Quan Tao1, Peiwei Yi1, Zimeng Cai1, Yingjie Mei2, Zhifeng Chen1, Ruiyuan Liu1, Wufan Chen1, and Yanqiu Feng1
1School of Biomedical Engineering, Southern Medical University, Guangzhou, China, 2Philips healthcare, Guangzhou, China
Synopsis
Chemical exchange saturation
transfer (CEST) as a novel molecule MRI technique, was approved to detect some
diseases, like tumor grading, kidney injure,
osteoarthritis (OA) and etc. However, it is crucial to purify the CEST signal. In
this study, we explored the perfusion contribution to the CEST signal acquired
by a UTE-CEST sequence.
Synopsis
Chemical exchange saturation
transfer (CEST) as a novel molecule MRI technique, was approved to detect some
diseases, like tumor grading[1], kidney injure[2],
osteoarthritis (OA) and etc. However, it is crucial to purify the CEST signal. In
this study, we explored the perfusion contribution to the CEST signal acquired
by a UTE-CEST sequence.Introduction
The
quantification and separation of the CEST signal accurately is important for developing
imaging metrics for various diseases diagnosis and grading, however, the ASL
technique was quite similar to the CEST method in the pulse sequence and image
post-processing. Both methods in ASL and CEST applied an pre-pulse to label and
saturation, and in the image processing, the difference between the labeled
(saturated) and unlabeled (unsaturated) image was
applied to obtain perfusion and labile protons information for ASL and CEST
respectively. Thus, the CEST signal may contain perfusion signal, especially in
the region where the blood supply is abundant. Although Feliks Kogan’s result
have showed that the perfusion has no effect on the in vivo CEST effect from Cr
(CrCEST) in skeletal muscles[3], however muscles have slow blood
flow. Here, we conduct the experiments in vitro and in rat kidney,
demonstrating the perfusion effect maybe pollute the CEST signal.Method
In vitro
experiments: an rubber tubing contains ioversol solution (120mM) connected
with peristaltic pump was to simulate the blood flowing (Figure 1.). Phantoms
were scanned on a 7T small-bore MRI scanner (Bruker Biospec, Billerica, MA)
equipped with a 40 mm volume coil. CEST imaging is under the flowing rate of 3
and 0 cm/s, and read out by ultrashort echo time (UTE). The
resolution of CEST image is 0.25 mm2, TR=40 ms, TE=0.4 ms,
saturation power=0.9 uT, saturation time=30 ms, the total saturation time for
each offset is 12 s, and z-spectrum is acquired form -8 ppm to +8 ppm, the
offset frequency is 0.2 ppm.
In vivo experiments:
rats
were scanned on a 7T small-bore MRI scanner (Bruker Biospec, Billerica, MA)
equipped with a 72mm volume coil and a rat heart array RF coil for signal
transmission and reception respectively. In vivo imaging also used a UTE
sequence for CEST signal readout. The resolution of CEST image is 0.4 mm2,
TR=40 ms, TE=0.4 ms, imaging frequency is ±3.5 ppm, saturation power=0.9 uT,
saturation time=30 ms, number of average=6. The resolution of ASL is the same
as CEST, and TR=10000 ms, TI= 30, 100, 200, 300, 500, 800, 1200, 2000, 3000,
5000 ms.
Data processing The
CEST signal was subtracted by MTR(±3.5ppm)=(M0-M±3.5ppm)/M0.Results
Figure
2 shown the CEST image of the phantom with the ioversol solution flowing with 3
and 0 cm/s. In figure 2(a), the MTR image on 4.3 ppm been shown, and (b), (c)
show the z-spectrum of the center and exterior static solution. And the
corresponding MTR image and z-spectrum of flowing solution also been shown in
figure 2(d), (e), (f). In flowing solution, the MTR value improved drastically
in the center, and smaller improved in the peripheral. Figure 3 shown the T2w,
APT, and ASL image of the rats’ kidney. The APT signal of cortex is outclass
that of medulla and pelvis, and highly consistent with ASL signal.Discussion and Conclusion
In
this study, we quantify the CEST signal of the ioversol under the static and
flowing situation, and found the CEST signal improvement if the ioversol
flowing, which proved our assumption, the ASL signal will be mixing in CEST
signal. And kidney experiment also be implemented, the CEST signal of cortex
was much higher, and highly agreed with ASL signal. However, it not clear that
if the perfusion effect alter with different CEST sequence, further detailed
studies are still needed to confirm this assumption and purify the CEST signal.Acknowledgements
This work was supported by the Guangdong Provincal Medical Foundation
(A2018268).References
[1] Zhou J,
Heo H Y, Knutsson L, et al. APT‐weighted MRI: Techniques, current neuro
applications, and challenging issues[J]. Journal of Magnetic Resonance Imaging,
2019.
[2] Liu J,
Han Z, Chen G, et al. CEST MRI of sepsis‐induced acute kidney injury[J]. NMR in
Biomedicine, 2018, 31(8): e3942.
[3] Kogan
F, Stafford R B, Englund E K, et al. Perfusion has no effect on the in vivo
CEST effect from Cr (CrCEST) in skeletal muscle[J]. NMR in Biomedicine, 2017,
30(1): e3673.