Hina Arif1, Stephane Chartier 2, Tanmay Tiwari3, Bobby Kalb2, and Manoj Saranathan2
1Medical Imaging, University of Arizona, Tucson, AZ, United States, 2University of Arizona, Tucson, AZ, United States, 3University of Arziona, TUCSON, AZ, United States
Synopsis
Uniform fat suppression is critical
for optimal MR assessment of lesions in the abdomen and pelvis. Recent advances
using SPAIR sequencing have improved fat suppression and increased the
signal-to-noise ratio versus older techniques. However, fat suppression still
remains challenging in anatomical regions away from the magnetic isocenter due
to field inhomogeneity. Here we demonstrate that continuous table movement
(CTM) is superior to fixed table (FT) technology for SPAIR-mediated fat
suppression across the entire image stack. As field inhomogeneity away from
isocenter is a common challenge, CTM may provide a solution to many frequently
encountered problems.
Abstract
Introduction:
Effective and uniform fat suppression
on axial T2-weighted single-shot imaging is critical for optimal assessment of
lesions in the abdomen and pelvis. Advances in single shot fast spin echo
(SSFSE) imaging, specifically the use of spectral adiabatic inversion recovery
(SPAIR), have shown superior fat suppression and increased signal-to-noise
ratio of liver lesions compared to older techniques such as inversion-recovery
(IR) SSFSE. However, in slice locations far from the table isocenter, effective
fat suppression remains challenging for all types of sequences including SPAIR.
Recently, continuous table movement (CTM) technology has been introduced on the
Siemens platform that enables every slice to be acquired at or near the
isocenter. We investigated the effect of CTM on the quality of fat suppression compared
to fixed table (FT) multi-station acquisitions .
Methods:
All
patients were scanned on a Siemens 1.5T MRI scanner. A total of 92 patients
were retrospectively analyzed : 43 female, age 16 – 87, average 62.5 and 49
male, age 35 – 87, average 62.2. Each patient was imaged with a SPAIR fat-saturated
T2-weighted SSFSE sequence using both a CTM and a FT multi-slice acquisition
mode. Acquisition scan parameters were kept identical between both techniques. For
each abdominal exam, slices at three different
locations (top, center, and bottom slices in the imaging slab) were selected,
with the center location representing the slice closest to magnetic isocenter.
Four ROIs (anterior, posterior, left, right) were then outlined within each
slice. The mean signal intensity was recorded, and the ratio of mean intensity of
FT to CTM was calculated for each of the ROIs for each slice. A t-test was used
to evaluate statistical significance.
Results:
Figure 1 shows representative images from FT (top) and CTM
(bottom) SSFSE for top, center, and bottom slice locations in the slab. Note
the significantly better fat suppression in CTM SSFSE especially for the outer
slices (bottom left, bottom right). Figure 2 tabulates FT to CTM signal
intensity ratios for the three slice locations for all 4 ROIs. The edge slices
have significantly higher ratios (p<0.006) indicating significantly worse
fat sat for FT in the left and right ROIs. The central location has comparable
fat sat for both methods due to proximity to isocenter.
Discussion:
Over the last decade, continuous table
movement (CTM) technology has been developed and refined to allow acquisition
of images while the patient is continuously moving through the magnet. In
contrast to the conventional multi-station fixed table (FT) approach, CTM
allows for every image to be acquired at or near the magnetic isocenter over
large z-fields- of- view (FOV) with a consequent reduction in image artifact
from field inhomogeneity and improved scan times. Sequential 2D and volumetric
3D techniques have been explored but not quantitatively in a clinical setting.
MR angiography with peripheral run-off and whole-body cancer staging comprise
the two major categories of clinical studies in which the potential benefits of
CTM have been investigated. Fewer studies have been performed to demonstrate
the utility of CTM technology in decreasing effects of field inhomogeneity on
image quality in anatomic regions which are conventionally positioned distant
from magnetic isocenter. CTM technology provides improved quality of fat
suppression compared to fixed table (FT) multi-station acquisitions by allowing
optimal slice homogeneity for spectral fat selection radiofrequency pulses. The
results from our quantitative analysis show the greatest CTM gains in
subcutaneous fat suppression and signal uniformity occur in the superior and
inferior beginning and end slices, locations furthest from magnetic isocenter
when acquiring conventional fixed-table multi-slice sequences. There were no
significant gains in fat suppression or uniformity in the mid slices.
Conclusion:
CTM technology affords consistent
imaging at or near the magnetic isocenter while also reducing the time required
for image acquisition. Our data reflects this by the improved fat suppression
(recorded as reduced mean signal intensity) compared to conventional FT
methodology, particularity at outer slice positions where field inhomogeneity
interrupts fat-water frequency shift.
Legends
Figure 1. Representative images from FT SSFSE (top) and CTM SSFSE (bottom) shown for top, center, and bottom slice locations in the slab. Note the significantly uniform and efficient fat suppression in CTM SSFSE esp. for the edge slices (bottom left, bottom right) and comparable fat suppression for the near isocenter center slice (middle column).
Figure 2. Table showing FT to CTM signal intensity ratio. The edge slices have higher ratios indicating significantly poor fat sat for FT in the left and right ROIs as also visualized in Figure 1. * indicates statistical significance of p<0.006.Acknowledgements
No acknowledgement found.References
Lauenstein et al., Evaluation of
optimized inversion-recovery fat-suppression techniques for T2-weighted
abdominal MR imaging. J Magn Reson Imaging. 2008 Jun;27(6):1448-54.
Sengupta S, Smith DS, Welch EB. Continuously moving table MRI with golden angle radial sampling. Magn Reson Med. 2014 Dec 02; [doi: 10.1002/mrm.25531]