Chloé Najac1, Vincent Boer2, Hermien E. Kan1, Andrew G. Webb1, and Itamar Ronen1
1C.J. Gorter Center for High Field MRI, Department of Radiology, Leiden University Medical Center, Leiden, Netherlands, 2Danish Research Centre for Magnetic Resonance, Copenhagen University Hospital Hvidovre, Copenhagen, Denmark
Synopsis
Spin-echo sequences suffer
from loss of signal due to J-modulation
for coupled spins. Previously, it was shown that adding a π/2 pulse between two π pulses in a double spin-echo
sequence partially refocuses some of the J-evolution.
However, the potential of such sequence at ultrahigh field is limited due to
the large chemical shift displacement error. Here, we propose a J-refocused variant of the sLASER
sequence (J-sLASER) to improve
quantification of J-coupled
metabolites at ultrahigh field. Significant improvement in quantitation of J-coupled metabolites is illustrated
using simulation, phantom and in vivo measurements in the human brain.
Introduction
Spin-echo
sequences suffer from loss of signal due to J-modulation
for coupled spins in metabolites of interest such as glutamate, lactate,
glutamine and others. Although shorter TE values improve signal-to-noise ratios
(SNR) and limits loss due to J-modulation,
it also results in significant confounding contributions from macromolecules
(MM). It was shown that an additional π/2 RF pulse between two π pulses
in a double spin-echo sequence refocuses some of the dephasing caused by J-evolution1,2. This has been
successfully implemented in the PRESS sequence, reducing much of the signal loss due to
homonuclear J-coupling3-6.
At ultrahigh field, poor localization accuracy due to large chemical shift
displacement error (CSDE) in the PRESS sequence remains problematic7.
Additionally, the increase in linewidth, typically about 10Hz and above, without
concomitant increase in J-splitting
results in a greater degree of signal cancellation due to anti-phase peaks
cancellation. Here, we propose a J-refocused
variant of the semi-LASER (sLASER) sequence, which provides both accurate
localization and significant recovery of signal from J-coupled systems. The sequence (J-sLASER) was first validated using simulation followed by phantom
acquisition at 7T. Finally, the sequence was compared to sLASER
sequence in the human brain, illustrating significant improvement in J-coupled metabolites quantification using J-sLASER. Materials and methods
J-sLASER sequence: The classical sLASER
sequence was modified to incorporate a slice-selective π/2 RF pulse and concomitant slice selection and refocusing gradients at
time of formation of the first echo (Figure 1). The phase of the added π/2
pulse was set to be orthogonal to the first π/2 pulse. Phase cycling was used
to suppress unwanted signal due to potential out-of-volume excitation, pulse
angle deviations or T1 relaxation in the first half of the sequence.
All experiments were
conducted on a Philips 7T whole body MRI scanner (Philips Healthcare, The
Netherlands) equipped with a volume transmit/32-channel receive head coil (Nova
Medical, USA).
Simulation: Resonances of protons
belonging to metabolites detectable in the human brain were simulated using
Matlab programs based on the density matrix formalism and published chemical
shifts and J-coupling values8.
The simulation did not include T2 relaxation. Signal was integrated
using Matlab and signal J-modulations
as a function of TE (up to 500ms) for both J-sLASER
and sLASER sequences were investigated.
Phantom validation: In vitro sequence validation was subsequently performed on a GE-MR
spectroscopy “Braino” phantom using both sequences and a range of TE values
from 44 to 164ms and with TR=4000ms, number of signal averaged (NSA)=8, bandwidth=3kHz,
2048 data points.
Spectra were corrected for eddy currents using unsuppressed water spectra. Metabolites signal was quantified using Matlab and LCModel9. A basis-set for each TE
value was created for both sequences.
In vivo measurement: In vivo
experiments were performed on 3 healthy volunteers. A 3D-T1W
gradient-echo acquisition (TR/TE=5/2ms, resolution 1x1x1mm) was used for
planning of the experiments. A 8mL volume-of-interest was positioned
in the posterior cingulate cortex (PCC, Figure
4). Metabolites and water spectra were acquired using both sLASER and J-sLASER sequence at TE=45 and 65ms (TR=5600ms/NSA=80/BW=3kH/np=1024). Spectra were corrected for eddy currents
using an in-house Matlab routine and quantified using LCModel9. A
basis-set including MM spectrum was created for both sequences and TE values. Results and discussion
Figure 2 shows results
of spectral simulations for some of the J-coupled
proton groups in brain metabolites for classic sLASER and J-sLASER sequences over a range of TE values. This illustrates the potential
to improve detection of metabolites such as lactate, glutamate and glutamine at
intermediate TE values.
Figure 3 shows
phantom results obtained with both sequences. It can be seen that the evolution
of the J-modulation for glutamate and
lactate follows the theoretical prediction in the simulation (Figure 2B). A slight decrease in the NAA
singlet signal with J-sLASER is due
the generation of residual stimulated echoes caused by the additional π/2 pulse and the slice
selection gradient.
In vivo measurements were performed in the PCC of the human brain (n=3). Figure 4
shows representative spectra acquired with J-sLASER
and sLASER at TE=45ms. The dramatic effect of the J-refocusing is seen on all glutamate peaks, as well as on the
lactate and glutamine. Finally, the CRLB values for three metabolites obtained
with LCModel from data acquired with both sequences at two different TE values
are reported in Table 1. While the overall
SNR was slightly lower due to loss of signal by the stimulated echo generated
by the added π/2 pulse, the CRLB decreased for glutamate and glutamine at both
TE values and for lactate at the shorter TE with the J-sLASER sequence. Conclusion
We
illustrated the potential of the J-sLASER
sequence to improve detection of J-coupled
resonances in the human brain at ultrahigh field. We found in vivo a strong improvement in glutamate detection. Further
investigations are needed to improve SNR, for example by using a non-selective π/2 pulse and thus minimizing the stimulated echo
generated by it, and aim to decrease TE below 45ms. The J-sLASER sequence offers benefits that may be used to improve the
measurement of small changes in metabolites in functional MRS experiments10,11,
as well as new perspective on spectral editing, by combining data from sLASER
and J-sLASER in conditions that
optimally maximize the differences for desired J-coupling systems. Acknowledgements
The authors would like to thank Drs. D.
Deelchand and P.G. Henry from Center of Magnetic Resonance Research at
University of Minnesota (USA) and Dr. Julien Valette at Atomic Energy and
Alternative Energies Commission in Paris (France) for sharing their Matlab
programs to create LCModel basis-sets.References
[1] P.
van Zijl et al. J.
Magn. Res. 1990; [2] K. Takegoshi et al. J. Magn. Res. 1989; [3] Pan et al.,
MRM 1996; [4] Pan et al., MRM 2010; [5] Mulkern et al., MRM 1996; [6] Chen et
al., NM in Biomed. 2013; [7] D.K. Deelchand et al., MRM 2017; [8] V. Govindaraju
et al., NMR in Biomed. 2000; [9] Provencher et al., MRM 1993; [10] Schaller et
al., NeuroImage 2014; [11] Bednařík et al., J. Cereb. Blood Flow
Metab. (2015)