Yuki Koshino1, Naoki Ohno2, Tosiaki Miyati2, Tetsuo Ogino3, Yu Ueda3, Naoki Hori1, Yukihiro Matsuura1, Toshifumi Gabata4, and Satoshi Kobayashi1,2,4
1Radiology Division, Kanazawa University Hospital, Ishikawa, Japan, 2Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Ishikawa, Japan, 3Philips Japan, Tokyo, Japan, 4Department of Radiology, Kanazawa University Graduate School of Medical Sciences, Ishikawa, Japan
Synopsis
To
investigate the potential benefit of second-order motion compensated
(2nd-MC) diffusion encoding scheme on intravoxel incoherent
motion (IVIM) analysis of the kidney, we compared the IVIM diffusion parameters
and repeatability between the 2nd-MC and conventional diffusion
gradients. Our results showed that the 2nd-MC diffusion gradients show better fitting accuracy
and repeatability of IVIM diffusion parameters of the kidney compared with the
conventional diffusion gradients. Therefore,
the 2nd-MC diffusion encoding scheme could enable us to reduce the
bulk motion effect mainly caused by cardiac and respiratory motion during
diffusion sensitization.
INTRODUCTION
Intravoxel incoherent motion (IVIM) analysis of the kidney provides both
perfusion and diffusion information and is useful to assess renal function and
classify renal tumor.1,2 However, the clinical application of the
IVIM analysis is still hampered by the low measurement accuracy and
repeatability. One possible reason for this limitation is bulk motion mainly
caused by cardiac and respiratory motion during diffusion sensitization,
resulting in signal loss.3 Previous studies have reported that second-order
motion compensated (2nd-MC) diffusion encoding scheme reduced the
bulk motion sensitivity in cardiac diffusion MRI.4.5 Therefore, to
investigate the potential benefit of the 2nd-MC diffusion encoding
on IVIM analysis of the kidney, we compared the IVIM diffusion parameters and
repeatability between 2nd-MC and conventional diffusion encoding
schemes.METHODS
On a 3.0-T MRI, coronal diffusion-weighted images
of the kidney were obtained in eight healthy volunteers (all men; mean age,
25.4 ± 3.8 years) using respiratory-synchronized single-shot diffusion echo-planar
imaging with the 2nd-MC and conventional diffusion gradients (Fig. 1). The imaging
parameters were as follows: repetition time, one respiration cycle; echo time,
shortest (83-85 ms for 2nd-MC and 50-52 ms for conventional diffusion
gradients); acquisition matrix, 96 × 119; b-values, 0, 10, 30, 50, 100, 200,
400, 600, and 800 s/mm2; field of view, 300 mm; slice thickness, 7
mm; number of signals averaged, 2; and sensitivity encoding factor, 2. All
subjects were asked to fast for at least 6 hours before the scan. The same scan
was repeated twice to assess the repeatability of the measurements. We
determined mean signal intensities in the renal cortex and medulla at each
b-value. Then, perfusion-related diffusion coefficient (D*), the fraction (F), and
restricted diffusion coefficient (D) were calculated from biexponential
fitting. Normalized root mean square errors (nRMSE) were obtained to assess the
fitting quality.6 Repeatability coefficients (RC) were calculated
from Bland-Altman plots to assess the repeatability of diffusion parameters.
These values were compared between the 2nd-MC and conventional schemes.RESULTS AND DISCUSSION
Tables 1 and 2 show the results of IVIM diffusion parameters and nRMSE
with 2nd-MC and conventional diffusion gradients in the renal cortex
and medulla, respectively. Representative images of diffusion parameters and
nRMSE with both schemes are presented in Figure 2. The F in the cortex and D*
in the medulla with the 2nd-MC were significantly lower than those
with the conventional diffusion gradients. These results could be explained by
the difference in signal attenuation caused by coherent motion and flow between
the 2nd-MC and conventional diffusion gradients. The cortical and
medullary D with the 2nd-MC were significantly higher than those
with the conventional diffusion gradients. This may be due to the dependence of
restricted diffusion on the difference in the diffusion encoding waveform between
both schemes. Moreover, the nRMSE in the cortex and medulla significantly
reduced when using the 2nd-MC diffusion gradients, suggesting
improved fitting accuracy. Table 3 shows the RC of diffusion parameters with both
schemes. The 2nd-MC diffusion gradients showed better repeatability
for D and F in the cortex and D* and F in the medulla compared with the
conventional diffusion gradients.CONCLUSION
The 2nd-MC diffusion gradients show better
fitting accuracy and repeatability of IVIM diffusion parameters of the kidney
compared with the conventional diffusion gradients.Acknowledgements
No acknowledgement found.References
1. Gaing
B, et al. Subtype differentiation of renal
tumors using voxel-based histogram analysis of intravoxel incoherent motion
parameters. Invest Radiol. 2015; 50:
144-152.
2. Bane
O, et al. Assessment of renal function using
intravoxel incoherent motion diffusion-weighted imaging and dynamic
contrast-enhanced MRI. J Magn Reson Imaging.
2016; 44: 317-326.
3. Jerome NP, et al. Comparison
of free-breathing with navigator-controlled acquisition regimes in abdominal
diffusion-weighted magnetic resonance images: Effect on ADC and IVIM
statistics. J Magn Reson Imaging.
2014; 39: 235-240.
4.
Nguyen C, et al. In vivo
diffusion-tensor MRI of the human heart on a 3 tesla clinical scanner: An
optimized second order (M2) motion compensated diffusion-preparation approach. Magn Reson Med. 2016; 76: 1354-1363.
5.
Aliotta E, et al. Convex optimized
diffusion encoding (CODE) gradient waveforms for minimum echo time and bulk
motion-compensated diffusion-weighted MRI. Magn
Reson Med. 2017; 77: 717-729.
6.
Lv J, et al. Performance of U-net based
pyramidal lucas-kanade registration on free-breathing multi-b-value diffusion
MRI of the kidney. Br J Radiol. 2018;
91: 20170813.