3214

Sensitivity of Quantitative Susceptibility Mapping to Temperature: A Phantom Study
Hirohito Kan1,2, Yuto Uchida3,4, Masahiro Takizawa5, Tosiaki Miyati6, Hiroshi Kunitomo7, Nobuyuki Arai7, Harumasa Kasai7, and Yuta Shibamoto2
1Radiological and Medical Laboratory Sciences, Nagoya University Graduate School of Medicine, Nagoya City, Japan, 2Department of Radiology, Nagoya City University Graduate School of Medical Sciences, Nagoya City, Japan, 3Department of Neurology, Nagoya City University Graduate School of Medical Sciences, Nagoya City, Japan, 4Department of Neurology, Toyokawa City Hospital, Toyokawa, Japan, 5Healthcare Business Unit, Hitachi, Ltd., Tokyo, Japan, 6Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Kanazawa, Japan, 7Department of Radiology, Nagoya City University Hospital, Nagoya City, Japan

Synopsis

To date, we are unaware of any reports regarding quantitative susceptibility mapping (QSM) at different temperatures. To clarify the temperature dependence of susceptibility estimated by QSM analysis, we investigated the relationships between temperature and susceptibility using a simple cylinder phantom with varying temperatures. This study has demonstrated that a significant inverse correlation was found between the temperature and the susceptibility value estimated by QSM analysis. This dependence might be the confounding factor of QSM-based iron estimation.

Introduction

Quantitative susceptibility mapping (QSM) analysis enables to detect iron concentration and distribution from multi-echo phase images applied several complicated post-processing [1]. A similar iron evaluation has also been performed by the R2* relaxometry in the clinical setting, such as the brain and liver. However, it is well-known that R2* value depends on the temperature, which is the confounding factor of R2* measurement [2]. To clarify the temperature dependence of susceptibility estimated by QSM analysis, we investigated the relationships between temperature and susceptibility and R2* values using a simple cylinder phantom with varying temperatures.

Material and Methods

Phantom preparation
A schematic drawing of the cylinder phantom is shown in Fig. 1a. The six solutions with various concentrations of superparamagnetic iron oxide (SPIO) nanoparticles (0.0125, 0.025, 0.05, 0.1, 0.2, and 0.4 mM) were employed. The solutions were sealed to plastic tubes (φ = 6 mm, wall thickness = 0.05 mm). These tubes were placed in a cylinder phantom and were filled with water around to minimize the interaction in the air-water boundary. Fig. 1b illustrates the experimental setup of the phantom. The cylinder phantom was placed on an axis parallel to the magnetic field. The desired temperature water was circulated by the pomp around the cylinder phantom from the water bath with temperature control. The temperature of the circulated water was adjusted to change the temperature in the cylinder phantom to 25.8, 29.8, 31.6, 33.6, 35.6, 38.0, 40.0, and 42.5 ℃. The heat-insulated container was used to help maintain the temperature in the cylinder phantom. The water temperature during the examination was assured to measure the temperature by thermometer before and after every scan.

MRI experiment and data analysis
The cylinder phantom was scanned in the transverse plane with a 3T MRI (Hitachi Ltd., Tokyo, Japan). The magnitude and phase images were obtained using 3D multiple spoiled gradient echo sequence with the following parameters: FOV, 160 × 160 ×140 mm3; matrix size, 80 × 80 × 70 (zipped 160 × 160 × 140); TR, 32 ms; TE, 4.2-­27.4 ms at 2.9-ms intervals; number of echo, 9; and flip angle, 20°. The pre-scan was performed at every changing temperature. The local field was calculated from the multiple phase images by two-step background field removal, consisting of the Laplacian boundary value method [4] followed by iterative spherical mean value method [5] with the filter size of 5 mm. Next, the susceptibility map was estimated by the two-step non-linear morphology enabled dipole inversion algorithm [6] like a streaking artifact reduction for the QSM algorithm [7] to minimize the streaking artifact induced by high concentration solutions. The mean susceptibility of water in the cylinder phantom was used for zero reference of susceptibility. The R2* map was calculated from the magnitude images using auto-regression on linear operations algorithm [8]. The relationships between temperature and susceptibility and R2* values were determined. Moreover, the temperature coefficients of susceptibility (Tc-χ) and R2* (Tc-R2*) [2] from measurements at different temperatures were calculated at each concentration, and the linearities in these indices against the SPIO concentration were validated.

Results

The resultant susceptibility and R2* maps in 25.8 ℃ are shown in Fig. 2a and b. Fig. 2c demonstrates that the susceptibility and R2* values at an SPIO concentration of 0.4 mM decreased with increasing temperature. There were clear temperature dependences of susceptibility and R2* values in each SPIO concentration (Fig.3). The significant inverse correlations were found between the temperature and susceptibility and R2* values at each SPIO concentration, except to 0.0125 mM in R2* (Table1). There were strong linearities between the SPIO concentration and Tc-χ (r = -0.99, P < 0.001) and Tc-R2* (r = -0.99, P < 0.001) (Fig. 4).

