Lumeng Cui1, Emily J McWalter1,2, Gerald Moran3, and Niranjan Venugopal4
1Division of Biomedical Engineering, University of Saskatchewan, Saskatoon, SK, Canada, 2Department of Mechanical Engineering, University of Saskatchewan, Saskatoon, SK, Canada, 3Siemens Healthcare Limited, Oakville, ON, Canada, 4Department of Radiotherapy Physics, CancerCare Manitoba, Winnipeg, MB, Canada
Synopsis
Ultrashort
echo time (UTE) pulse sequences have the unique ability to visualize short T2
tissues. While several UTE techniques have been implemented and demonstrated
good clinical results, there currently does not exist a flexible and robust UTE
pulse sequence which can easily manipulate multiple parameters “on-the-fly“ (i.e.
selection of trajectories, excitation pulses, acquisition parameters, etc.),
and obtain several complimentary scans in a single scan session. In this work
we present a newly developed hybrid UTE pulse sequence that includes multiple
excitation pulses, and varying trajectories, allowing for novel investigations
in studying short T2 tissues.
Introduction
Short T2 tissues appear as a
signal void on conventional MRI images due to their short relaxation times1;
however, there are several applications in which signal is required from these
tissues. Signal can be obtained from short T2 tissues using Ultrashort
Time-of-Echo (UTE) sequences2 that employ approaches such as non-Cartesian
trajectories and a half-pulse or hard pulse to reduce TE to the system limits2-5.
Tissues such as bone, tendons, ligaments and deep layers of cartilage have been
successfully visualized using UTE approaches however it is not clear what
particular trajectories, pulses, and associated parameters are optimal for this
visualization. This is because a single sequence for direct comparisons is
lacking. Thus, presented in this study, is a flexible robust UTE sequence that
incorporates varying combinations of trajectories, pulses, and voxel sizes which
enable multiple UTE data acquisitions in a single scanning session.Methods
A flexible UTE sequence was
developed (and tested on a 3T MRI (MAGNETOM Skyra, VE11C, Siemens Healthcare,
Erlangen, Germany) using an in-house phantom that contained both short and long
T2 properties. The current sequence can generate on-the-fly, radial and spiral
trajectories using the inline GUI. 2D acquisitions were developed using radial
and spiral trajectories and employed a half-Sinc excitation pulses. Similarly, 3D
acquisitions were developed using radial and spiral trajectories, but have the
option of using a half-Sinc or hard excitation pulses. The images were
reconstructed using standard methods. To measure the limits of the sequence,
and its execution on the 3T MRI, data was collected by 1) Varying the numbers
of spokes/interleaves and cones that comprised the radial and spiral
trajectories, 2) Performing an ultrashort and a long TE acquisitions, and 3) Switching
between the use of the half-Sinc versus hard pulses (See details in Figure 1). To
evaluate the acquired data, we observed the variation of the image artifacts
for 1), conducted a dual-echo subtraction (ultrashort and long) for 2), and
reported Signal-to-Noise-Ratio (SNR) in Figure 5.b using the two-region method6
for 3).Results
When the number of shots was
increased, the artifacts observed in fewer shots start disappearing for radial
and spiral trajectories in the 2D acquisition (Figure 2.a-n); and their results
at a high shot number (96/1024 shots for spiral/radial) are comparable to the
GRE image. In the 3D spiral-cone trajectory, when the cone number is increased, the
image quality improves with less artifacts (Figure 3.a.1-6). In terms of
spoke-cone combination, 512*256 or 362*362 suggests a better image quality than
other combinations (Figure 3.b.1-8 and c.1-8). With both the radial and spiral
trajectories it is feasible to visualize short T2 components (Figure 4.a-f). When the
half-Sinc and the Hard pulses were compared the spiral trajectories generally
had higher SNR (Figure 5.a.1-8 and b).
Discussion
We showed that it was possible
to develop and use a flexible UTE sequence to image short T2 substances. When
comparing the radial and spiral trajectory (Figure 2.a-n), the undersampled
shots result in the undersampling artifact for radial trajectory; while for
spiral trajectory, long spiral gradient readout due to fewer shots cause the
off-resonance artifact as observed. However, increasing the number of shots can
effectively suppress the artifacts, and the images acquired with a high shot
number are comparable to the GRE image while achieving an ultrashort TE (0.1 –
1 ms1). In a 3D trajectory, the artifacts are also from fewer cones in
addition to fewer spokes/interleaves. For example, with an insignificant
off-resonance artifact (Figure 3.a.1-6), the artifacts are successfully reduced
from 32 cones to 512 cones. Even though a larger number of spokes and cones are
more desirable for the 3D radial to obtain a high-quality image, there is a
limit for the number of gradients that the MR system can execute. Therefore, we
need to determine an optimal combination between spokes and cones (Figure
3.b.1-8 and c.1-8); the results suggest that spokes have more impact on the
image quality than the cone number. 3D radial and spiral can both capture more
short-T2 signals with an ultrashort TE (0.06 ms) than a longer TE (3.09 ms)
(Figure 4); the short T2 details (arrows) that are observed as voids in a long TE
image can be further enhanced by taking the subtraction between two TEs. For
the radial trajectory, the half-Sinc pulse’s SNRs on short T2, long T2, and the
whole region are higher than the SNRs of the hard pulses (with different flip
angles), but the total acquisition time of half-Sinc excitation is twice that
of the hard pulse excitation (Figure 5). For the spiral trajectory, there is an evident
difference of SNR on the short T2 and whole region between the half-Sinc pulse
and hard pulses, and the hard pulses with high flip angles demonstrate a higher
SNR on the long T2 region. Conclusion
The newly developed hybrid UTE
sequence demonstrated efficacy in preliminary phantom tests. Additionally, the
new sequence provided a flexible platform to adapt the pulse sequence for a
variety of UTE applications by enabling the end-user to employ varying trajectories,
dimensions, and excitation pulses.
Acknowledgements
No acknowledgement found.References
- Bydder
GM. The Agfa Mayneord lecture: MRI of short and ultrashort T 2 and T 2*
components of tissues, fluids and materials using clinical systems. The British
journal of radiology. 2011 Dec;84(1008):1067-82.
- Gold
GE, Pauly JM, Macovski A, Herfkens RJ. MR spectroscopic imaging of collagen:
tendons and knee menisci. Magnetic resonance in medicine. 1995
Nov;34(5):647-54.
- Bergin
CJ, Pauly JM, Macovski A. Lung parenchyma: projection reconstruction MR
imaging. Radiology. 1991 Jun;179(3):777-81.
- Glover
GH. Simple analytic spiral k‐space algorithm. Magnetic Resonance in Medicine:
An Official Journal of the International Society for Magnetic Resonance in
Medicine. 1999 Aug;42(2):412-5.
- Cline
HE, Zong X, Gai N. Design of a logarithmic k‐space spiral trajectory. Magnetic
Resonance in Medicine: An Official Journal of the International Society for
Magnetic Resonance in Medicine. 2001 Dec;46(6):1130-5.
- Dietrich
O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO. Measurement of signal‐to‐noise
ratios in MR images: influence of multichannel coils, parallel imaging, and
reconstruction filters. Journal of Magnetic Resonance Imaging: An Official
Journal of the International Society for Magnetic Resonance in Medicine. 2007
Aug;26(2):375-85.