Tustomu Tamada1, Yu Ueda2, Kosuke Ito3, Takeshi Fukunaga1, Ayumu Kido1, Tomohiro Mochizuki4, and Akira Yamamoto1
1Radiology, Kawasaki Medical School, Kurashiki City, Japan, 2Philips Japan, Minato-ku, Japan, 3Kawasaki Medical School, Kurashiki City, Japan, 4Philips Japan, Osaka, Japan
Synopsis
T2WI is a key component of
mpMRI in prostate. 2D T2WI obtained in multi plane is recommend in PI-RADS v2,
resulting in prolonged examination time. The usefulness of 3D-VRFA-TSE,
which has a potential to reduce scan
time by multiplanar postprocessing reconstruction of images into any desired
plane, has been reported. As an further effective way to accelerate scan time
is provided by C-SENSE, which uses undersampling of the k-space, we compared 3D-VRFA-TSE with 3D-VRFA-TSE
with C-SENSE. C-SENSE enables
a reduction in acquisition time of 40% in 3D-VRFA-TSE,
maintaining image quality and lesion conspicuity except for clarity of
prostatic capsule.
Body of the abstract
INTRODUCTION: T2-weighted imaging (T2WI) which
is used for anatomical evaluation of prostate is key components of prostate multiparametric MRI (mpMRI) in patients with prostate cancer (PC) that contributes to tumor
detection as well as assessment of extracapsular extension (ECE).1,2
The T2WI acquisition in mpMRI is commonly performed with T2-weighted
two-dimensional turbo spin-echo (2D-TSE) repeated in multiple planes1.
The 2D TSE sequence are adversely
affected by relatively thick slices and as voxels are not isotropic, several acquisitions are necessary to enable displaying the prostate
in multiple planes. Therefore comparatively
long examination times that are prone to motion artifacts have to be scheduled.
Thus, using a three-dimensional (3D)
TSE sequence would be a
possible method to reduce total imaging time because it can serve as source data for multiplanar
postprocessing reconstruction of
images into any desired plane.3 A recent study with comparison
between 2D-TSE and 3D-TSE has been demonstrated that 3D-TSE with tissue-specific
variable refocusing flip angle (TS-VRFA, Philips Medical Systems, Best, The
Netherlands) could potentially replace multiplane 2D-TSE for prostate
cancer diagnosis while maintaining good
image quality.3 Furthermore, compressed SENSE (C-SENSE)
which is a combination of the parallel imaging technique SENSE (sensitivity
encoding) together with compressed sensing allow for a reduction of the amount
of acquired k-space data therefore enabling speeding up of MRI acquisitions.4 Thus, the aim of this study was to compare the image
quality and PC detection ability between 3D-TSE with TS-VRFA (3D-VRFA-TSE) and 3D-VRFA-TSE
with C-SENSE at 3-T system.
METHODS: A total of 21 patients with elevated PSA
levels (mean age, 71 years) underwent mpMRI including DWI on a 3-T with a
32-channel phased-array coil (Ingenia Elition 3.0T; Philips Medical Systems).
Axial 3D-VRFA-TSE and 3D-VRFA-TSE with C-SENSE was acquired with the imaging parameters
described in Figure 1. In qualitative visual
assessment, contrast of transitional zone (TZ) and peripheral zone (PZ) (CTP,
1=none, 2= slightly high, 3= moderately high, 4=very high), clarity of
prostatic capsule (CAP, 1=poor, 2= fair, 3=good, 4=excellent), blurring and
motion artifacts (BMA, 1=strong, 2= moderate, 3=weak, 4=none), overall image
quality (OIQ, 1=poor, 2= fair, 3=good, 4=excellent), and contrast of tumor and
normal tissue (CTN, 1=poor, 2= fair, 3=good, 4=excellent) were assessed, and in
quantitative analysis, signal-to-noise ratio (SNR) (SNR= mean signal intensity
(SI) in normal prostate / standard deviation (SD) of SI in internal obturator
muscle), and tumor-to-normal tissue ratio (TNR) (TNR=mean
SI in tumor / mean SI in normal prostate) were measured. CTP, CAP, BMA, OIQ, CTN, SNR, and TNR were compared between 3D-VRFA-TSE and
3D-VRFA-TSE with C-SENSE. The Friedman
and Wilcoxon signed rank tests were used for statistical analysis.
