) and vary labeling time (1500ms-4000ms with 500ms interval). The corresponding ASL data is shown in Fig.2. The first method was chosen for better
signal in the posterior brain. As the PLDs were smaller than mean transit time
(MTT), tracer stayed mainly in artery and capillary space5.
Therefore, CBV was set to the blood volume of artery (CBVa),
which was 1% for gray matter and 0.75% for white matter6. A T1w image was used to generate gray/white matter segmentation and the CBV
value for each voxel. 7 healthy subjects were scanned on
a 3T GE scanner, using the PCASL sequence (voxel
size mm3, field of view 24cm, number of slices 36, 6
post label delays). The flow vector $$$\boldsymbol{q(\boldsymbol{\xi})}(mm^{3}/s)$$$ from QTM was converted to
CBF (in mL/100g/min) using: $$CBF=\frac{\sqrt{q^x(\xi)^2+q^y(\xi)^2+q^z(\xi)^2}}{\rho V}\cdot 100g\cdot 60s,$$ where $$$\rho$$$ is blood density. CBF
generated using Kety’s method9 was used as comparison in both simulation and in vivo data.Results
Figure
1 shows high accuracy of QTM flow mapping (Figure 1c), but substantial error in Kety’s flow (Figure 1d). Figure 3
shows an in vivo comparison of CBF obtained using QTM (Figure 3c) and Kety’s method (Figure 3d). Flow using QTM was more uniform in the
posterior gray matter, but lower near the circle of Willis,
compared to Kety’s flow (Figure 3, arrows). A weak correlation between QTM and Kety’s flow was observed (Figure 4). The average flow in gray matter was
63±15 mL/100g/min for QTM and 65±14 mL/100g/min for Kety’s method, while in
white matter it was 42±17 mL/100g/min for QTM and 46±14 mL/100g/min for Kety’s method. Figure 5
shows the QTM flow vector map overlayed onto the tracer concentration map. QTM captured the in-plane flow from gray matter into white matter, while the
flow vertical to the plane needs to be validated because the slice thickness is
large (4mm). Discussion and Conclusion
In simulation, QTM based flow was more accurate than Kety’s. In healthy subjects, QTM showed a more uniform flow in the posterior gray matter,
likely due the Kety method's reliance on the M0 map (Figure 3a) and there is
artifact in M0 near the edge of the brain. QTM estimated flow is low near the circle of Willis, possibly due the inability to capture flow in large arteries with the used frame rate. While QTM and Kety’s method produced
similar average flow for global gray and white matter, QTM flow vector
map captures blood flow from gray matter to white matter. Further validation using
flow phantoms and more realistic vasculature models are warranted to evaluate the QTM flow vector map.Acknowledgements
References
1. Kety SS. The theory and applications of the exchange of
inert gas at the lungs and tissues. Pharmacol Rev 1951;3(1):1-41.
2. Zhou L, Spincemaille P, Zhang Q, Nguyen T, Doyeux V,
Lorthois S, Wang Y. Vector Field Perfusion Imaging: A Validation Study by Using
Multiphysics Model. ISMRM, 2018; Paris, France. p 1870
3. Zhou L, Zhang Q, Spincemaille P, Nguyen T, Wang Y.
Quantitative Transport Mapping (QTM) of the Kidney using a Microvascular
Network Approximation. ISMRM, 2019 4/26/2019; Montreal. p 703.
4. Dai W, Robson P M, Shankaranarayanan A, et al. Reduced
resolution transit delay prescan for quantitative continuous arterial spin
labeling perfusion imaging[J]. Magnetic resonance in medicine, 2012, 67(5):
1252-1265.
5. Buxton RB. Introduction to Functional Magnetic Resonance
Imaging. Cambridge University Press. (2009) ISBN:1139481304.
6. Hua J, Liu P, Kim T, et al. MRI techniques to measure
arterial and venous cerebral blood volume[J]. Neuroimage, 2019, 187: 17-31.
7. Karch R, Neumann F, Neumann M, et al. A three-dimensional
model for arterial tree representation, generated by constrained constructive
optimization[J]. Computers in biology and medicine, 1999, 29(1): 19-38.
8. Lorthois S, Cassot F, Lauwers F. Simulation study of
brain blood flow regulation by intra-cortical arterioles in an anatomically
accurate large human vascular network: Part I: methodology and baseline flow.
Neuroimage 2011;54(2):1031-1042.
9. Buxton R B. Quantifying CBF with arterial spin
labeling[J]. Journal of Magnetic Resonance Imaging: An Official Journal of the
International Society for Magnetic Resonance in Medicine, 2005, 22(6): 723-726.