Yiwen Qian1, Kai Ruppert1, Faraz Amzajerdian1, Yi Xin1, Hooman Hamedani1, Luis Loza1, Tahmina S. Achekzai1, Ryan J. Baron1, Ian F. Duncan1, Harrilla Profka1, Mehrdad Pourfathi1, Federico Sertic1, Stephen Kadlecek1, and Rahim R. Rizi1
1Radiology, University of Pennsylvania, Philadelhia, PA, United States
Synopsis
The fleeting nature of hyperpolarized xenon-129 magnetization
necessitates a separate flip angle calibration procedure based on the decay of
the gas-phase signal. This calibration is subsequently assumed to be valid for
imaging of the xenon dissolved in the lung tissue despite the rapid exchange
between the parenchyma and the alveolar gas volume and despite that the T2* is
much shorter in the former than in the latter. In this work we implemented a 1-dimensional
gas exchange simulation and investigated to what extent the effective flip
angle at the dissolved-phase frequencies is affected by exchange and
acquisition parameters.
Purpose
Unlike
in proton MRI, the fleeting nature of hyperpolarized xenon-129 (HXe) magnetization
necessitates a separate flip angle calibration procedure during which a series
of RF pulses is applied at the gas-phase frequency and the flip angle is
extracted from the decay of the gas-phase signal. This calibration is
subsequently assumed to be valid for imaging of the xenon dissolved in the lung
tissue despite the rapid exchange between the parenchyma and the alveolar gas
volume and despite that the T2* is much shorter in the former than in the
latter. In this work we implemented a 1-dimensional gas exchange simulation and
investigated to what extent the effective flip angle at the dissolved-phase
(DP) frequencies is affected by exchange and acquisition parameters.Methods
A 1-dimensional numerical
lung model was used for the simulation of xenon gas exchange between the
tissue-plasma (TP) and red blood cell (RBC) compartments comprising a single
alveolus (Fig. 1). At the start of the simulation one million xenon were
randomly distributed across the compartments in Fig. 1 such that 98% were in
the alveolar gas phase (GP) and the remainder in the DP. Random diffusional
motion was simulated at 1 ms time increments with diffusion constants of 0.68 10-5 m2s-1
in the GP and 0.33 10-9 m2s-1 in the DP, respectively. To maintain the
prescribed concentration ratio the probability for entering the DP was set to
5.7%. For each time increment the effect of a Gaussian RF pulse (duration 1-12
ms, flip angle 5°-180°) and B0-inhomogeneity induced T2* on the magnetization
vectors of the atoms were calculated. Four conditions were simulated: no gas
exchange, no T2* effects, both effects, or neither effect. gas exchange and T2*
and without gas exchange or T2*. Also, the center frequency of the RF pulse was
varied between 198 ppm (TP), 218 ppm (RBC) and 208 ppm (in between DP
resonances). To isolate the impact of exchange and T2*, the RF pulse was
assumed to leave the GP magnetization unaffected and no blood flow was
simulated. The effective longitudinal and transverse flip angles as a function
of the simulated nominal flip angles was calculated for all conditions.Results and Discussion
Figure 2 shows the impact of gas exchange and
T2* on the effectiveness of a Gaussian RF pulse centered at the TP frequency. Without
T2* and gas exchange the effective flip angles were identical to the nominal
ones (Fig. 2a). flip angles applied in the simulations when the Rf pulse was at
the tissue frequency (Fig 2a). When T2* is taken into consideration the effective
longitudinal flip angle is still fairly close to its nominal value but the
transverse flip angle diverges as a function of RF duration. Gas exchange (Fig.
2c) strongly affects longitudinal and transverse magnetization and both effects
combined compound the observed deviations. Figure 3 demonstrates that T2* and
gas exchange affect the effective flip angle at RBC and TP resonance
differently with larger deviations found for the TP resonance due to a more
direct impact of gas exchange between TP and GP. Conclusion
We demonstrated that gas exchange and T2*
strongly impact the effective flip angle for the DP magnetization. This effect
is of particular importance but largely underappreciated for precise measurements
of pulmonary gas exchange.Acknowledgements
Supported by NIH grants
R01 EB015767, R01 HL129805, S10 OD018203 and R01 CA193050. References
No reference found.