Collin J Buelo1,2, Ruiyang Zhao1,2, Ante Zhu2,3, Scott B. Reeder1,2,3,4,5, and Diego Hernando1,2,3
1Medical Physics, University of Wisconsin-Madison, Madison, WI, United States, 2Radiology, University of Wisconsin-Madison, Madison, WI, United States, 3Biomedical Engineering, University of Wisconsin-Madison, Madison, WI, United States, 4Medicine, University of Wisconsin-Madison, Madison, WI, United States, 5Emergency Medicine, University of Wisconsin-Madison, Madison, WI, United States
Synopsis
Chemical
shift-encoded (CSE) MRI can be used to simultaneously obtain multi-parametric
maps, including R2*, proton density fat fraction (PDFF), and tissue
susceptibility, to aid in the diagnosis of diffuse liver disease. However, the
presence of fat in tissue can confound both R2* and B0 field estimation and
consequently quantitative susceptibility mapping (QSM). To validate this multi-parametric mapping, comparison of
quantitative maps to phantoms with known iron concentrations were performed.
This work demonstrated the feasibility of accurate PDFF mapping, and R2* and susceptibility
measurements that are linearly dependent on both iron and fat concentration.
Introduction
Chemical
shift-encoded (CSE) MRI techniques enable the simultaneous quantification of proton-density
fat-fraction (PDFF, an established biomarker of triglyceride concentration),
and R2* (an emerging biomarker of liver iron content). Further, CSE-MRI also enables
mapping of the B0 field, which can be used for the estimation of tissue susceptibility
in quantitative susceptibility mapping (QSM). Importantly, tissue
susceptibility is also showing promise as an emerging biomarker of iron
quantification, which may provide highly complementary information with R2* regarding
iron deposition[1].Thus, comprehensive
assessment of liver iron content using R2* and susceptibility simultaneously
measured using CSE-MRI
is of great interest. However,
R2*, susceptibility and PDFF may confound each other [2] and previous studies on
susceptibility-based iron quantification have not considered the effect of fat[3]. Highly-controlled
validation of the accuracy of R2*, susceptibility and PDFF from CSE-MRI
acquisitions is highly desirable. Therefore,
the purpose of this work was to validate CSE-MRI techniques for simultaneous
estimation of PDFF, R2*, and susceptibility in a fat-iron phantom that closely
mimics liver MRI signals in the presence of fat and iron deposition.Methods
Twenty
four vials were constructed to mimic the simultaneous presence of fat and iron
using a recently proposed phantom formulation[4]. Four sets of vials, each
with constant fat fraction, were constructed. The nominal fat fractions
designed in each set were 0%, 10%, 20%, and 40%. Each set contained six vials of
different iron concentration, obtained by varying the concentration of iron
particles (2.9 μm diameter magnetite
spheres, COMPEL, Bangs Labs, Fishers, IN). The four sets were separated into
two batches of 12 vials each, which were scanned sequentially. The 12 vials in each
batch were placed in a holder within a custom designed spherical housing filled
with distilled water. Phantoms were placed in the scanner with the long axis of
the vials parallel to the main magnetic field.
Images
were acquired on a clinical 3T MR system (GE Healthcare Signa Premier,
Waukesha, WI) using a 48 channel phased array head coil. 3D multi-echo spoiled
gradient recalled echo (SGRE) images were acquired using the following imaging
parameters: 25x25x17.2 cm3 FOV, 128x128x86 imaging matrix,
1.95x1.95x2 mm3 resolution, 6 echoes with initial echo time 1.1 ms, 0.9
ms echo spacing, echo train length of 3, repetition time 7.2 ms, and 1° flip angle to avoid T1-related bias.
Image
reconstruction and data analysis were performed offline. Images of the
multichannel data are reconstructed using the Walsh method[5]. Signal fitting was
performed using a graphcut method to avoid fat-water swaps [6] followed by
voxel-independent fitting of a signal model including R2*, B0 field, and PDFF [7]. Subsequently, tissue
susceptibility was calculated using an iterative L2 regularized reconstruction[3].
Region-of-interest
(ROI) measurements of PDFF, R2*, and susceptibility values were performed at
the center of each vial. The susceptibility of each vial was measured using the
adjacent water bath as the susceptibility reference. Linear regressions were
performed to validate PDFF measurements with the known fat fraction, and to assess
the relationship between R2* and iron concentration. A multi-variable linear
regression was performed to assess the relationship between susceptibility and
both iron concentration and PDFF.Results
Reconstructed
R2*, PDFF, and susceptibility images can be seen in Figure 1. Accurate PDFF
values were obtained at varying iron concentrations (Figure 2). Fat-corrected R2*
measurements ranged from 27 s-1 to 811 s-1 and were observed
to correlate linearly with iron concentration across PDFF values
(slope
= 7567 s-1 ml/mg, intercept = 27.2 s-1, r2 =
0.99) (Figure 3).
