Tanja Uhrig1, Frank Zöllner1, and Lothar Schad1
1Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
Synopsis
No uniform consensus on the optimal sampling
resolution for quantitative DCE MRI exists. In order to increase the sampling
resolution, accelerated parallel imaging techniques (PAT) can be used. The inter- and intra-measurement precision of the
plasma flow (PF) was determined in a phantom depending on the sampling resolution
(1.9-3.8s), achieved by varying the PAT factor (1-6). A significant influence of the
PAT factors and the sampling resolution on the PF itself and the robustness and
reproducibility of the fit could be observed, even with small sampling rates as recommended
in literature.
Introduction
The arterial input function (AIF) has a strong
influence on the results of quantitative DCE MRI using pharmacokinetic modeling
[1]. Fast and stable approaches such as the population average method or the
reference region method do not account for pathophysiological properties of the
arteries of the individual patient's organ to be studied, such as stenosis.
Therefore, if possible, it is preferable to measure the AIF directly. One
parameter that affects the AIF and the goodness of fit is the temporal
resolution Ts. It is commonly recommended that Ts should
be smaller than 10s [2]; yet, no uniform consensus on the optimal Ts
exists [3]. In order to increase the sampling resolution, accelerated parallel
imaging techniques (PAT) can be used.
However, a compromise between acquisition speed and image quality must always
be found, since the signal-to-noise-ratio (SNR) decreases with greater
parallelization. Here, we determined the inter- and intra-measurement precision
of the plasma flow (PF) in a customized two-compartment perfusion phantom for
quantitative DCE MRI depending on the sampling resolution. Different Ts
were achieved by varying the PAT factor.Materials and Methods
Phantom
The phantom is constructed from a dialysis
filter (Helixone FX800, Fresenius Medical Care, Germany) to imitate tissue
characteristics with fibers (diameter = 200µm) close to the capillary size.
The construction principle of the dialyzer enables the transfer of contrast
agent (CA) through a semipermeable membrane (pore diameter = 1.8nm) from
within the fibers to the surrounding space and thus provides two compartments
(Fig. 1). Water (v=20ml/s, typical blood flow velocity in larger human arteries
[4]) and automatically injected CA (Dotarem®, Guerbet, France, (1ml CA and 10ml NaCl)) were pumped through the phantom with an injection rate of 1ml/s. A
tube in front of the phantom inlet served as an artificial artery where the
first pass of CA signal was measured and used to determine the AIF.
Imaging Protocol
DCE MRI data were acquired using a 3T scanner
(Magnetom Skyra, Siemens Healthineers, Germany) and a 15-channel phased-array
knee coil (Siemens Healthineers, Germany). A gradient-echo imaging sequence
(TWIST) with the following scan parameters was used: TR = 3.8ms, TE = 1.4ms,
flip angle (FA) = 15°, matrix size = 128×128×80mm3, spatial
resolution = 1×1×4mm3. The total measurement time for each
measurement was 1 minute, whereby the different temporal resolutions Ts =
[3.8/2.7/2.3/2.1/2.0/1.9]s were achieved. These corresponded to parallel
imaging factors PAT = [1/2/3/4/5/6].
Analysis
The data analysis was based on a manually
selected ROI in the AIF and a ROI in the phantom that remained constant for all
measurements (Fig. 2a). The two-compartment exchange model with four free
parameters was used for pixel-wise determination of PF using Horos (horosproject.org,
USA) and the corresponding perfusion plugin [5]. (1) The intra measurement
precision was determined by the standard deviation σintra of the PF in each ROI of the pixel-wise fit
such as the mean fit quality $$$\mathrm{Chi}^{2} _{intra}$$$. (2) Measurements were repeated five times for
each PAT setting to determine the inter measurement precision by the standard deviation
of the mean values $$$\overline{\mathrm{PF}}_{inter} \pm \sigma_{inter}$$$ among each other. Statistical analysis was performed for all above
mentioned quantities with an unpaired t-test (p<0.05) in Matlab (Matlab 2013a, The
Mathworks, USA).Results
Exemplary fit maps for PF and resulting Chi2 are shown in Fig
2b,c.
