Joshua D Kaggie1, Mary McLean1, Rolf F Schulte2, Dimitri A Kessler1, Frances Henson3, Fiona J Gilbert1, Martin J Graves1, and Ferdia A Gallagher1
1Radiology, University of Cambridge, Cambridge, United Kingdom, 2GE Healthcare, Munich, Germany, 3Veterinary Medicine and Divisino of Trauma and Orthopaedic Surgery, University of Cambridge, Cambridge, United Kingdom
Synopsis
We demonstrate
natural abundance deuterium imaging of the normal human brain in vivo, without the requirement of an
injected deuterated molecule, and within a 10-minute chemical shift imaging acquisition.
Introduction
Deuterium
(2H) has a 0.01% natural abundance, which is low in comparison to
that of proton (1H), but comparable to the concentrations of other
ions, such as intracellular 23Na. The resulting lower sensitivity of
2H-MRI compared to 1H-MRI has resulted in a paucity of
research in this area. There has been a
resurgence of interest in 2H-MRI with the advent of deuterium
metabolic imaging (DMI) and the imaging of 2H-labelled glucose. 2H-labelled
molecules that measure metabolic activity are particularly interesting for
measuring the Warburg effect in brain tumours. Recent work has shown MRI of 2H-labelled
water and [6,6′-2H2]glucose at concentrations as high as
5%. 2H-labelled glucose has been shown to metabolise into lactate
and glutamine/glutamate in the normal human brain and in brain tumours. 2H-MRI
with [6,6′-2H2]glucose has the potential to rival FDG-PET
imaging without the need for ionising radiation
(1).
One of the
challenges of DMI is that it is difficult to optimise and standardise 2H-MRI in the absence of endogenous signal. Here we demonstrate natural abundance 2H-MRI, which can be used
for optimization and quality assurance of hardware, pulse sequence, and
reconstruction methods. This work will also aid the acquisition of signal
following the administration of 2H-labelled molecules.Methods
We modified a 16-rung
birdcage coil to resonate at 19.8 MHz for 2H-MRI (Figure 1) for a 3T
MRI system (MR750, GE Healthcare, Waukesha, WI). The birdcage connected to a quadrature
2H transmit/receive switch (Clinical MR Solutions, Brookfield, WI).
The 28 cm diameter birdcage included copper conductor rungs with 8 cm length, 1
cm width, and with 2 cm end-ring widths.
We imaged natural
abundance 2H in the brains of two cadaveric sheep and in three normal
volunteers with 3D chemical shift imaging (CSI). Volunteer imaging occurred
with informed consent and local ethical approval. The 3D density-weighted CSI sequence acquired
either 2 averages of a 12x12x12 matrix size, 2405 FID acquisitions, scan time
9.5 minutes; or 1 average of a 16x16x16 matrix size, 5408 transients, scan time
10.8 minutes. Both used field-of-view = 240 mm3, TR = 120 ms, flip
angle = 45°, bandwidth = ±2.5 kHz, with 512 spectral points between -128 and +128
ppm per voxel, and both were zero-filled once in the time domain and once
spatially in the X and Y planes for analysis. For viewing images based on the
maximal spectral intensity, an additional 4x4 interpolation was applied with
3-point spline interpolation using SciPy (2).
3D gradient echo T1-weighted
images were also obtained with the 1H body coil without
repositioning of the subjects.
Results
Figure 2 shows
sample spectra from a 5x5x2 cm3 region in the center of the brain. The
spectra are dominated by a central heavy water peak, and several spectra show a
possible fat peak at around -4ppm relative to water.
Figure 3 shows the
maximum signal in two axial slices of a normal volunteer. The cerebral
ventricles demonstrate hyperintensity.
Figure 4 shows a
sagittal view of the brain with the voxel-wise mean of the CSI spectra,
demonstrating good delineation of the brain in 3D.
Figure 5 shows the
MRF-derived 1H density and relaxation maps, which should relate to
the endogenous images obtained with 2H.Discussion
Despite
the low natural abundance of 2H, we demonstrated that endogenous 2H
signal of normal brains can be imaged in
vivo using MRI.
This
work demonstrates that the evaluation of 2H methods can be performed
without the need for increased signal following the administration of 2H-labelled
molecules. Natural abundance imaging simplifies the development and
optimization of new methods, such as the evaluation of more sensitive
radiofrequency coils.
Baseline
measurements with natural abundance imaging also establish the feasibility of
performing dynamic metabolic measurements. Administration of 2H-labelled molecules would significantly
increase the signal-to-noise above this baseline, e.g., using [6,6′-2H2]glucose.
Resolution of isotopically-labelled metabolite peaks from the nearby naturally
abundant heavy water signal will require more careful optimization of field
homogeneity than that achieved in the current study using only linear shim
adjustment.
Conclusion
We have demonstrated natural abundance deuterium imaging in vivo in normal human brains,
which can be useful for methodological evaluation of pulse sequences and radiofrequency
coils, and for baseline measurements before dynamic imaging.
Acknowledgements
This work acknowledges
support from the National Institute of Health Research Cambridge Biomedical
Research Centre, Addenbrooke's Charitable Trust, Cancer Research UK, and from the European Union's Horizon 2020
research and innovation programme under grant agreement no. 761214.References
1. De Feyter HM,
Behar KL, Corbin ZA, Fulbright RK, Brown PB, McIntyre S, Nixon TW, Rothman DL,
de Graaf RA. Deuterium metabolic imaging (DMI) for MRI-based 3D mapping of
metabolism in vivo. Science advances 2018;4(8):eaat7314.
2. Virtanen
P, Gommers R, Oliphant TE, Haberland M, Reddy T, Cournapeau D, Burovski E,
Peterson P, Weckesser W, Bright J. SciPy 1.0--Fundamental Algorithms for
Scientific Computing in Python. arXiv preprint arXiv:190710121 2019.