Susceptibility
was observed to linearly correlate with both iron concentration and PDFF
(Figure 4a). A multi-variable linear fit was obtained: $$$ \chi = -0.23+0.011PDFF+19.2\rho_{Fe} $$$ with high r2 = 0.982. If PDFF is not considered in the regression, a lower
coefficient of determination is observed (r2 = 0.903) (Figure 4b). Previous works show human fat tissue to have a
higher susceptibility than water [8][9], which agrees with the direction of the shift observed
in these phantom experiments.Discussion
This
work validated the
accuracy of simultaneous
CSE-based mapping of R2*, susceptibility and PDFF in a tissue-mimicking
fat-iron phantom. Importantly, the presence of fat affects tissue
susceptibility,
and consequently susceptibility-based iron quantification if unaccounted for. Therefore,
correction for the effects of fat may be needed for susceptibility-based iron
quantification[10]. Fortunately, PDFF is
readily estimated from the same CSE-MRI acquisition, providing additional
synergy between the estimated parameters. These results suggest the potential
for comprehensive tissue characterization from a single CSE-MRI acquisition in
diffuse liver disease.
A limitation of this work is that the experiments were
performed solely on phantoms, in order to enable highly controlled fat and iron
concentration. Future work includes validation of this approach in-vivo in
patients with fat and iron deposition in the liver. Additionally, in this
phantom iron was added in the form of magnetite microspheres. Different forms
of iron will generally result in different MRI properties.Acknowledgements
We wish to acknowledge support from the NIH (R01 DK083380, R01 DK088925, K24 DK102595) and from the University of Wisconsin SEED grant program. Further, we wish to acknowledge GE Healthcare and Bracco Diagnostics who provides research support to the University of Wisconsin. Finally, Dr. Reeder is a Romnes Faculty Fellow, and has received an award provided by the University of Wisconsin-Madison Office of the Vice Chancellor for Research and Graduate Education with funding from the Wisconsin Alumni Research Foundation.References
[1] T. J. Colgan, G.
Knobloch, S. B. Reeder, and D. Hernando, “Sensitivity of quantitative
relaxometry and susceptibility mapping to microscopic iron distribution,” Magn.
Reson. Med., vol. 83, no. 2, pp. 673–680, 2020.
[2] S. B. Reeder, I.
Cruite, G. Hamilton, and C. B. Sirlin, “Quantitative Assessment of Liver Fat
with Magnetic Resonance Imaging and Spectroscopy,” J. Magn. Reson. Imaging
JMRI, vol. 34, no. 4, p. spcone, Oct. 2011.
[3] S. D. Sharma, D.
Hernando, D. E. Horng, and S. B. Reeder, “Quantitative susceptibility mapping
in the abdomen as an imaging biomarker of hepatic iron overload: QSM as a
Biomarker of Hepatic Iron Overload,” Magn. Reson. Med., vol. 74, no. 3,
pp. 673–683, Sep. 2015.
[4] R. Zhao, G. Hamilton,
J. H. Brittain, S. B. Reeder, and D. Hernando, “A Quantitative MRI Phantom to
Mimic the Simultaneous Presence of Fat, Iron, and Fibrosis,” presented at the
ISMRM, 2019, vol. 1777.
[5] D. O. Walsh, A. F.
Gmitro, and M. W. Marcellin, “Adaptive reconstruction of phased array MR
imagery,” Magn. Reson. Med., vol. 43, no. 5, pp. 682–690, May 2000.
[6] D. Hernando, P.
Kellman, J. P. Haldar, and Z.-P. Liang, “Robust water/fat separation in the
presence of large field inhomogeneities using a graph cut algorithm,” Magn.
Reson. Med., p. NA-NA, 2009.
[7] D. Hernando, J. H.
Kramer, and S. B. Reeder, “Multipeak fat-corrected complex R2* relaxometry:
Theory, optimization, and clinical validation,” Magn. Reson. Med., vol.
70, no. 5, pp. 1319–1331, 2013.
[8] C. Wang, B. A. Ratliff,
C. B. Sirlin, S. B. Reeder, and D. Hernando, “Monte Carlo Modeling of Liver MR
Signal in the Presence of Fat,” presented at the ISMRM, 2018, vol. 2264.
[9] S. M. Sprinkhuizen, C.
J. G. Bakker, J. H. Ippel, R. Boelens, M. A. Viergever, and L. W. Bartels, “Temperature
dependence of the magnetic volume susceptibility of human breast fat tissue: an
NMR study,” Magn. Reson. Mater. Phys. Biol. Med., vol. 25, no. 1, pp. 33–39,
Feb. 2012.
[10] J.
Li et al., “Quantitative susceptibility mapping (QSM) minimizes
interference from cellular pathology in R2* estimation of liver iron
concentration,” J. Magn. Reson. Imaging, vol. 48, no. 4, pp. 1069–1079,
2018.