Hidehiro Watanabe1, Nobuhiro Takaya1, and Fumiyuki Mitsumori1
1Center for Environmental Measurement and Analysis, National Institute for Environmental Studies, Tsukuba, Ibaraki, Japan
Synopsis
The method for absolute quantitation of 1H
MRS in a human brain at high B0 field was proposed. A water phantom
as a concentration reference and a human brain area measured separately. The
ratio of the reception sensitivities between uniform areas in the phantom and
in the human brain can be computed from measured B1+s. The ratio
between the VOI and the uniform area in the human brain can be computed by our
previously reported ratio map method. Then, concentration in the VOI can be calculated.
Our method was demonstrated in the phantom experiments.
Introduction
High field 1H
MR spectroscopy in human brain has advantages in high sensitivity and spectral
resolution. This feature can lead to accurate quantitation. In absolute
quantitation of metabolites of human brains in 1H MRS, tissue water
is widely used as an internal concentration reference [1]. Water content in a
volume of interest (VOI) is estimated from fractions of GM, WM and CSF with the literature values. However, pathological changes of water content will cause
quantitation error.
This problem can be resolved using an external
concentration reference (Fig. 1). However, this method cannot be used at high B0 field due
to B1 inhomogeneity.
One of important findings on B1 that
the transmission RF field and the reception sensitivity are represented as B1+ and B1-, respectively and that these fields differ in spatial distribution at high B0 field even when a transceiver
RF coil is used [2]. While B1+
can be measured, reception sensitivity of B1- cannot be measured in
human brains because it is always associated with signal intensities. Then, reception sensitivities between the VOI in a human
brain and the external reference cannot be compared at high B0 field.
We have reported that the linear
relationship exists between B1+ and B1-
around a uniform area even at high B0 field [3]. Water content maps could be measured in human brains using this relationship at 4.7T. In this work, we will apply this relationship to absolute
quantitation in 1H MRS at high B0 field and will demonstrate
measurement of concentration in experiments using water phantoms.Methods
Proposed
absolute quantitation method
In our proposed method, a water phantom as a reference and a human brain
are measured separately (Fig. 2). The following measurements are required to obtain a
reception sensitivity in the VOI of a human brain; B1+ maps both in
the phantom and in the human brain, and localized water signal in a uniform
area in the phantom. Form the finding that the linear relationship exists between B1+
and B1- around a uniform area [3], the ratio of reception sensitivities of B1-s in uniform areas (red squares in Fig. 2), denoted as Coefuniform can be expressed by the ratio of measured B1+s.
$$\begin{align}Coef_{uniform}&=\frac{B_{1\ \ brain.uniform}^{-}}{B_{1\ \ ref.uniform}^{-}}\\&=\frac{B_{1\ \ brain.uniform}^{+}}{B_{1\ \ ref.uniform}^{+}}\tag{1}\end{align}$$
where subscript texts of brain.uniform and ref.uniform denote the uniform area inside the human brain and that inside the reference phantom, respectively. Next, B1-s can be compared between the VOI (yellow dotted square) and the uniform area (red dotted square) in the human brain using our previously reported ratio map method [4].
$$Coef_{VOI}=\frac{B_{1\ \ brain.VOI}^{-}}{B_{1\ \ brain.uniform}^{-}}.\tag{2}$$
From these equations, reception
sensitivities can be compared between the VOI in the human brain and the
uniform area (red square) in the reference phantom.
$$\begin{align}Coef&=Coef_{VOI}\times
Coef_{uniform}\\&=\frac{B_{1\ \ brain.VOI}^{-}}{B_{1\ \
ref.uniform}^{-}}.\tag{3}\end{align}$$
Then, concentration of a metabolite in the human brain denoted as Concmetab can be computed by
$$Conc_{metab}=Conc_{ref}\times \frac{PA_{brain.VOI.metab}}{Coef\times PA_{ref.uniform}}\tag{4}$$
where Concref and PAref.uniform denote the concentration of water that is 55.5 M and a peak area of water in the uniform area in the phantom, respectively, and PAbrain.VOI.metab denotes PA of a metabolite in the localized 1H spectrum of the human brain.
Experiments
All experiments were performed using a 4.7T
whole-body MR system (INOVA, Agilent) by using a volume TEM coil both for
transmission and reception. To demonstrate the proposed absolute quantitation
method, we performed experiments with a cylinder water phantom with 150-mm
diameter and 170-mm length as a reference (Fig. 3a) and a spherical saline
phantom with 130-mm diameter for measuring (Fig. 3b). Absolute B1+
maps were measured by the phase method where a set of two SE images were
acquired using hyperbolic secant pulses form both excitation and refocusing [5].
The same conditions of RF power were used both in the phantom and in the human
brain. Instead of the ratio map in a human brain, a reception sensitivity map in the saline phantom for measuring was measured by an
adiabatic SE sequence [6] to avoid B1+
inhomogeneity with condition of
TR/TE = 4000/26 ms. Localized water signals both in the uniform area (red square in Fig. 3a) in the
reference phantom and in the VOI (yellow dotted square) in the saline phantom were measured by the STEAM
sequence with TR/TE = 15,000/4 ms. The VOI was set up around the edge of the
phantom of non-uniform area (Fig. 3b). The voxel volumes were 8ml in the both phantoms.Results & Discussion
Amplitudes of B1+ in the uniform areas are
0.234 [KHz] in the reference phantom and 0.123 [KHz] in the saline phantom for
measuring, respectively. The ratio of the VOI to the uniform area calculated
from measured B1- map in the saline phantom was 0.776 (Fig. 3). The peak areas of the water spectra in arbitrary units were
141.8 in the reference phantom and 56.5 in the saline
phantom, respectively.
Then, a concentration of water
in the saline phantom for measuring was calculated as 54.2 [M] using Eq. 4
whose error was -2.3% to 55.5 [M].Conclusions
Our proposed method is
useful for absolute quantitation in 1H MRS at high B0 field. This can also be applied in a human brain.Acknowledgements
No acknowledgement found.References
- Barker PB., Soher
BJ., Blackband SJ., Chatham JC., Mathews VP., Bryan RN., Quantitation of proton
NMR spectra of the human brain using tissue water as an internal concentration
reference. NMR Biomed. 1993;6:89-94.
- Hoult DI., The principle
of reciprocity in signal strength calculations – a mathematical guide. Concepts
Magn. Reson. 2000;12:173-187.
- Watanabe H.,
Takaya N., Mitsumori F., Development of quantitative water content mapping in
human brain at high magnetic field. Proc. Intl. Soc. Mag. Reson. Med. 2019;27:4406.
- Watanabe H.,
Takaya N., Mitsumori F. Non-uniformity
correction
of human
brain
at high field
by RF field mapping
of B1+ and B1-. J. Magn. Reson. 2011;212(2):426-430.
- Lee Y., Han Y.,
Park H.W., Watanabe H., Garwood M., Park J.Y. New phase-based B1
mapping method using two-dimensional spin-echo imaging with hyperbolic secant
pulses. Magn. Reson. Med. 2015;73(1):170-181.
- Watanabe H. Experimental demonstration of the proportionality of the RF reception field to a complex conjugate of B1-. Magn. Reson. Med. Sci. 2012;11(3):193-196.