Lucia Bossoni1, Andrew Webb1, Loes Huijnen2, Remco Overdevest2, and Wyger Brink1
1Leiden University Medical Center, Leiden, Netherlands, 2Leiden University, Leiden, Netherlands
Synopsis
Phase-only Helmholtz-Electrical
Property Tomography (PO-HEPT) has recently shown promise for the detection and
grading of diffuse glioma. However, a brain glioma is usually located away from
the centre of the brain where the transceive phase and phase-only approximations
used in reconstruction may not be valid. Here we assessed the sensitivity of PO-HEPT
in an anatomically realistic brain phantom where an off-centre glioma-compartment
was incorporated, and using a clinically applicable MR sequence. Our results
show that while the accuracy of the PO-HEPT deteriorates, its sensitivity is
mostly unaffected, thus allowing correlative studies of tumour grading.
Introduction
Gliomas are known to show higher electrical
conductivity values than the normal brain parenchyma1. Recently, phase-only
Helmholtz-Electrical Property Tomography (PO-HEPT) was employed to distinguish high
grade from lower-grade gliomas with high efficiency2. However, the validity of the assumptions of
PO-HEPT has only been tested on highly symmetrical and homogeneous phantoms.
This study aimed to assess whether the PO-HEPT
reconstruction had enough sensitivity to accurately report conductivity values in
a brain phantom designed to mimic the human brain anatomy, within an off-centre
ROI where the transceive phase approximation (TPA) is less accurate, and using
a clinically relevant sequence taking less than 3 mins of scan time. Methods
A brain phantom
was designed by segmenting the MRI image of a patient with a glioma. The tissue
boundaries were 3D printed with polypropylene, and the CSF, white matter (WM) and grey matter (GM) compartments were filled with a mixture of 2% agarose,
demi-water and NaCl to reach the conductivity of the specific tissue type at 128 MHz: 2.1 S/m (CSF), 0.34 S/m (WM) and 0.6 S/m (GM)3. The
conductivity of the glioma-compartment was varied from 0.6 S/m to 2 S/m, in
steps of 0.2 S/m and consisted of a water-NaCl solution to allow in-situ replacement
in the glioma-compartment.
MRI data were
acquired on a 3T clinical scanner (Ingenia, Philips, Best, The Netherlands)
with a 3D-bSSFP sequence. The scan parameters were: FOV (mm)=240x240x37.5;
acquired voxel size (mm)=2x2x1.5; reconstructed voxel size (mm)=1.25x1.25x1.5; flip
angle 10 degrees; 16 signal averages; TR/TE=3/1.5 ms; Dummy pulses were
included before each acquisition to minimize transient magnetization effects4;
total scan duration was 2:23 mins.
The signal was
received with a 32-channel array head coil, while a body coil was used for
transmission and phase referencing. The phantom was placed at the centre of the
array.
The electrical conductivity
maps were reconstructed from the phase data and from the homogeneous Helmholtz
equation5:
$$\sigma = -\frac{1}{\mu_0 \omega} \left(\frac{\nabla^2e^{i\phi+}}{e^{i\phi+}}\right) $$
where ω is the Larmor frequency, μ0 the magnetic vacuum permeability and φ+
the transmit phase, which was assumed to be φ+= φ±/2 (TPA),
where φ± is the transceive phase.
The Laplacian was implemented
by convolving the phase-data with a 7x3x3 kernel, for in-plane derivatives, and
a 5x3x3 kernel for out-of-plane derivatives6. Conductivity maps were
subsequently convolved with a 2D median filter to mitigate noise-related
reconstruction artefacts, and subsequently three ROIs were drawn in the glioma
compartment. Finally an additional ROI was placed in the ventricles, and this
was used as a control. An ANOVA test with Bonferroni post-hoc correction for
multiple comparisons was performed on the eight different glioma-conductivity
values, for each different post-processing choice. The same statistical test
was used to judge the stability of the method across different acquisitions in
the ventricles' ROI.Results
A
picture of the phantom and the acquired magnitude and phase data are shown
in Fig.1. Raw and filtered conductivity maps are shown
in Fig.2. The accuracy and sensitivity of the method were
tested against different choices of ROI and with different median filter sizes.
