2736

A self-compensated spin-locking scheme for quantitative R dispersion in articular cartilage
YUXI PANG1
1Department of Radiology, University of Michigan, Ann Arbor, MI, United States

Synopsis

A self-compensated spin-locking (SL) scheme for quantitative R dispersion in cartilage has been developed. The performance of this new method was evaluated by Bloch simulations and R dispersion (with 6 SL RF strengths ranging from 50 to 1000 Hz) studies on agarose (1-4%, w/v) phantom and on one healthy human knee in vivo at 3T, with respect to three reported SL approaches. The simulated and experimental results indicate that the proposed SL method was less susceptible to B0 and B1 field artifacts for a wide range of SL strengths, and thus more suitable for quantitative R dispersion in ordered tissue.

Introduction

Quantitative R dispersion could provide a unique structural information on collagen integrity in cartilage.1 Substantial artifacts could arise from non-uniform B0 and B1 fields in varying R-weighted images, particularly at higher B0 with an extreme spin-lock (SL) RF strength (ω1/2π).2-4 Recently, a few self-compensated SL methods have been developed; however, none of these existing approaches would be robust enough for a wide range of ω1/2π.5 Hence, this work aimed to introduce a new SL scheme for quantitative R dispersion in cartilage.

Methods

(1) Spin-locking schemes: Figure 1 sketches the reported (A-C)2-4 and the proposed (D) SL schemes. The digital numbers under these sketches indicate the specific time points on which the simulated spin trajectories will be visualized. (2) Spin dynamics simulations: A spin dynamics was simulated for 4 SL schemes with ω1/2π progressively increased from 0 to 1000 Hz in 101 steps and likewise with Δω0/2π from 0 to 250 Hz. The nominal FA α and β were scaled down 90% to mimic B1 inhomogeneity. (3) Spin trajectory visualizations: Some simulations were highlighted with an ensemble of 101 spins that were respectively exposed to a nominal ω1/2π of 500 Hz with B1 homogeneity progressively improved from 90% to 100% of the nominal value, and to a progressively increased Δω0/2π from 0 to 250 Hz. The spin dynamics were visualized in Figure 3 as follows: A0 -> A1 -> A2 (A); A0 -> A1 -> B2 -> B3 (B); A0 -> C1 -> C2 -> C3 -> C4 -> C5 (C) and A0 -> C1 -> D2 -> D3 -> D4 -> D5 -> D6 (D). (4) MR imaging protocol: R-prepared signals using 4 SL schemes were imaged by 3D TFE following an elliptical centric phase-encoding (low-high) order on a 3T scanner using a 16-ch T/R knee coil. Other parameters: ω1/2π=50, 125, 250, 500, 750, 1000 Hz; shot interval = 2 s; TFE factor = 64; TR/TE = 5.4/2.8 ms; FA = 10° ; voxel size = 1.0*1.0*3.0 mm3; CS-SENSE factor = 2.5; acquisition duration = 40 s per 3D R-weighted image scan. B0 and B1 fields were mapped on phantom using conventional dual-echo ( ΔTE of 3 ms) and dual-TR (TR=30 and 150 ms) methods. (5) R dispersion modeling: A voxel intensity S (in a logarithmic scale) in R-weighted image of cartilage at 3T could be expressed using the following Eq.,
$$ ln(S) = P-R_{2a}/(1+4ω_1^2τ_b^2) $$
where P stood for $$$ln(S_0)/TSL-R_{2i}$$$, S0 and TSL (=40 ms) were constant, R2i and R2a were two predominant contributions to R2, and ω1 and τb were a SL RF power and an effective “slow” correlation time.1 A sum of squared errors (SSE) was used to assess the discrepancies between the measured and the modeled data in vivo. When R dispersion was not expected for agarose (1-4%, w/v) gels,6 their signal fluctuations were quantified with coefficients of variation CVs (%) as ω1 altered. All data analysis and visualization were carried out with a customized software developed in IDL 8.5 (Harris Geospatial Solutions, Inc, Boulder, CO).

Results and Discussion

Figure 2 presents the simulated longitudinal magnetizations. Compared with the reported methods A-C, the proposed scheme D demonstrated markedly reduced sensitivity to ΔB0 and ΔB1. This finding could be visualized on Bloch spheres as shown in Figure 3, where the resulting spin trajectories just before α-y pulse were represented in A2, B3, C5 and D6, indicating that the least and most close trajectories with respect to A0 were A2 and D6, respectively. In other words, the scheme D could offer not only a relatively higher tolerance to ΔB0 and ΔB1, but also a lesser signal loss. Figure 4A-D show one exemplary R-weighted image slice of phantom. Qualitatively, the scheme D demonstrated fewer banding artifacts, particularly when using a lower ω1/2π. Figure 4E presents quantitatively that the method D (filled black circle) provided the least signal fluctuations with the smallest average CV (%), based on the measurements from 2 ROIs located inside and outside agarose tubes. ΔB0 (ppm) and B1 (%) field maps for the same image slice are respectively shown in Figure 4F and 4G. The observed B1 field was very close to its normal value (i.e. ~100%); however, the measured ΔB0 was markedly varied as much as more than 1 ppm (i.e. >128 Hz) near the interfaces in which the banding artifacts became more prominent. Figure 5 demonstrates one exemplary R-weighted image slice of human knee. The image banding artifacts could barely be recognized except the image acquired with the scheme A using ω1/2π of 125 Hz. Two exemplary signal intensities were derived from 2 ROIs located on the posterior tibial (PTC) and the central femoral (CFC) cartilage. The measured and fitted R dispersion profiles were compared in the 3rd row. The observed SSE (*10-3) for SL schemes A-D were respectively 1.2, 0.4, 0.1, and 0.3 for CFC (blue lines). In comparison, those values were 7.2, 3.2, 2.8, and 0.4, for PTC (red lines). These fitting results suggest that the proposed SL scheme D could improve R dispersion quantification accuracy of human knee cartilage.

