Jullie W Pan1, Victor Yushmanov1, Chan H Moon1, Frank Lieberman2, and Hoby P Hetherington1
1Radiology, University of Pittsburgh, Pittsburgh, PA, United States, 2Neurology, UPMC, Pittsburgh, PA, United States
Synopsis
The
increased SNR at 7T and high efficiency of spatial encoding from fast readout trajectories
means that substantially accelerated spectroscopic imaging acquisitions are practical.
In this report, we implement targeted multi-slice Hadamard encoded cascaded and
simultaneous acquisitions. With the cascaded rosette, minimal CSDE is incurred,
generating 4slices in 8min12s. The J-refocused sequence achieves excellent CRLB
in <2.3min for a single slice. With a B1 reduction strategy that allows
multiple simultaneous Hadamard encoding, the J-refocused sequence is also demonstrated
with 4slice acquisitions. As performed in brain tumor patients, the targeted Hadamard
rosette is well tolerated and sensitive for detection of pathology.
Introduction
For extended volume spectroscopic imaging in the human brain, an
acceptable clinically relevant goal is to acquire single and multiple axial
slices at moderate to high resolution in-plane in under 10min. To achieve this
goal, fast in-plane encoding and clean axial slice profiles (thereby minimizing
partial volume overlap) are needed. The greater than 2.3-fold increase in SNR
at 7T relative to 3T makes fast trajectories especially advantageous. We
describe implementation of the rosette trajectory (1) with multi-slice
capabilities at 7T for moderate spin echo and j-refocused sequences.
Methods
For multi-slice MRSI acquisitions at 7T,
the slice encoding can be achieved
either by conventional phase encoding, or by use of cascaded or simultaneous RF
based Hadamard encoding. At low slice counts (e.g., 2 to 4), conventional phase
encoding suffer from poor slice profiles and substantial slice profile bleed
which is eliminated by Hadamard encoding (2). The RF-based Hadamard cascaded
multi-slice acquisition maintains optimal chemical shift dispersion error
(CSDE), but each slice has a slightly different TE. Due to B1
limitations, for the simultaneous Hadamard acquisition, the base pulse duration
was doubled, and an encoding pattern was used to further reduce peak B1
requirements and reduce CSDE (Fig. 1). We demonstrate the cascaded and
simultaneous multi-slice Hadamard rosette for moderate spin echo and j-refocused
polarization transfer sequences. For the rosette trajectory, multiple shots of
the circular rosette (1) are used with radial and angular oscillation
frequencies of 625Hz with two temporal interleaves (spectral bandwidth of
2500Hz, Gmax 5.5mT/m; Smax 40.2mT/m/ms). The data were reconstructed as
previously described (3) with matched scout images used to weight and phase the
data for coil recombination.
All studies were performed with a Siemens Magnetom Step 2.3 7T 8 channel system
equipped with an Avanto body gradient coil, maximum slew rate of 200mT/m/ms,
Gmax of 40mT/m and an 8x2 transceiver array (2 rows, 8coils/row, RRI Inc.). RF
shimming is performed using two spatial B1+
distributions, targeting the whole brain (“homogeneous”) and extracerebral
regions (“ring”) with the ring B1+ used for outer volume
suppression (4). Non-iterative B0 shimming is performed using a very
high order shim insert (VHOS, RRI Inc) providing 1st-4th order shims and two
5th order shims. We demonstrate this sequence in healthy controls and brain
tumor patients.Results
Fig. 2 compares data from a 4 slice cascaded moderate TE spin
echo Hadamard acquisition, time matched (14.5min) with a single slice
conventional phase encoded spectroscopic image (CSI). There is high consistency
between these acquisitions, with a voxel-to-voxel analysis of Cr and NAA
amplitudes showing a very significant correlation, R2=0.92. Spectral
quality as assessed by the CRLBs for tNAA was not significantly different at
2.2±0.8 and 1.8±0.8 for the CSI and RSI respectively (2 subjects, n=140 voxels
total). The J-refocused sequence is a phase sensitive double spin echo homonuclear
polarization transfer acquisition. Fig. 3 shows that spectra from a 9mm thick
2.2min single slice J-refocused RSI acquisition have spectral quality that is
comparable to a 9mm thick 14.5min phase encoded CSI acquisition of equivalent
volume resolution. Despite the significant difference in acquisition time, LCM analysis
shows no significant difference between the CRLB for glutamate, 4.8±1.7% (rosette)
and 5.8±2.1% (CSI; n=2 subjects, ~130 pixels). Analyses of these data showed
significant gray matter regression for the Glu/tNAA parameter for all subjects,
intercept 0.49±0.15 and slope 0.58±0.06. To multi-slice the J-refocused
acquisition requires the use of simultaneous (rather than cascaded) multi-slice
excitation RF pulses. Using the reduced B1 Hadamard slice encoding
for a 4-slice J-refocused acquisition, there is a chemical shift dispersion
error in these data of 2.2mm between tNAA and Creatine (Fig. 4). The SNR is
consistent with the 4fold longer acquisition time (8.8min) used for the
multi-slice study in comparison with the matched 2.2min single slice
acquisition.
Fig. 5 shows data from a 4 slice Hadamard encoded moderate spin echo rosette
acquisition in a 69yo tumor patient whose tumor was initially treated in 2009.
Low grade progression was noted in 2017 and again with the present study
identifying increased Ch/NAA at the gyral brain edge. This patient ultimately
proceeded to additional chemotherapy.Discussion
We have applied a rosette trajectory to achieve rapid MRSI acquisitions
at 7T with <1cc resolution in single (2.2min) to four and eight slice
acquisitions (8, 16min respectively). With two temporal interleaves as a 2D
method, the rosette has very low gradient demands (Smax 40.2mT/m/ms, 20% of
maximum to achieve a 9mm nominal isotropic resolution), is identical in
gradient amplitude from shot to shot, and does not require additional eddy
current corrections. We have shown that the SNR and CRLB values from the
rosette are not significantly different from single slice conventional phase
encoded acquisitions of equal duration, consistent with minimal loss due to the
trajectory encoding. As applied in tumor patients, the rosette trajectory is well
accepted by subjects and as shown, is sufficiently sensitive to demonstrate the
pertinent pathology in neuro-oncology patients.Acknowledgements
This work supported by
National Institutes of Health, NINDS R01NS090417,
R01EB024408 and R01EB011639.References
1.
Noll DC IEEE Med Imag 1997 16:372.
2.
Gonen O Mag Res Med 1998 39:34
3.
Schirda C Mag Res Med 79:2470
5.
4. Hetherington HP Mag Res Med 63:9