RESULTS: Acquisition time was 5
min 48 s in 3D-VRFA-TSE and 3 min 30 s in 3D-VRFA-TSE with C-SENSE. Of the 21
patients, 14 were diagnosed with PC and 7 with benign conditions. The PC was in
the PZ in 57% (8 of 14) and in the transition zone (TZ) in 43% (6 of 14). There was no significant difference in CTP, BMA, OIQ, CTN, SNR, and TNR
between 3D-VRFA-TSE and 3D-VRFA-TSE with C-SENSE (2.81 ± 0.96 vs.
2.90 ± 0.92, P = 0.16, 3.76 ± 0.43 vs. 3.81 ± 0.39, P=0.56, 3.24 ± 0.68 vs.
3.14 ± 0.56, P = 0.41, 2.21 ± 0.94 vs. 2.14 ± 0.83, P = 0.56, 18.1 ± 6.96 vs.
17.2 ± 6.81, P = 0.19,and 0.38 ± 0.29 vs. 0.38 ± 0.25, P = 0.93), whereas CAP
in 3D-VRFA-TSE was significantly higher than that of 3D-VRFA-TSE with C-SENSE
(3.38 ± 0.79 vs. 3.10 ± 0.68, P = 0.01) (Fig. 2, 3).
DISCUSSION:
There was no significant difference in
SNR, TNR, and image quality except for CAP between 3D-VRFA-TSE and 3D-VRFA-TSE with C-SENSE. These
results suggest that qualitative and quantitative assessment of 3D-VRFA-TSE
with C-SENSE are almost the same as that
of 3D-VRFA-TSE while reducing 40% in acquisition time and keeping isotropic
acquisition. CAP in 3D-VRFA-TSE with
C-SENSE was significantly lower than that of 3D-VRFA-TSE. It is assumed that
the decrease of clarity in prostatic capsule is due to blurring caused by denoising
in C-SENSE. Therefore, further optimization of C-SENSE factor and denoising
level in 3D-VRFA-TSE with C-SENSE will be needed to accurately assess ECE in
PC.
CONCLUSION: Our
study showed that C-SENSE can provide a reduction in acquisition
time of 40% when applied to 3D-VRFA-TSE MRI of the prostate and demonstrated
that there was no difference in the image quality and lesion conspicuity except
for clarity of prostatic capsule. Further clinical investigation and
optimization of imaging parameters is needed to assess if 3D-VRFA-TSE with
C-SENSE can consistently provide diagnostic performance comparable to 3D-VRFA-TSE.Acknowledgements
NoneReferences
-
Weinreb JC,
Barentsz JO, Choyke PL, et al. PI-RADS Prostate Imaging - Reporting
and Data System: 2015, Version 2. Eur Urol. 2016;70(5):e136.
- Gupta RT, Spilseth
B, Patel N, et al. Multiparametric prostate MRI: focus on T2-weighted imaging
and role in staging of prostate cancer. Abdom Radiol (NY). 2016;41(5):831-843.
- Tanaka U, Ueno
Y, Morinaga Y, et al. Value of three-dimensional
T2-weighted turbo spin-echo imaging with tissue-specific variable refocusing
flip angle for 3-T magnetic resonance imaging of prostate cancer: comparison
with conventional two- and three-dimensional T2-weighted turbo spin-echo
imaging. Jpn J Radiol. 2017;35(12):707-717.
- Sartoretti T, Sartoretti E, Wyss M,
et al. Compressed SENSE accelerated 3D T1w black blood turbo spin echo versus
2D T1w turbo spin echo sequence in pituitary magnetic resonance imaging. Eur J
Radiol. 2019 Nov;120:108667. doi: 10.1016/j.ejrad.2019.108667. Epub 2019 Sep
14.