Arun Joseph1,2,3, Tom Hilbert4,5,6, Gabriele Bonanno1,2,3, Emilie Mussard4,5,6, Christoph Forman7, Ashley Stewart8,9, Kieran O’Brien8, and Tobias Kober4,5,6
1Advanced Clinical Imaging Technology, Siemens Healthcare AG, Bern, Switzerland, 2Translational Imaging Center, Sitem-Insel, Bern, Switzerland, 3Departments of Radiology and Biomedical Research, University of Bern, Bern, Switzerland, 4Advanced Clinical Imaging Technology, Siemens Healthcare AG, Lausanne, Switzerland, 5Department of Radiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland, 6LTS5, École Polytechnique Fédérale de Lausanne (EPFL), Lausanne, Switzerland, 7Siemens Healthcare AG, Erlangen, Germany, 8Siemens Healthcare Pty Ltd, Brisbane, Australia, 9Queensland Brain Institute, The University of Queensland, Brisbane, Australia
Synopsis
Quantitative
susceptibility mapping exploits magnetic susceptibility to probe biological
tissue. The obtained quantitative information carries additional clinical
value, e.g. for the diagnosis of neurodegenerative diseases. However, standard
protocols have long acquisition times which impedes their use in clinical
routine. Here, we propose a compressed sensing acquisition based on a Cartesian
spiral-phyllotaxis readout scheme to drastically reduce the total acquisition
times from 20 minutes to 2 minutes for a 1mm isotropic susceptibility map of
the whole brain. A preliminary qualitative and quantitative validation is
performed on healthy subjects.
Introduction
Quantitative Susceptibility Mapping (QSM) is a
technique that provides information on underlying tissue properties such as
iron distribution, (de-)oxyhaemoglobin, calcification, and others1-3.
It has, amongst others, the potential to provide new clinical insights for
neurodegenerative diseases such as multiple sclerosis and Parkinson’s. Current
implementations of QSM exhibit, however, long scan times (~10-20 mins) with
sufficiently high spatial resolution; this can be mitigated using classical
parallel imaging techniques, but scan times remain long for application in
clinical protocols.
In this study, we implement a compressed sensing QSM
gradient echo sequence based on a Cartesian spiral-phyllotaxis readout4
to achieve a clinically feasible QSM acquisition time. A preliminary
qualitative and quantitative validation is performed on healthy subjects.Methods
A Cartesian spiral-phyllotaxis
undersampling
scheme was implemented in a prototype gradient echo (GRE) sequence. Three subjects were
measured at 3T (MAGNETOM Prisma, Siemens Healthcare, Erlangen, Germany) using a
64-channel head/neck-coil after obtaining written informed consent. The protocol
consisted of undersampled acquisitions using the above prototype sequence with
the following scan parameters: TR 25 ms, TE 20 ms, flip angle 12°, 1 mm
isotropic resolution, FOV 256 mm, number of slices 160, slice oversampling 20%.
The measurements were performed in the transversal plane with undersampling
factors of [2, 4, 6, 8], yielding total acquisition times of [8:06, 4:06, 2:46,
2:02] minutes, respectively. For comparison, a reference measurement with the
same parameters was performed using a product GRE without and with GRAPPA x2
acceleration resulting in acquisition times of 20:30 and 11:09 min,
respectively. Magnitude and phase images as well as QSM maps were directly obtained
at the scanner after image reconstruction followed by a prototype QSM algorithm8,9. Additionally, an MPRAGE
sequence was acquired and used as anatomical reference (TR 2300 ms, TE 2.32 ms,
TI 900 ms, flip angle 8°, 1 mm isotropic resolution).
The compressed sensing reconstruction5-6
process consisted of iteratively minimizing the cost function enforcing both
consistency with the acquired data and sparsity in the wavelet domain following:
$$\underset{x}{min} ||PF\{CX\} - Y||_2^2 + \lambda |\Psi X|_1$$
with P being the sampling mask, F the discrete Fourier
transform, C complex coil sensitivities, Y the undersampled k-space, λ a
regularization parameter, Ψ the wavelet-transform and Х the estimated image. The
complex coil sensitivities were computed by low-pass filtering and normalizing
the individual coil images of the calibration lines as it was done in the
original SENSE publication7. The inline reconstruction time for
compressed sensing reconstructions was ~3 minutes. The algorithm for QSM
estimation was comprised of Laplacian phase unwrapping, V-SHARP background
phase removal, and dipole inversion using the single-step L2 algorithm with
quadratic smoothing regularization8,9.
