Baiyan Jiang1 and Weitian Chen1
1Department of Imaging and Interventional Radiology, The Chinese University of Hong Kong, Hong Kong, Hong Kong
Synopsis
The
measurement of R1rho (1/T1rho) spectrum and its asymmetry have several
advantages over Chemical Exchange Saturation Transfer (CEST) to probe chemical
exchange effect. Previously reported AC-iTIP approach is able to obtain
R1rho-spectrum and R1rho asymmetry robustly. However, it suffers from long scan
time. In this work, we proposed a new AC-iTIP approach to reduce the scan time
by approximately 30%~40%, without compromising the robustness.
Introduction
The MRI imaging based on endogenous
chemical exchange mechanism, such as Chemical Exchange Saturation Transfer
(CEST), can be used to probe metabolites and proteins level in tissue. The R1rho
asymmetry imaging may have certain advantages compared to CEST.1,2 However, these
approaches suffer from B1 RF and B0 field inhomogeneities. Recently, an
approach termed Adiabatic Continuous wave constant amplitude spin-lock for
irradiation with Toggling Inversion Preparation (AC-iTIP) was proposed to
improve the robustness of R1rho asymmetry imaging.3 However, this approach
suffers from a relatively long scan time. In this work, we report a revised pulse
sequence of AC-iTIP to reduce its scan time by approximately 30%~40%. Simulations,
phantom, and in vivo experiments were used to demonstrate the proposed approach.Methods
In
AC-iTIP, a toggling inversion pulse1 and crushers are applied
preceding ACCSL RF pulses.4,5 Although ACCSL is able to lock the
spins along the effective spin-lock field in presence of B1 and B0
inhomogeneities, the B0 inhomogeneity can result in a discontinuity of the
Z-spectrum, whereas R1rho-spectrum is not affected.3 In the original
iTIP approach,1 R1rho asymmetry can be calculated without obtaining
a full R1rho-spectrum. However, due to the use of adiabatic RF pulses in AC-iTIP,3 additional
data acquisition is needed to obtain a full R1rho-spectrum to calculate R1rho asymmetry, which prolongs the scan time.
When
using AC-iTIP and there is no B0 field inhomogeneity, the adiabatic pulse will
align the spins with the effective spin-lock field in parallel if frequency
offset (FO) is positive, and in anti-parallel if FO is negative. The flip angle
is smaller than 90 degrees at any FO. However, consider the scenario in Figure 1:
positive B0 inhomogeneity and negative FO, with the sum of them bigger than 0. Now
the adiabatic pulse will align the spins in anti-parallel due to negative FO. Therefore,
instead of path 1, the spins will follow path 2. The magnetization flips more
than 90 degrees until the sum of B0 and FO is smaller than 0. This results in
distortion of the spectrum between FO=0Hz and FO+B0=0Hz. The similar situation
applies to the negative B0 inhomogeneity and positive FO.
The discontinuity can be corrected by reversing the FM component of the adiabatic
pulse, as shown in Figure 2. Namely, the adiabatic pulse will align the spins
with the effective spin-lock field in anti-parallel if FO is
positive, and in parallel if FO is negative. The reversing equation is
as follow:$$FM2=\left\{\begin{matrix}-FM1-2\times \begin{vmatrix}FO\end{vmatrix}, if FO\leq 0\\ -FM1+2\times \begin{vmatrix}FO\end{vmatrix}, if FO> 0\end{matrix}\right.\ ,[1]$$where
FM1 is the standard FM waveform, FM2 is the reversed waveform, and FO is the
frequency offset. Combined with a B0 map acquisition, we can replace the
distorted regions with the correct ones.
We performed the proposed approach using
simulation and compared both the AC-iTIP3 and the proposed
approach using phantom and in vivo experiments. Simulations were performed
using parameters of cartilage, five pool-population ratios of metabolites were selected
from 0 to 0.01. Three groups of B1 RF and B0 field inhomogeneities were simulated:
control (no B1 RF and B0 inhomogeneity), moderate (90% B1 and 50Hz B0) and severe
(80% B1 and -75Hz B0). Simulation data were performed at FO from -500 to 500Hz. For moderate and severe cases, reversed FM data were simulated at FO from -50 to 50Hz and -75 to 75Hz, respectively. Phantoms and in vivo knee data sets were acquired from a Philips
Achieva TX 3.0T system (Philips Healthcare, Best, the Netherlands) using
eight-channel head and knee coils, respectively. For both phantom and in vivo
experiments, scans were performed at FO from -300 to 300Hz with 25Hz increment,
reversed FM data were collected at FO ±100Hz, ±75Hz,
±50Hz
and ±25Hz.
Six tubes of myo-inositol with concentration 0, 20, 30, 50, 100, 150 and 200mM were dissolved in phosphate-buffered saline (pH=7.4), and added with 0.2mM MnCl2 to modulate the water relaxation R1 and R2 to ~1.6 and 20Hz, respectively. Volunteer scan was conducted under the approval of the
Institutional Review. 2D Fast Spin Echo was used for imaging data acquisition.
Frequencies of spin-lock (FSL) for simulation, phantom and in vivo experiments
was 150Hz. Time of spin-lock was 60ms. Scan time reduction was about 30%. An
order of 15 polynomial fitting was used to fit the spectrum to perform
asymmetry analysis. R1rho asymmetry was calculated at 0.9ppm to 1.1ppm.Results and Discussion
Figure
3 shows the simulation results. The reversed FM is able to correct the discontinuity
and obtain a smooth Z-spectrum. Note R1rho asymmetry signal reduces with increasing B1 RF inhomogeneity, which affects the optimal FSL.3 Figure 4 shows the phantom results. Note both AC-iTIP and the proposed approach show a similar linear relationship between the R1rho asymmetry and the concentration of metabolites. Figure 5 shows the in vivo results.
Note the proposed approach resulted in a comparable R1rho asymmetry map
compared to AC-iTIP, but the scan time is reduced by 30%.Conclusion
The
preliminary results demonstrated the proposed approach can significantly reduce
the scan time of AC-iTIP without comprising its robustness in the presence of field
inhomogeneities. The proposed approach has the potential to probe metabolites in
vivo.Acknowledgements
This study is supported by a grant
from the Research Grants Council of the Hong Kong SAR (Project SEG CUHK02), and
a grant from the Innovation and Technology Commission of the Hong Kong SAR
(Project MRP/001/18X). References
- Jin T, and Kim S-G. Quantitative Chemical Exchange Sensitive MRI
Using Irradiation with Toggling Inversion Preparation. Magnetic Resonance in Medicine, 2012,
68(4), pp. 1056–1064., doi:10.1002/mrm.24449.
- Ling
W, Nicholls, F. J, Jin T, et al. Selective Detection of Chemical Exchange
Specific R1rho by iTIP gagCEST. ISMRM 2014.
- Jiang, B.,
Jin, T., Blu, T., & Chen, W. (2019). Probing chemical exchange using
quantitative spin-lock R1ρ asymmetry imaging with adiabatic RF
pulses. Magnetic Resonance in Medicine, 82(5), 1767–1781. doi:
10.1002/mrm.27868.
- Jiang B and Chen W. On-Resonance and off-Resonance Continuous
Wave Constant Amplitude Spin-Lock and T1ρ Quantification in the Presence of B1
and B0 Inhomogeneities. NMR in
Biomedicine, 2018, 31(7), doi:10.1002/nbm.3928.
- Chen
W. Artifacts correction for T1rho imaging with constant amplitude spin-lock. J
Magn Reson. 2017;274:13‐23. https://doi.org/10.1016/j.jmr.2016.11.002.