(1) Intra measurement variability: For each individual measurement, σintra of the PF within the phantom and the fit quality $$$\mathrm{Chi}^{2} _{intra}$$$ were calculated and are depicted
in a boxplot (Fig. 3a, b) to determine the robustness and performance of the
fit. The statistical evaluation showed that σintra of the PF is significantly smaller for PAT=5 and PAT=6.
For $$$\mathrm{Chi}^{2} _{intra}$$$ the values for PAT=1 are
significantly higher.
(2) Inter measurement variability: The median and interquartile range of
the mean values $$$\overline{\mathrm{PF}}_{inter} $$$ within the measurement series are
shown in the box plot (Fig. 3c). The unpaired t-test reveals a significant
lower value for PAT=1 and PAT=3 measurements compared to PAT=5, where also a
slightly elevated median is noticeable in the boxplot.Discussion
Reproducible values for the PF
in a human like phantom can be achieved with the applied pharmacokinetic model
[6]. This allowed to perform a patient-independent quantitative protocol review
which examined the influence of parallel imaging and thus the sampling resolution
on AIF and tissue function. A significant influence of PAT and TS
on the PF could be observed. Smaller PAT factors seem to lead to a lower fit
quality whereas higher PAT factors showed lower standard deviations within a
ROI. The higher sampling resolution in this cases will possibly compensate for
the SNR loss from parallel imaging. A comparison with a gound truth would be
necessary in the next step in order to better understand the significant
difference between the mean values of the five measurement series. Furthermore,
the sampling resolutions in this study were very high and still within
the recommended literature range. It remains to be investigated how it compares to an even
slower sampling resolution.Conclusion
Significant differences were
observed in both intra- and inter-measurement variability, which makes clear
that the results for quantitative perfusion parameters depend on modifications
of the PAT settings and TS even within the recommended literature
range.Acknowledgements
This research project is partly
supported of the Research Campus M2OLIE funded by the German Federal Ministry
of Education and Research (BMBF) within the Framework “Forschungscampus:
public-private partnership for Innovations” under the funding code 13GW0388A.References
[1] Luypaert R, Sourbron S, Makkat S, de Mey J. Error estimation for
perfu-sion parameters obtained using the two-compartment exchange model
indynamic contrast-enhanced MRI: a simulation study.Phys Med
Biol.2010;55:6431–6443.
[2] DCE‐MRI Technical Committee. DCE‐MRI
Quantification Profile, Quantitative Imaging Biomarkers Alliance. Version 1.0.
Publicly Reviewed Version. QIBA. 2012. Available from RSNA.org/QIBA
[3] Ingrisch M, Dietrich O,
Attenberger UI, et al. Quantitative pulmonaryperfusion magnetic
resonance imaging: influence of temporal resolutionand signal-to-noise
ratio.Invest Radiol. 2010;45:7–14.
[4] Nichols, Wilmer W., Michael
F. O'Rourke, and M. McDonald’s. "Blood flow in arteries." Lea and Fiberger, Philadelphia, PA
(1990): 270-281.
[5]
F. Zöllner, M. Daab, S. Sourbron, L. Schad, S. Schoenberg and G. Weisser. An open source software for analysis of dynamic
contrast enhanced magnetic resonance images: UMMPerfusion revisited. BMC Med
Imaging, 2016, 16 (7), pp.1-13
[6] Gaa T, Schad L
and Zöllner F., Feasibility Study of a Dialyzer as a Multi-Compartment
Perfusion Phantom for Microvascular Tracer Kinetic Modelling. Proc. Int. Soc.
Magn. reson. Med. , Honolulu. HI. USA , (2017) 25, p.1922 .