Lucas Soustelle1,2, Samira Mchinda1,2, Thomas Troalen3, Maxime Guye1,2, Jean-Philippe Ranjeva1,2, Gopal Varma4, David C Alsop4, Guillaume Duhamel1,2, and Olivier M Girard1,2
1Aix-Marseille Univ, CNRS, CRMBM, Marseille, France, 2APHM, Hôpital Universitaire Timone, CEMEREM, Marseille, France, 3Siemens Healthcare SAS, Saint-Denis, France, 4Division of MR Research, Radiology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA, United States
Synopsis
Inhomogeneous Magnetization Transfer (ihMT) is a novel myelin
imaging technique. As for many other MRI techniques special attention must be
paid to local B1+ variations in applications at high field. In this work, a
detailed analysis of the ihMT ratio sensitivity to B1+ inhomogeneities is
performed in the context of human brain imaging at 3T. It is shown that
concentrating RF power is an efficient way to reduce B1+ induced bias and that
the overall performance of the technique lies on a tradeoff between
high ihMT ratios (sensitivity) and reduced sensitivity to B1+ inhomogeneities.
Introduction
Inhomogeneous Magnetization Transfer (ihMT) imaging is a recent technique
that has demonstrated good sensitivity to myelination1,2. When performed at relatively low field (e.g. 1.5T) or in small
field of view (e.g. in small animal applications) a rather homogeneous B1+
environment is typically observed. However, special
attention must be paid to the local B1+ variations when applications at high
field or on large samples are targeted, to prevent shadowing artifacts inherent
to B1+ inhomogeneities.
Mchinda et al. have evidenced different B1+
dependency regimes of the ihMT ratio metric
(ihMTR) as a function of the RF
energy concentration of MT saturation pulses using a sensitivity-boosted ihMT gradient echo (ihMT-GRE) sequence at
1.5T3. It was shown that concentrating RF power (i.e. use of long TR) may
yield lower relative variations of ihMTR with B1+ variations (illustrated on
Figure 1 with simulated ihMTR data), hence holding promise for high field
applications. These findings have been confirmed experimentally in a recent preliminary study performed
on brain at 3T4.
In this work, we expand on these results and present
a thorough analysis of the ihMTR metric sensitivity to local B1+ values in
order to provide hints for the optimization of ihMT-GRE sequences at 3T for
human brain applications.Materials and methods
Experiments were performed on a 3T clinical scanner (Verio, Siemens Healthineers,
Erlangen, Germany) with body coil transmission and a 32-channel receive head
coil on healthy volunteers. Every protocol included a 3D anatomical sequence
(MPRAGE) as well as a B1+ mapping sequence (pre-saturated turbo FLASH). Two
sets of 3D ihMT-GRE sequences (referred hereafter to as Configuration #1 and #2
with parameters listed in Table 1) were acquired on three volunteers each, with
a constant nominal B1,RMS of 2.7 µT. Both configurations mostly differ
in their number of RF pulses per saturation burst3, with Configuration #2 allowing for stronger RF energy concentration
(achieved with longer TR) considering the peak power limitations of the system.
Each ihMT experiment was performed twice: 1) with a nominal B1+ adjustment, and
2) with a B1+ distribution deliberately attenuated by 20%, achieved by setting
the transmitter reference voltage to 80% of its nominal value (respectively
referred hereafter to as Vref and Vref,80%).
