YUXI PANG1
1Department of Radiology, University of Michigan, Ann Arbor, MI, United States
Synopsis
A self-compensated
spin-locking (SL) scheme for quantitative R1ρ dispersion
in cartilage has been developed. The performance of this new method was
evaluated by Bloch simulations and R1ρ dispersion (with 6 SL RF strengths ranging
from 50 to 1000 Hz) studies on agarose
(1-4%, w/v) phantom and on one healthy human knee in vivo at 3T, with respect
to three reported SL approaches. The simulated and experimental results
indicate that the proposed SL method was less susceptible to B0 and B1 field
artifacts for a wide range of SL strengths, and thus more suitable for
quantitative R1ρ dispersion
in ordered tissue.
Introduction
Quantitative R1ρ dispersion could provide
a unique structural information on collagen integrity in cartilage.1
Substantial artifacts could arise from non-uniform B0 and B1 fields in
varying R1ρ-weighted images, particularly at higher B0 with an
extreme spin-lock (SL) RF strength (ω1/2π).2-4 Recently, a few self-compensated
SL methods have been developed; however, none of these existing
approaches would be robust enough for a wide range of ω1/2π.5 Hence, this work aimed to introduce a new SL scheme
for quantitative R1ρ dispersion in
cartilage.Methods
(1) Spin-locking
schemes: Figure 1 sketches the reported (A-C)2-4
and the proposed (D) SL schemes. The digital numbers under these sketches indicate
the specific time points on which the simulated spin trajectories will be
visualized. (2) Spin dynamics simulations: A spin dynamics was simulated for 4 SL schemes with ω1/2π progressively
increased from 0 to 1000 Hz in 101 steps and likewise with Δω0/2π from 0 to 250
Hz. The nominal FA α
and β
were scaled
down 90% to mimic B1 inhomogeneity. (3) Spin trajectory
visualizations: Some simulations
were highlighted with an ensemble of 101 spins that were respectively exposed
to a nominal ω1/2π
of 500 Hz
with B1 homogeneity progressively improved from 90% to 100% of the nominal
value, and to a progressively increased Δω0/2π from 0 to 250
Hz. The spin dynamics were visualized in Figure 3 as follows: A0 -> A1 -> A2 (A); A0 -> A1 -> B2 -> B3 (B); A0 -> C1 -> C2 -> C3 -> C4 ->
C5 (C) and A0 -> C1 -> D2 -> D3 -> D4 -> D5 -> D6 (D). (4) MR
imaging protocol: R1ρ-prepared signals using 4 SL schemes were imaged by 3D TFE following an
elliptical centric phase-encoding (low-high) order on a 3T scanner using
a 16-ch T/R knee coil. Other parameters: ω1/2π=50, 125,
250, 500, 750, 1000 Hz; shot interval = 2 s; TFE factor = 64; TR/TE = 5.4/2.8
ms; FA = 10°
; voxel size = 1.0*1.0*3.0 mm3; CS-SENSE factor = 2.5; acquisition duration = 40 s
per 3D R1ρ-weighted image scan. B0 and B1 fields were mapped
on phantom using conventional dual-echo (
ΔTE of 3 ms) and dual-TR (TR=30 and 150 ms) methods. (5)
R1ρ dispersion modeling: A voxel intensity S (in a
logarithmic scale) in
R1ρ-weighted image of cartilage at 3T could be
expressed using the following Eq.,
$$ ln(S) = P-R_{2a}/(1+4ω_1^2τ_b^2) $$
where P stood for $$$ln(S_0)/TSL-R_{2i}$$$, S0 and TSL (=40
ms) were constant, R2i and R2a were two
predominant contributions to R2, and ω1 and τb were a SL RF
power and an effective “slow” correlation time.1 A sum of squared
errors (SSE) was used to assess the discrepancies between the measured and the
modeled data in vivo. When R1ρ dispersion
was not expected for agarose (1-4%, w/v) gels,6 their signal fluctuations
were quantified with coefficients of variation CVs (%) as ω1 altered. All
data analysis and visualization were carried out with a customized software
developed in IDL 8.5 (Harris Geospatial Solutions, Inc, Boulder, CO).Results and Discussion
Figure 2
presents the simulated longitudinal magnetizations. Compared with the reported
methods A-C, the proposed scheme D demonstrated markedly reduced sensitivity to
ΔB0
and ΔB1. This finding could be visualized on Bloch spheres
as shown in Figure 3, where the resulting spin trajectories just before α-y pulse were
represented in A2, B3, C5 and D6, indicating that the least and most close
trajectories with respect to A0 were A2 and D6, respectively. In other words, the
scheme D could offer not only a relatively higher tolerance to ΔB0 and ΔB1, but also a
lesser signal loss.