A quantitative region of interest (ROI) analysis was
performed using the prototype MorphoBox10 segmentation algorithm on five structures whose
susceptibility is of interest, namely: thalamus, caudate, putamen, pallidum,
amygdala. Of note, the segmentation was performed on the MPRAGE image and the
resulting label maps were copied into the space of each individual QSM maps
using rigid registrations11. The median values from all regions were
extracted using the obtained masks. Subsequently, the median QSM of the
cerebellum was subtracted from the ROI values to correct a systematic offset
originating from the inverse nature of the QSM reconstruction. The mean and
standard deviation across subjects for the different structures and acceleration
methods were compared in a bar-plot.Results and Discussion
Figure 1 shows the measured k-space with the Cartesian spiral-phyllotaxis
patterns for the applied acceleration factors. The
magnitude, phase and QSM maps obtained for reference, GRAPPAx2 and compressed
sensing reconstructions with acceleration factors 4 and 8 (results for x2 and
x6 similar) are shown in Figure 2. The magnitude and QSM maps are qualitatively
consistent and robust even for the higher acceleration factors, providing good anatomical
details. Differences can be observed between the phase images of the reference,
GRAPPAx2 and the compressed sensing reconstructions, with the reference showing
large phase variations. These differences are due to different type of coil
sensitivity estimations. Figure 3 shows the QSM maps obtained for reference,
GRAPPAx2 and compressed sensing reconstructions with all tested acceleration
factors. The QSM maps obtained from compressed sensing reconstructions are
qualitatively comparable to the reference; expectedly, increased blurring is observed
with higher acceleration factors. Figure 4 shows the mean of the median QSM
values over three
subjects obtained for the probed regions of the brain. While thalamus, putamen,
caudate and pallidum showed positive QSM median values and similar results,
amygdala showed negative values. The values obtained from compressed sensing
reconstructions were higher in comparison to the reference and GRAPPAx2 acquisitions.Conclusion
We
implemented and tested a Cartesian spiral-phyllotaxis undersampling
pattern in combination with a QSM processing. The preliminary data shows that
compressed sensing reconstruction provides consistent data among different
acceleration factors. Notably, typical compressed sensing blurring
increases with higher acceleration factors, which should hence be adapted to
the intended application. Further studies with a larger cohort including
patients are required for further validation.Acknowledgements
No acknowledgement found.References
-
Duyn J, MR
susceptibility imaging. J. Magn. Reson. 2013; 229:198–207.
- Haacke EM,
Liu S, Buch S, et al. Quantitative susceptibility mapping: current status and
future directions. Magn. Reson. Imaging 2015; 33(1):1-25.
- Wang Y,
Liu T, Quantitative susceptibility mapping (QSM): Decoding MRI data for a
tissue magnetic biomarker. Magn. Reson. Med. 2015;73: 82–101.
- Forman C,
Piccini D, Grimm R, et al. High-resolution
3D whole-heart coronary MRA: a study on the combination of data acquisition in
multiple breath-holds and 1D residual respiratory motion compensation. MAGMA. 2014;27(5):435-43.
- Wetzl J, Forman C, Wintersperger BJ, et al. High-resolution dynamic CE-MRA of the thorax enabled
by iterative TWIST reconstruction. Magn Reson Med. 2017;77(2):833-840.
- Ma LE,
Markl M, Chow K, et al. Aortic 4D flow MRI in 2 minutes using compressed
sensing, respiratory controlled adaptive k-space reordering, and inline
reconstruction. Magn Reson Med. 2019;81(6):3675-3690.
- Pruessmann KP, Weiger M, Scheidegger MB, et al.
SENSE: sensitivity encoding for fast MRI. Magn Reson
Med. 1999;42(5):952-62.
- Chatnuntawech
I, McDaniel P, Cauley SF et al. Single-step quantitative susceptibility mapping
with variational penalties. NMR Biomed. (2016). doi:10.1002/nbm.3570
- Li W, Wu
B, Liu C. Quantitative susceptibility mapping of human brain reflects spatial variation
in tissue composition. NeuroImage 2011; 55:1645–1656.
- Schmitter D, Roche
A, Maréchal B et al. An evaluation of volume-based morphometry for prediction
of mild cognitive impairment and Alzheimer's disease. Neuroimage Clin. 2014 ; 7:7-17.
-
Klein
S, Staring M, Murphy K, Viergever MA, Pluim JPW, "Elastix: a toolbox for
intensity based medical image registration," IEEE Transactions on Medical
Imaging 2010; 29(1):196-205.