Steffen Ringgaard1 and Anne Dorte Blankholm2
1MR Research Centre, Aarhus University, Aarhus, Denmark, 2Department of Radiology, Aarhus University Hospital, Aarhus, Denmark
Synopsis
When doing
arterial spin labeling perfusion measurements in the kidneys, respiratory motion
is an important issue. We compared different methods for doing retrospective
motion correction of single-shot ASL images. We evaluated the temporal signal
variation in the cortex after correction by translation, translation+rotation and
affine transformation and found that the variation was lowest when using full
affine correction. Besides, we also compared different similarity measures as
used during the motion correction.
Introduction
Measurement
of blood perfusion in the kidneys is important for diagnosing kidney diseases, and
arterial spin labeling (ASL) is preferable over contrast agent based methods.
Measurement can be performed with either the FAIR or pCASL methods, and for
both methods, quantitative perfusion maps is based on subtraction of label and
control images. For obtaining sufficient signal-to-noise ratio, a number of
repetitions are needed, because the signal difference between label and control
images is only 4-6 % of the equilibrium magnetization. Usually 20-30
repetitions are applied.
Due to
respiration, the kidneys are moving several centimeters, mainly in foot-head
direction. Usually prospective motion correction (such as gating to a
respiration sensor) is applied, but still severe motion of the images is seen. The
perfusion is mainly in the renal cortex, which is only 3-6 mm thick, and for
obtaining reliable perfusion values, careful retrospective motion correction is
therefore necessary. This can be done in various ways and both manual
correction1, rigid body motion correction2 and correction
using affine transformation3 of the kidneys has been applied.
We have
performed a systematic analysis of the effect of various motion corrections on
the robustness of the resulting perfusion maps.Methods
Kidney
perfusion was assessed in 19 healthy volunteers using a 1.5T Philips Achieva
dStream MR scanner. Images were acquired in an oblique coronal orientation
using FAIR labelling and single-shot spin echo data sampling. Five slices were
sampled with a slice thickness of 5 mm and 3x3 mm in-plane resolution. Twenty
sets of label and control images were acquired over 7 minutes. Respiration
gating was used with a respiration sensor placed on the abdomen.
Motion
correction was made using affine transformation of the kidneys and similarity
with the first control image was obtained for the various affine transformations.
Different sets of transformations were investigated:
- No retrospective correction.
- Correction by combined rigid body
translation for both kidneys.
- Correction by rigid body translation
separately for each kidney.
- Correction by individual rigid body
translation + rotation.
- Full affine individual transformation.
This included translation, rotation, scaling and shearing (Figure 1).
Different
ways to assess the
similarity with
the reference kidney was also investigated. We compared three measures:
- Mean absolute pixel difference.
- Mean squared pixel difference.
- Mean absolute pixel difference of
image gradient maps.
The
correction algorithm was implemented as an iterative search for maximum
similarity, and took about 10 secs per slice.
The accuracy
of the various motion corrections was assessed by segmenting out the cortex (6
mm rim in the kidney edges) in the reference image. The standard deviation of
pixel intensities over the 20 repetitions was calculated for each correction
method. This was used as a measure of noise in the calculated perfusion maps.
Results
The
standard deviation of the signal in the cortex through the 20 dynamics can be
seen in Figure 2. There was a statistically significant difference using global
translation and no correction (p<0.001), between individual rigid body
motion correction for each kidney and global translation (p<0.001), and between
full affine correction and rigid body correction (translation+rotation)
(p<0.001). No difference was found between individual kidney translation and
individual kidney translation+rotation.
Comparing the
three different similarity measures we found that using absolute difference
performed slightly better than using the squared difference and also than using
absolute difference of gradient maps (Figure 3).
By proper
motion correction, clearer perfusion maps can be generated. In Figure 4, an
example of perfusion maps generated by combined kidney rigid body motion
correction and full affine corrected are compared.Discussion
The main
conclusions from the study are that retrospective motion correction of ASL
perfusion measurement is necessary prior to quantitation of perfusion. Significant
improved correction is obtained by individual correcting each kidney,
indicating that the two kidneys are displaced differently by respiration. A
further improvement was found, when full affine correction was used.
The study
also showed that the most efficient way to assess the similarity of two kidneys
as applied during the correction process, is to use either the absolute
difference or squared difference and not use gradient maps.
The main
limitation of this work is that we did only in-plane motion correction and did
not take through-plane motion into account. This was partially because the
motion in the through-plane direction was small using the oblique-coronal slice
orientation, and partially because the slice thickness was larger (5 mm) than the
pixel size. To obtain better motion correction, 3D imaging with retrospective
correction in all 3 directions might be used, but this is difficult to do with
single-shot imaging, and with multi-shot imaging the motion artifacts are much
more difficult to handle.
Without
careful motion correction, the edges of the kidneys can be enhanced in the
perfusion maps due to the motion, and because the blood perfusion is mainly in
the cortex, this might be confused with high perfusion values.
During development of the methods, we also tried to use non-affine correction, but
this was not robust enough, and we continued with the affine correction method. Conclusion
Retrospective
motion correction is necessary for kidney ASL measurement, and full affine correction
was optimal.Acknowledgements
NoneReferences
- Buchanan CE, Cox EF, Francis ST.
Evaluation of 2D imaging schemes for pulsed arterial spin labeling of the human
kidney cortex. Diagnostics 2018;8:43.
- Artz
NS, Sadowski EA, Wentland AL, et al. Arterial spin labeling MRI for assessment of perfusion in native and
transplanted kidneys. Magn Reson Imaging 2011;29:74-82.
- Gardener AG, Francis ST. Multislice
perfusion of the kidneys using parallel imaging: Image acquisition and analysis
strategies. Magn Reson Med. 2010;63:1627–1636.