As a
“control” to validate the stability of the algorithms for a constant
conductivity, the reconstructed conductivity values in the ventricles' ROI was
quantified and it showed no significant changes in mean conductivity across
different acquisitions, for all sample pairs except for one (Fig. 3).
The
best results in terms of sensitivity were obtained with ROI1 and a median
filter of 4x4 pixels and show that, despite the rather large standard deviation
(∼0.2 S/m), the differences in glioma-conductivities
were statistically significant also between the two samples which differed only
by ∼ 0.1 S/m. The best results in terms of accuracy were
obtained with ROI2. However, these results are much less precise (larger
standard deviation), due to the leakage of the boundary effects in the ROI. Finally, ROI3, although being far from the boundaries
by several pixels (i.e. >> kernel size), was statistically the same as
the other two ROI choices.Discussion
In all ROI
analyses, the conductivity was consistently underestimated and the error
increased with the nominal conductivity (Fig.4),
which is in line with literature6. One explanation could be that the
TPA deteriorates at the periphery of the FOV, as previously reported7,8.
In one study7, when using PO-HEPT, σ underestimated independent conductivity-measurements by
approximately 40% in peripheral regions, in agreement with our findings.
Additionally, 3D-printed boundaries may have a detrimental effect on the
spatial homogeneity of B0 and B1, therefore introducing further
phase errors. Regardless of the above limitations, PO-HEPT was able to detect
statistically significant changes in conductivity across different glioma-samples,
thus showing promise of the approach of using3D bSSFP and PO- HEPT for clinical
glioma applications1. We anticipate that the accuracy of the method
will improve in an in-vivo experiment, where hard boundaries with abrupt
permittivity and magnetic susceptibility changes are absent8.Conclusions
In this study we
assessed the sensitivity of phase-only EPT in a realistic brain phantom with a
glioma compartment. Despite the rather large σ errors, which we speculated to originate from the
partial violation of the TPA phase and phase-only approximations, the method
was able to measure significant differences across samples, therefore
validating its potential use in the clinic. Acknowledgements
This work was supported by the Netherlands Organization for
Scientific Research (NWO) through a VENI fellowship to L. B (016.Veni.188.040)
and to W. B (016.Veni.188.040 and TTW.16820). H. van de Stadt and J. Verhart
are acknowledged for assistance in the phantom preparation. J. Bresser provided
help in the design of the brain phantom. R. Leijsen is thanked for useful
discussions. E. Ercan, T. Ruytenberg and
J. Vonk-van Oosten helped with the data acquisition. References
-
T. Voigt, Imaging
Conductivity using Electric Properties Tomography – Initial Clinical
Results in Glioma Patients, 2011 XXXth URSI General Assembly and
Scientific Symposium, 2011
- K.K Tha et
al, Noninvasive electrical conductivity measurement by MRI: a test of its
validity and the electrical conductivity characteristics of glioma, Eur
Radiol (2018) 28:348–355
- S Gabriel
et al., The dielectric properties of biological tissues: III. Parametric
models for the dielectric spectrum of tissues,1996 Phys. Med. Biol. 41
2271
-
S.
Gavazzi et. al. Transceive
phase mapping using the PLANET method and its application for conductivity
mapping in the brain, Magn Reson Med. 2019;00:1–18.
- E M
Haacke, Extraction of conductivity and permittivity using magnetic
resonance imaging, 1991 Phys. Med. Biol. 36 723
- A.L.H. M. W. van Lier et al.,B1+ Phase Mapping at
7T and its Application for In Vivo Electrical Conductivity Mapping, Magn
Reson Med. 67:552–561 (2012)
- E.
Balidemaj et al, Feasibility of Electric Property Tomography of Pelvic
Tumors at 3T, Magnetic Resonance in Medicine 73:1505–1513, 2015 (2014)
-
A.L.H. M.
W. van Lier et al., Electrical Properties
Tomography in the Human Brain at 1.5, 3, and 7T: A Comparison Study, Magnetic
Resonance in Medicine 71:354–363 (2014)