Sivakami Avadiappan1, Melanie A Morrison1, Yicheng Chen1, Angela Jakary1, Christopher P Hess1, and Janine M Lupo1
1University of California San Francisco, San Francisco, CA, United States
Synopsis
Although susceptibility weighted imaging
(SWI) has been the standard method for quantifying and evaluating cerebral
microbleeds, more recently total lesion susceptibility obtained from quantitative
susceptibility mapping (QSM) has been found to correlate with measures of
disease burden and can provide more accurate estimates of microbleed volume.
This study found that 2.5 times as many CMBs can be detected on SWI compared to
QSM images, with the number of CMBs missed by QSM increasing as a function of
decreasing CMB size, highlighting the importance of including SWI for a more
accurate assessment of disease burden in these patients.
Introduction
Cerebral
microbleeds (CMBs)
appear as small hypointense lesions, with spherical shape and
a maximum diameter of <10mm on T2*-weighted
gradient echo (GRE) magnitude images1. Representing focal
accumulations of hemosiderin deposits2 with paramagnetic properties that
cause signal loss due to susceptibility effects, CMBs are commonly observed in patients with cerebrovascular and
neurodegenerative diseases1 such as stroke, Alzheimer’s disease, cerebral
amyloid angiopathy,
traumatic brain injury (TBI), irradiated brain tumors, dementia, and
also in normal aging. Quantification of CMB properties such as their number,
size, and location have been associated with disease severity and neurocognitive
decline. Although
susceptibility weighted imaging (SWI) has been the standard method for
quantifying and evaluating CMBs, more recently total lesion susceptibility
obtained from quantitative susceptibility mapping (QSM) has been found to correlate
with measures of disease burden3. Because QSM reconstructs the
source of a local susceptibility perturbation rather than the resulting dipole
field, the technique can also provide more accurate estimates of CMB size
compared to SWI. Although prior studies have directly compared CMB diameter measurements
from QSM and SWI images from CMBs that were large enough to be visualized on
QSM images3,4, we hypothesize that smaller CMBs may go undetected on
QSM without the “blooming” effect that is inherent in the magnitude images
underlying the SWI. The goal of this
study was to directly compare the number of CMBs detected on QSM and SWI images
and assess whether a relationship exists between CMB conspicuity and size. Methods
Subjects: A total of ten
patients with radiation-induced CMBs due to prior radiation therapy of a glioma
between 2 and 15 years prior were scanned on a GE MR950 7T scanner. A subset of these patients also underwent 3T MR
imaging on the same day with a similar protocol. This mixed
population allowed for a wide range of CMB locations, contrasts, and sizes,
thus creating a broad spectrum of detection sensitivities from which the QSM
and SWI images could be compared.
MRI Protocol: At 7T, a 3D multi-echo gradient-recalled sequence (4
echoes, TE = 6/9.5/13/16.5ms, TR=50ms, FA=20°, bandwidth=50kHz, 0.8x0.8x1mm resolution, FOV=24x24x15cm) was
performed for SWI using a 32-channel phase-array coil. GRAPPA-based parallel
imaging with an acceleration factor of 3 and 16 auto-calibration lines was
adopted to reduce the scan time. At 3T, an 8-echo scan was used (TE1/ΔTE/TE8= 14/3.4/37.8ms, TR=56ms, 0.9x0.9x2mm
resolution).