Conclusion

In conclusion, a robust spin-locking scheme has been developed for quantitative Rdispersion on human knee cartilage.

Acknowledgements

This work was supported in part by the Eunice Kennedy Shriver National Institute of Child Health & Human Development of the National Institutes of Health under Award Number R01HD093626 (Prof. Riann Palmieri-Smith). The content is solely the responsibility of the author and does not necessarily represent the official views of the National Institutes of Health. The author would also like to thank Prof. Thomas Chenevert for his support and encouragement, and Suzan Lowe and James O’Connor for their assistance in collecting human knee images.

References

1. Pang Y. An order parameter without magic angle effect (OPTIMA) derived from R1ρ dispersion in ordered tissue. Magn Reson Med 2019. DOI: 10.1002/mrm.28045.

2. Charagundla SR, Borthakur A, Leigh JS, Reddy R. Artifacts in T-1 rho-weighted imaging: correction with a self-compensating spin-locking pulse. J Magn Reson 2003;162(1):113-121.

3. Zeng H, Danie lG, Gochberg C, Zhao Y, Avison M, Gore JC. A composite spin-lock pulse for ΔB0 + B1 insensitive T1rho measurement. In: Proceedings of the 14th Annual Metting of ISMRM, Seattle, Washington, USA, 2006. (abstract: 2356).

4. Mitrea BG, Krafft AJ, Song RT, Loeffler RB, Hillenbrand CM. Paired self-compensated spin-lock preparation for improved T-1 rho quantification. J Magn Reson 2016;268:49-57.

5. Chen W. Errors in quantitative T1rho imaging and the correction methods. Quant Imag Med Surg 2015;5(4):583-591.

6. Andrasko J. Water in agarose gels studied by nuclear magnetic resonance relaxation in the rotating frame. Biophys J 1975;15(12):1235-1243.

Figures

FIGURE 1. The previous (A-C) and proposed (D) spin-locking preparation schemes. α, β, τ and x/-x/y/-y are respectively the flip angles for tip-down/up and refocusing RF pulses, and the spin-lock RF pulse duration and phases. The digital numbers below the diagrams represent the time points at which the spin trajectories will be visualized on Bloch spheres.

FIGURE 2. Simulated longitudinal magnetizations after 4 spin-locking schemes (A-D). The nominal spin-locking RF strengths (ω1/2π) were limited to [0, 1000] Hz with B1 homogeneity set to 90% of its nominal value, and the chemical shifts (Δω0/2π) limited to [0, 250] Hz to mimic B0 inhomogeneity reported for human knee cartilage in vivo at 3T.

FIGURE 3. Snapshots of spin-locking prepared spin trajectories (n=101) on Bloch spheres at the time points specified in Figure 1. These simulated spins were “locked” by 4 schemes (ω1/2π =500 Hz), subjected to B1 homogeneity progressively increased from 90% (black) to 100% (red), and to B0 inhomogeneity increased from 0 (black) to 250 Hz (red). The orders of the simulated spin trajectories are as follows, A0->A1->A2 for the method A; A0->A1->B2->B3 for the method B; A0->C1->C2->C3->C4->C5 for the method C; and A0->C1->D2->D3->D4->D5->D6 for the proposed method D from this work.

FIGURE 4. Exemplary constant-time (TSL=40 ms) R-weighted images (A-D) of an agarose phantom acquired with 4 spin-locking RF powers (i.e. ω1/2π=50, 125, 500, 1000 Hz) and 4 spin-locking methods at 3T. Average signal intensities were measured from 2 ROI (red and black) highlighted by red arrows on upper-right images, and their CVs (%) were compared (E), which were derived from six R-weighted images (i.e. ω1/2π=50, 125, 250, 500, 750, 1000 Hz) for different agarose (1-4%) gels. B0 inhomogeneity (in ppm) and B1 homogeneity (%) maps for the same image slice are presented in Figure 4F-G.

FIGURE 5. Exemplary constant-time (TSL=40 ms) R-weighted images of human knee in vivo at 3T acquired with 4 spin-locking methods and two ω1/2π values of 125 (1st row) and 500 Hz (2nd row). Two ROIs were highlighted by yellow arrows on the posterior tibial and the central femoral cartilage locations as shown on the upper-left image, and R-weighted signal intensity dependencies on ω1/2π from these two ROIs were modeled and compared in the 3rd row, with filled symbols and solid lines standing for the measured and the fitted for the femoral (blue) and tibial (red) cartilage.

Proc. Intl. Soc. Mag. Reson. Med. 28 (2020)
2736