Eugene Milshteyn1,2, Bastien Guerin1,2, and Lawrence L. Wald1,2,3
1Athinoula A. Martinos Center for Biomedical Imaging, Charlestown, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Harvard-MIT Division of Health Sciences and Technology, Cambridge, MA, United States
Synopsis
The
current standard methods for magnetic resonance imaging of superparamagnetic
iron oxide nanoparticles (SPIONs) can be subject to “1/f” physiological
fluctuations that can confound pre/post comparisons. We present a method to
combine Block Design contrast Modulations and statistical analysis with the
bSSFP sequence (BDM-bSSFP) in order to mitigate this effect. To do this, we
leverage the frequency response of the bSSFP sequence to the local fields
created by SPIONs. We show the ability of this method to detect SPIONs down to 10
μg/mL through simulations and in-vitro
acquisitions.
Introduction
The
current standard for magnetic resonance imaging (MRI) of superparamagnetic iron
oxide nanoparticles (SPIONs) involves pre- and post-injection comparisons of
T2- or T2*-weighted acquisitions acquired several minutes apart. However, this
approach can be confounded by various “1/f” physiological fluctuations that
occur between the pre- and post-injection images,
induced by subject
and respiratory motion.1,2 Others have developed
entirely post-injection methods in order to generate positive contrast images
from SPIONS,3-5 including using the
balanced steady-state free precession (bSSFP) sequence.6-9 Recently, Zhu et al.10 demonstrated a method
to externally modulate and selectively image the SPION contrast which allowed analysis of
a “block design” acquisition similar to an fMRI analysis. The resulting
statistical analysis potentially avoids the 1/f noise of long time-scale comparisons.
In this study we achieve
the advantages of block design modulation analysis by periodically shifting the
unique frequency response of the Block Design Modulated bSSFP sequence (BDM-bSSFP). This selectively modulates the water signal near the SPIONs. We
demonstrate the utility of this approach through both simulations and in-vitro experiments, and show detection
of SPIONs down to 10μg/mL.Methods
Theory: The frequency shift $$$Δf(r)$$$ of a sphere of SPIONs can
be approximated by:6
$$∆f(r)= \frac{γ\cdot∆χ}{6π}(\frac{R}{r})^3(3\cos^{2}\theta-1)B_{0}$$
Figure
1A shows the B0 map for a r=5.3mm sphere and ∆χ=1ppm. Figure 1B shows the response of a bSSFP
sequence to such a perturbation given a repetition time (TR) of 4.3ms and flip
angle (FA)=5°. The result is the well-known relatively flat
passband and a rapidly varying transition band. If maximal signal change is
sought, water near SPIONs will experience local fields that would move the
resonance into the transition band of the bSSFP frequency response (blue circle).
Conversely, water far from the SPIONs will lie in the passband (red star) and
experience little effect from shifting the sequence’s response. Thus, by modulating
the bSSFP frequency response in a block design manner (by modulating the
scanner Tx and Rx center frequency), we anticipate significant signal
modulations in the water near SPIONs without significant modulation of the
water signal from water far from the SPIONs (Figure 1C). These “block-design”
modulations can be robustly detected through statistical analysis.
Simulations: We initially simulated
the field distribution of an r=5.3mm sphere for four different TRs (3.4, 4.3,
6, and 10ms) with an isotropic resolution of 0.8mm and ∆χ=1ppm. First, we simulated the bSSFP signal for
each voxel with FA=5°, and T1/T2=300/75ms. Then we took 20 timepoints
at the center frequency (0Hz) followed by 20 timepoints at X Hz off-center and
repeated this 10 times to generate 400 timepoints. X ranged from 5-50Hz, and
gaussian noise was added to simulate experimental data. The correlation coefficient (R) at each voxel was calculated by detrending the
signal, and then calculating R assuming a typical “on-off” response. Voxels with
abs(R)>0.8 were considered “Activated”. Subsequently,
we calculated R for ∆χ=0.25-40ppm for TR=4.3ms and 30 Hz shift off-center. We calculated percent activation for each ∆χ via the following: (# “Activated” voxels)/(# image voxels - # sphere voxels). For comparison, we also calculated the
difference between the two frequencies (0 and X) using the mean of each
frequency across all 200 timepoints. In this analysis, "Activated" voxels were
those with difference>0.65*max(difference). All analysis was performed in MATLAB.