Analyses were performed on the whole white matter area segmented from
the MPRAGE using FreeSurfer5, and projected over ihMTR and resampled B1+ maps. All extracted
voxels were pooled across the three subjects for each generated map. Resulting
distributions were used for the computation of mean and standard deviation of
ihMTR from the Vref datasets. Composite maps of voxel-wise ihMTR
relative variation between Vref and Vref,80% (δihMTR) were used to
compute the following statistics: i) mean and standard deviation of δihMTR;
ii) voxel fraction with a relative variation within a 10% error margin: δihMTR-[-10%;10%]; iii) root-mean-square error with respect to the ideal case of
a 0% ihMTR relative variation (perfect immunity to B1+ variations): δihMTR-RMS. Insight into the B1+
spatial distribution was further provided by investigating the same metrics in
clusters of WM voxels with specific B1+ values ranging from 0.86 to 1.16 (1.00
referring to the nominal B1+ value), with a class size of 0.02.Results
Representative images and metrics
pertaining to the whole WM area are reported in Figure 2. Overall the highest
average ihMTR were obtained in both configurations for an intermediate TR of
145 ms. Regarding immunity to B1+ variations the best scores (minimal bias: lowest
δihMTR and δihMTR-RMS, highest δihMTR-[-10%;10%]) were
unambiguously obtained at TR=265 ms for Configuration #1. For Configuration #2,
the best settings lie in the range 265-ms ≤ TR ≤ 345-ms depending
on the considered figure of merit. Of interest, the B1+-clustered data analysis,
shown in Figure 3 for the absolute ihMTR measured at Vref and on
Figure 4 for δihMTR, is generally consistent with the
simulation behaviours from Figure 1: short TRs leading to quasi-linear increase
of ihMTR with B1+ in the tested range, and associated with a high B1+-induced
bias, and long TRs leading to a bell-shaped dependency of ihMTR with B1+,
associated with minimal B1+-induced bias.Discussion and conclusion
Sensitivity
of ihMTR with B1+ variations has been investigated experimentally and agrees
well with theoretical expectations. From the simulations presented in Figure 1,
it can indeed be qualitatively understood that RF energy concentration should
reduce the sensitivity to B1+. Noteworthy the bell-shaped curve obtained in the
more concentrated case explains that δihMTR may change sign (getting negative) in areas of
strong B1+, consistent with the cluster analysis displayed in Figure 4
(Configuration #2).
This study indicates that it is not possible to cancel the B1+ induced bias in all voxels simultaneously with the
proposed approach, suggesting that the fraction of voxels within acceptable
error margins or the mean root-mean-square error should be better indicators of the overall performance
of ihMTR than the more usual mean ihMTR relative variation. Noteworthy, the
settings the most robust to B1+ variations do not provide the highest average
ihMTR values. A tradeoff is therefore to be made between high ihMTR and reduced
sensitivity to B1+ inhomogeneities.Acknowledgements
This
work was supported by the SATT Sud-Est (France), the French Association pour la
Recherche sur la Sclérose En Plaques (ARSEP), Roche Research Foundation
(Switzerland) and French National Research Agency, ANR [ANR‐17‐CE18‐0030].
This work was performed by a
laboratory member of France Life Imaging network (grant ANR-11-INBS-0006).References
1. Van Obberghen
E, Mchinda S, le Troter A, et al. Evaluation of the Sensitivity of
Inhomogeneous Magnetization Transfer (ihMT) MRI for Multiple Sclerosis. Am. J.
Neuroradiol. 2018;39:634–641 doi: 10.3174/ajnr.A5563 SMASH.
2. Duhamel G,
Prevost VH, Cayre M, et al. Validating the sensitivity of inhomogeneous
magnetization transfer (ihMT) MRI to myelin with fluorescence microscopy. Neuroimage
2019;199:289–303 doi: 10.1016/j.neuroimage.2019.05.061 SMASH.
3.
Mchinda S, Varma G, Prevost VH, et al. Whole brain inhomogeneous
magnetization transfer (ihMT) imaging: Sensitivity enhancement within a
steady-state gradient echo sequence. Magn. Reson. Med. 2018;79:2607–2619 doi:
10.1002/mrm.26907 SMASH.
4. Mchinda S,
Varma G, Draveny R, et al. Whole brain inhomogeneous Magnetization Transfer
(ihMT) imaging at 3T: concentrating RF power to mitigate RF inhomogeneities
effects. In: Proceedings 26th Scientific Meeting, International Society for
Magnetic Resonance in Medicine. ; 2018. p. 786.
5. Dale AM,
Fischl B, Sereno MI. Cortical Surface-Based Analysis. Neuroimage 1999;9:179–194
doi: 10.1006/nimg.1998.0395 SMASH.