Matthew R Orton1, Mihaela Rata2, Nina Tunariu2, Andra Curcean2, Julie Hughes2, Erica Scurr2, James D'arcy1, Matthew D Blackledge1, and Dow-Mu Koh2
1Division of Radiotherapy and Imaging, Institute of Cancer Research, Sutton, United Kingdom, 2Department of Radiology, Royal Marsden NHS Foundation Trust, Sutton, United Kingdom
Synopsis
Accurate characterisation of bone metastases and
their response to treatment remains an unmet need in oncology. Magnetic
resonance fingerprinting (MRF) is an imaging technique that yields inherently
co-registered quantitative images from a single acquisition, so this work aims
to evaluate measurement repeatability of MRF when measuring T1 and T2
relaxation times in metastatic bone disease.
We found good measurement repeatability of 6% for lesion mean T1 and 23%
for lesion mean T2.
Background
Magnetic resonance fingerprinting (MRF) is an
imaging technique that yields inherently co-registered quantitative images from
a single acquisition1. Evaluations
of its application in a number of soft-tissue oncology settings are ongoing,
but there is also potential in using this technology to better characterise metastatic
bone disease, since accurate characterisation of bone disease and its response
to treatment remains an unmet need in oncology. In quantitative imaging, whilst absolute
accuracy remains the gold-standard performance criterion, measurement
repeatability is also of major interest when assessing the utility of such
measures in clinical practice.Aim
To evaluate the repeatability
of MRF-derived T1 and T2 relaxation times in metastatic bone disease in
patients with primary prostate cancer.Methods
All imaging was performed on a
1.5T MAGNETOM Aera (Siemens Healthineers, Erlangen, Germany) according to an ethics
and research board approved protocol. In
an ongoing study, nine prostate cancer patients (mean age 69.8 years, range
58-83 years) were imaged using a prototype MRF-FISP sequence2 with
Gadgetron reconstruction3 which yields maps of T1 and T2 relaxation
times, and the proton density M0. To aid
lesion localisation, diffusion-weighted images (DWI) and T1-weighted Dixon
images were also acquired using the sequences described in Table 1. MRF
acquisitions were performed twice, between which the patients were removed from
the scanning table and allowed to relax before being returned to the table. Care was taken to match the patient and
pelvic coil positioning for the second scan as closely as possible to the first
scan. Five contiguous 5mm thick slices
were positioned over the pelvis to include lesions that appeared hyper-intense
on high b-values images and hypo-intense on the post-processed relative fat-fraction
Dixon images.
Volumetric regions of interest
(ROIs) were drawn for multiple lesions per patient on the MRF M0 image, using
the DWI and relative fat-fraction Dixon images as a visual guide. ROIs drawn on the first MRF scan were copied
to the second scan and manually adjusted to accommodate any changes in pose and
position. Contours were drawn using
Horos (Horosproject.org), saved as DICOM-RT structure sets using a pyOsiriX
plugin4, and image statistics were obtained using MATLAB (R2019a The
MathWorks, Inc.). The following lesion statistics
were computed from the voxel values in each lesion: mean, median, standard
deviation, first quartile (Q1), third quartile (Q3) skewness and kurtosis.
Repeatability was visually
assessed with Bland-Altman plots, and statistically measured using the
repeatability coefficient (r), and intra-class correlation (ICC). In addition, the cohort mean across all
lesions of each statistic was recorded, from which a relative repeatability
coefficient can be derived using r% = r/(cohort mean)×100%. The between-lesion (sB) and within-lesion (sW)
standard deviations were recorded separately, from which ICC = sB2/(sB2+sW2).Results
The median number of lesions studied per patient
was 2 (range 1-7), and the median lesion volume was 1106 ml (range 171 – 11,976
ml). The smallest lesion was visibly
impacted by partial volume effects, and so was excluded from the repeatability
analysis, giving a total of 26 lesions. Figure 1 shows a single slice from an example patient with three lesions. Repeatability statistics for all lesion
statistics are detailed in Table 2. The following discussion focusses on r% and
ICC as these are dimensionless quantities, but for completeness, all relevant
statistics are included in Table 2. The Bland-Altman plots for the mean
lesion statistics are shown in Figure 2. Also included in these plots are points
corresponding to the mean voxel value per patient, i.e. averaged across all
lesions per patient.Discussion
The
relative repeatability (r%) of T1 mean is 6.22%, T2 mean is 22.8% and M0 mean
is 18.0%, implying that proportional changes less than these values can be
attributed to measurement error, while excess changes are likely to be
treatment related. The corresponding ICCs are 0.95, 0.88 and 0.87,
and taken with r%, these statistics indicate that T1 mean statistics are more
repeatable than both T2 and M05.
A similar MRF repeatability study in brain6 also reported better
repeatability for T1 than T2 (3.4% and 8.0%), although the proportional increase in
T2 r% compared to T1 r% observed here is larger than that reported in the brain. For T2 and M0, the lesion statistics std, Q1
and Q3 maintain similar r% and ICC to the mean, although for skewness and
kurtosis the repeatability tends to be degraded. For T1, r% for lesion standard deviation is
around three times larger than for the lesion mean, implying that assessment of
T1 heterogeneity is more prone to measurement variations than mean T1. The Bland-Altman plots indicate that the
repeat measures differences and cohort variations are well behaved, and both
sources of variation are smaller when the mean value per patient is
considered.Conclusions
Repeatability of T1, T2 and M0 measurements
in metastatic prostate cancer bone disease are in line with, but somewhat
larger than those previously reported in phantom5 and brain studies6.
Further work is ongoing to assess the magnitude of treatment related changes in
T1 and T2 that occur in bone metastases compared with the repeatability limits
reported here.Acknowledgements
The MRF sequence used
in this work was provided by Prof Vikas Gulani (University of Michigan) and
Prof Mark Griswold (Case Western Reserve University), with additional support
from Dr Yun Jiang. We acknowledge NHS
funding to the NIHR Biomedical Research Centre and the Clinical Research
Facility. This study represents independent research
part funded by the National Institute for Health Research (NIHR) Biomedical
Research Centre at the Royal Marsden NHS Foundation Trust and the Institute of
Cancer Research. The views expressed are those of the author(s) and not
necessarily those of the NHS, the NIHR or the Department of Health.References
- Ma D, et
al., Magnetic
resonance fingerprinting. Nature. 2013 14;495(7440):187–192.
- Jiang Y, et
al. MR Fingerprinting Using Fast Imaging with Steady State Precession (FISP)
with Spiral Readout, Magn. Reson. Med. 2015;74(6):1621-1631.
- Lo
W-C, et al. MR Fingerprinting using a
Gadgetron-based reconstruction, Proc. ISMRM 2018, #3525.
- Blackledge
MD, et
al., Rapid
development of image analysis research tools: Bridging the gap
between researcher and clinician with pyOsiriX.
Comput Biol Med. 2016;69:203-12.
- Jiang
Y, et al., Repeatability of magnetic resonance fingerprinting T1 and T2
estimates assessed using the ISMRM/NIST MRI system phantom. Magn Reson Med.
2017 78(4):1452-1457.
- Körzdörfer
G, et al., Reproducibility and Repeatability of MR Fingerprinting Relaxometry
in the Human Brain. Radiology. 2019 Aug;292(2):429-437.