Experiments: Six
concentrations (0, 5, 10, 20, 50, and 100µg/mL) of 10nm SPIONs (Ocean
Nanotech) were put into r=5.3mm microcentrifuge tubes and placed into an agar
phantom (Figure 4A). The phantom was placed into a Siemens 32-channel head coil
with the microcentrifuge tubes orthogonal to B0 and all experiments
were performed on a 3T Siemens Prisma scanner. We used the product bSSFP sequence
that features an option to modulate (shift) the scanner base-frequency and thus
the sequence frequency response by a specified value. For the four TRs listed
above, we acquired a single 5mm sagittal slice at two different frequency
offsets (0 and X Hz; 200 times at each frequency offset) with FA=5°. Then we reordered the data
into blocks of 20 timepoints at 0 Hz and 20 timepoints at X Hz, for a total of
400 timepoints (10 on-off blocks) and calculated R.Results
Figure
2 shows the R maps, a visualization of activated voxels, and representative
signal blocks from a voxel far away and close to the SPIONs for different TRs
and different frequency modulations. All four cases presented here seem to
indicate robust activation near the SPIONs. Unsurprisingly, Figure 3 shows that
the activation strength increases with increasing magnetic susceptibility
(i.e., SPIONs concentration). Additionally, the correlation coefficient
analysis provided more robust detection of the SPIONs location than a simple
difference image analysis. Figure 4 shows the experimental results for TR=4.3ms and 30Hz modulation,
with detection of 50 and 100µg/mL tubes clearly visible. Furthermore, the correlation
coefficient analysis (C/D) yielded a greater activation volume than the difference
analysis (F/G). Figure 5 shows similar results for the other acquired TRs, which
is in agreement with simulations. Conclusion
We
present a new approach to positive contrast imaging of SPIONs using a block
design modulation of a low flip angle bSSFP acquisition. Future analysis will focus on optimizing the combination of flip angle, TR, and frequency modulation that will provide the most sensitivity to SPION detection.Acknowledgements
NIH
F32EB027571.References
1.
Weisskoff RM, Baker J, Belliveau J, et al. Power Spectrum Analysis of
Functionally-Weighted MR Data: What’s in the Noise? In: Proc. Intl. Soc. Mag.
Reson. Med. 1 (1993) 0007.
2. Aguirre GK, Zarahn E, D’Esposito M. Empirical Analyses of
BOLD fMRI Statistics. Neuroimage 1997;5:199–212.
3. Cunningham CH, Arai T, Yang PC, Mcconnell M V, Pauly JM,
Conolly SM. Positive Contrast Magnetic Resonance Imaging of Cells Labeled with
Magnetic Nanoparticles. Magn. Reson. Med. 2005;53:999–1005.
4. Zhao Q, Langley J, Lee S, Liu W. Positive contrast
technique for the detection and quantification of superparamagnetic iron oxide
nanoparticles in MRI. NMR Biomed. 2011;24:464–472.
5. Harris SS, Mao H, Hu XP. Adiabatic Pulse Preparation for
Imaging Iron Oxide Nanoparticles. Magn. Reson. Med. 2012;67:1133–1137.
6. Çukur T, Yamada M, Overall WR, Yang P, Nishimura DG.
Positive Contrast with Alternating Repetition Time SSFP (PARTS): A Fast Imaging
Technique for SPIO-Labeled Cells. Magn. Reson. Med. 2010;63:427–437.
7. Ingle RR, Çukur T, Nishimura DG. The Central Signal
Singularity Phenomenon in Balanced SSFP and Its Application to
Positive-Contrast Imaging. Magn. Reson. Med. 2012;67:1673–1683.
8. Mascheri N, Dharmakumar R, Zhang Z, Paunesku T. Fast
low-angle positive contrast steady-state free precession imaging of
USPIO-labeled macrophages: theory and in vitro experiment. Magn. Reson. Imaging
2009;27:961–969.
9. Heyn C, Bowen C V, Rutt BK, Foster PJ. Detection Threshold
of Single SPIO-Labeled Cells With FIESTA. Magn. Reson. Med. 2005;53:312–320.
10. Zhu B, Witzel T, Jiang S, Huang SY, Rosen BR, Wald LL.
Selective magnetic resonance imaging of magnetic nanoparticles by acoustically
induced rotary saturation. Magn. Reson. Med. 2016;75:97–106.