Discussions

It was successful in depicting the susceptibility of iron changed by the temperature using QSM analysis. The negative correlations between the susceptibility and R2* values can be explained by the decrease of paramagnetic iron susceptibility with increasing temperature based on Curie’s law [2, 8]. The susceptibility estimated by QSM analysis is a relative value based on proton resonance frequency (PRF) of water changed by the temperature [9]. However, the effect of the PRF shift of water to QSM analysis could be minimized by the use of water susceptibility in the cylinder phantom as zero reference in this study.
Moreover, Tc-χ and Tc-R2* linearly decreased with SPIO concentration. The volume susceptibility (χ) in a particular volume (V) considering the temperature is approximated as
$$\chi\approx\frac{\sum_{n=1}^k(\alpha_{n}T+\chi_{n, 0})V_{n}}{V}$$
where k is the number of particles, T is the Celsius temperature, αn is the Tc-χ of particle n, Vn is the volume of particle n, and χn,0 is the susceptibility of particle n at 0 ℃. According to the above equation, the susceptibility in the voxel is proportional to the temperature and the number of particles (i.e., iron content in this study). Therefore, since the Tc-χ in the voxel varies considerably with the iron contents, the temperature might be the confounding factor of QSM-based iron estimation, similar to the R2* measurement.

Conclusion

There is a significant inverse correlation between the temperature and the susceptibility value estimated by QSM analysis. This dependence may be the confounding factor of the QSM-based iron estimation.

Acknowledgements

This work was supported by JSPS KAKENHI Grant Number JP17K15805.

References

[1] Deistung A, Schweser F, Reichenbach JR. Overview of quantitative susceptibility mapping. NMR Biomed 2017;30(4):e3569.

[2] Birkl C, Carassiti D, Hussain F, Langkammer C, Enzinger C, Fazekas F, et al. Assessment of ferritin content in multiple sclerosis brains using temperature-induced R*2 changes. Magn Reson Med 2018;79(3):1609-15.

[3] Zhou D, Liu T, Spincemaille P, Wang Y. Background field removal by solving the Laplacian boundary value problem. NMR Biomed 2014;27(3):312-9.

[4] Wen Y, Zhou D, Liu T, Spincemaille P, Wang Y. An iterative spherical mean value method for background field removal in MRI. Magn Reson Med 2014;72(4):1065-71.

[5] Liu T, Wisnieff C, Lou M, Chen W, Spincemaille P, Wang Y. Nonlinear formulation of the magnetic field to source relationship for robust quantitative susceptibility mapping. Magn Reson Med 2013;69(2):467-76.

[6] Wei H, Dibb R, Zhou Y, Sun Y, Xu J, Wang N, et al. Streaking artifact reduction for quantitative susceptibility mapping of sources with large dynamic range. NMR Biomed 2015;28(10):1294-303.

[7] Pei M, Nguyen TD, Thimmappa ND, Salustri C, Dong F, Cooper MA, et al. Algorithm for fast monoexponential fitting based on Auto-Regression on Linear Operations (ARLO) of data. Magn Reson Med 2015;73(2):843-50.

[8] Birkl C, Langkammer C, Krenn H, Goessler W, Ernst C, Haybaeck J, et al. Iron mapping using the temperature dependency of the magnetic susceptibility. Magn Reson Med 2015;73(3):1282-8.

[9] Rieke V, Butts Pauly K. MR thermometry. Journal of magnetic resonance imaging: J Magn Reson Imaging 2008;27(2):376-90.

Figures

Fig. 1 (a) Schematic drawing of the cylinder phantom. The Cylinder phantom consists of six plastic tube sealed solutions of various concentrations of superparamagnetic iron oxide (SPIO) nanoparticles. (b) Experimental setup of the phantom. The warmed water was circulated around the cylinder phantom from the water bath with temperature control to adjust the temperature in the cylinder phantom. The heat-insulated container helped to maintain the temperature in the cylinder phantom during the examination.

Fig. 2 Representative (a) susceptibility and (b) R2* maps of the cylinder phantom at 25.8 ℃. (c) Susceptibility (χ) and R2* values at superparamagnetic iron oxide (SPIO) nanoparticles of 0.4 mM decreased with increasing the temperature.

Fig. 3 Relationships between temperature and (a) susceptibility and (b) R2* values. There were the inverse correlations between the temperature and susceptibility and R2* values.

Fig. 4 Linear regression analyses of the temperature coefficient of susceptibility (Tc-χ) and R2* (Tc-R2*) at each concentration of superparamagnetic iron oxide (SPIO). The Tc-χ and Tc-R2* were inversely proportional to the iron content in the solution.

Table 1 Linear regression equations between the temperature and susceptibility and R2* values at each concentration of superparamagnetic iron oxide (SPIO) nanoparticles.

Proc. Intl. Soc. Mag. Reson. Med. 28 (2020)
3214