Figure 4A-D show
one exemplary R1ρ-weighted image slice of phantom. Qualitatively, the
scheme D demonstrated fewer banding artifacts, particularly when using a lower ω1/2π. Figure 4E presents quantitatively that the method
D (filled black circle) provided the least signal fluctuations with the
smallest average CV (%), based on the measurements from 2 ROIs located inside and
outside agarose tubes. ΔB0 (ppm) and B1 (%) field
maps for the same image slice are respectively shown in Figure 4F and 4G. The
observed B1 field was
very close to its normal value (i.e. ~100%); however, the measured ΔB0 was markedly
varied as much as more than 1 ppm (i.e. >128 Hz) near the interfaces in
which the banding artifacts became more prominent.
Figure 5
demonstrates one exemplary R1ρ-weighted image slice of human knee. The image banding
artifacts could barely be recognized except the image acquired with the scheme
A using ω1/2π of 125 Hz. Two
exemplary signal intensities were derived from 2 ROIs located on the posterior
tibial (PTC) and the central femoral (CFC) cartilage. The measured and fitted R1ρ dispersion
profiles were compared in the 3rd row. The observed SSE (*10-3)
for SL schemes A-D were respectively 1.2, 0.4, 0.1, and 0.3 for CFC (blue lines). In
comparison, those values were 7.2, 3.2, 2.8, and 0.4, for PTC (red lines). These
fitting results suggest that the proposed SL scheme D could improve R1ρ dispersion
quantification accuracy of human knee cartilage.Conclusion
In conclusion, a
robust spin-locking scheme has been developed for quantitative R1ρ dispersion on
human knee cartilage.Acknowledgements
This work was supported in part by the Eunice Kennedy
Shriver National Institute of Child Health & Human Development of the
National Institutes of Health under Award Number R01HD093626 (Prof. Riann
Palmieri-Smith). The content is solely the responsibility of the author and
does not necessarily represent the official views of the National Institutes of
Health. The
author would also like to thank Prof. Thomas Chenevert for his support
and encouragement, and Suzan Lowe and James O’Connor for their assistance in
collecting human knee images.References
1.
Pang Y. An order parameter without magic angle effect
(OPTIMA) derived from R1ρ dispersion in ordered tissue. Magn Reson Med 2019.
DOI: 10.1002/mrm.28045.
2.
Charagundla SR, Borthakur A, Leigh JS, Reddy R.
Artifacts in T-1 rho-weighted imaging: correction with a self-compensating
spin-locking pulse. J Magn Reson 2003;162(1):113-121.
3.
Zeng H, Danie lG, Gochberg C, Zhao Y, Avison M, Gore
JC. A composite spin-lock pulse for ΔB0 + B1 insensitive T1rho measurement. In:
Proceedings of the 14th Annual Metting of ISMRM, Seattle, Washington, USA,
2006. (abstract: 2356).
4.
Mitrea BG, Krafft AJ, Song RT, Loeffler RB, Hillenbrand
CM. Paired self-compensated spin-lock preparation for improved T-1 rho quantification.
J Magn Reson 2016;268:49-57.
5.
Chen W. Errors in quantitative T1rho imaging and the
correction methods. Quant Imag Med Surg 2015;5(4):583-591.
6.
Andrasko J. Water in agarose gels studied by nuclear
magnetic resonance relaxation in the rotating frame. Biophys J
1975;15(12):1235-1243.