Image Processing: SWI and QSM images were
reconstructed according to the pipeline in Figure 1. SWI processing utilized in-house
software,5 while QSM processing employed existing algorithms: a
3D Laplacian-based method for phase unwrapping6, V-SHARP7
for background field removal, and iLSQR8 for dipole field inversion
and reconstruction of the susceptibility maps. CMBs were manually labeled on
QSM images by an experienced rater, whereas a semi-automatic algorithm
developed previously9,10 was used to segment CMBs on SWI images. This
algorithm provides accurate characterization of the overall CMB burden and the
final output is a binary mask of individual CMBs that was then visualized by
overlaying on SWI images. The
number of CMBs detected from QSM and SWI images were compared using a Wilcoxon
signed rank test and plotted as a function of CMB size discretized into 4
distinct bins. Results
Although the number of CMBs detected on SWI
and QSM were highly correlated (r=0.9025, p<0.0005), there were 2.5 times
(range:1.33-5.25) as many CMBs detected on SWI compared to QSM (p<0.002;
Figure 2). The total susceptibility of CMBs varied between 0.26-0.609 ppm, with
a mean value of 0.45. As shown in Figure 3, for CMB volumes smaller than 5mm3,
only 15% of the microbleeds were detected in QSM. The detection accuracy on QSM
images then improved with increasing CMB volume, with 83% of CMBs larger than
50mm3 detected. Figure 4 depicts representative examples of CMBs
detected on both QSM and SWI, only SWI and not QSM, and two that were
mislabeled as a vein on SWI but correctly identified as two CMBs on QSM. All of
the CMBs that were detected on the QSM images were also present on the SWI,
except for the example in Figure 4C. Reviewing the subset of patients who had
3T scans as well revealed even fewer CMBs on QSM images, an example of which is
shown in Figure 5.Discussion & Conclusions
Although QSM is
beneficial for the evaluation of CMB susceptibility and for accurate volume
quantification, SWI is more sensitive to detecting smaller CMBs due to the
blooming effect, regardless of field strength. However, if many larger CMBs are
present close enough together, this same enhancement can incorrectly cause
these CMBs to be misclassified as a vessel on SWI. For studies that require QSM
to evaluate total susceptibility of small lesions, an approach that first uses
SWI for identifying lesions would provide a more accurate assessment of disease
burden in these patients.Acknowledgements
The authors would like to acknowledge the support received from the
following grant R01HD079568.References
1. Charidimou A, Werring DJ. Cerebral microbleeds: detection,
mechanisms and clinical challenges. Future Neurology. 2011;6:587–611.
2. Fazekas
F, Kleinert R, Roob G, et al. Histopathologic analysis of foci of signal loss
on gradient-echo T2*-weighted MR images in patients with spontaneous
intracerebral hemorrhage: evidence of microangiopathy related microbleeds. AJNR
Am J Neuroradiol. 1999;20(4):637–642.
3. Liu
T, Surapaneni K, Lou M, Cheng L, Spincemaille P, Wang Y. Cerebral microbleeds:
burden assessment by using quantitative susceptibility mapping. Radiology.
2012;262(1):269–278.
4. Cronin
MJ, Wang N, Decker KS, Wei H, Zhu WZ, Liu C. Exploring the origins of
echo-time-dependent quantitative susceptibility mapping (QSM) measurements in
healthy tissue and cerebral microbleeds. Neuroimage. 2017;149:98–113.
5. Bian W, Hess CP, Chang SM, Nelson SJ, Lupo JM.
Susceptibility-weighted MR imaging of radiation therapy-induced cerebral
microbleeds in patients with glioma: a comparison between 3T and 7T. Neuroradiology.
2014;56(2):91–96.
6. Li W, Wu B, Liu C. Quantitative susceptibility
mapping of human brain reflects spatial variation in tissue composition. Neuroimage. 2011; 55: 1645– 1656.
7. Wu B, Li W, Guidon A, Liu C. Whole brain
susceptibility mapping using compressed sensing. Magn Reson Med.
2012;67:137–147
8. Li W, Wang N, Yu F, Han H, Cao
W, Romero R, Tantiwongkosi B, Duong TQ, Liu C. A
method for estimating and removing streaking artifacts in quantitative
susceptibility mapping. Neuroimage. 2015; 108: 111– 122.
9.
Bian
W, Hess CP, Chang SM, Nelson SJ, Lupo JM. Computer-aided detection of
radiation-induced cerebral microbleeds on susceptibility-weighted MR images. NeuroImage Clin. 2013; 2:282–290.
10. Morrison
MA, Payabvash S, Chen Y, et al. A
user-guided tool for semi-automated cerebral microbleed detection and volume
segmentation: Evaluating vascular injury and data labelling for machine
learning. NeuroImage Clin. 2018;
20:498–505.