Gang Zheng1, Wai Tung Lee2,3, Xin Tong4, Tim D’Adam2, Steven Parnell5, Michael de Veer1, Shenpeng Li1, Hongxin Wang1, Richard Mcintyre1, Francis Thien6, and Gary Egan1
1Monash Biomedical Imaging, Monash University, Clayton, Australia, 2Australian Nuclear Science and Technology Organisation, Lucas Heights, Australia, 3Instrument Technologies Division, European Spallation Source ERIC, Lund, Sweden, 4China Spallation Neutron Source, Dongguan, China, 5Delft University of Technical, Delft, Netherlands, 6Eastern Clinical School, Monash University, Box Hill, Australia
Synopsis
We describe a custom-designed
spin-exchange-optical pumping (SEOP) polarizer which can reliably produce
hyperpolarized 129-Xe (HP-Xe) gas. Imaging of a HP-Xe phantom showed phase
images were less sensitive to depolarization and could provide on/off images of
HP-Xe. In a human lung, phase image provided higher imaging contrast in low
HP-Xe polarization states.
Introduction
HP-Xe can be rapidly depolarized by oxygen and radiofrequency
(RF) pulses during magnetic resonance (MR) imaging in lung1,2. When HP-Xe
flows into the lung, the distances of gas path are different in different
regions which could cause different depolarization of HP-Xe. It is difficult to
distinguish the signal loss on routine magnitude images is related with disease
or caused by depolarization. The aim of this study was to use the phase image
to image HP-Xe in a human lung.Materials and methods
129Xe gas is polarized by a customized spin-exchange-optical
pumping (SEOP) polarizer by batches rather than a
continuous-flow process. Its diagram is shown in Figure 1. The polarizer consists of a gas-polarizing unit and a gas
injection/recycle unit. To reduce the interactions, the two units were placed
in two different rooms (See Figure 2).
The gas-injection circuit sequentially flows Xenon and nitrogen into the OPC
and the ratio between them is 1:4. After polarizing for 30-40 minutes, the gas
is transferred from OPC to the transport cell by a piston
and immediately taken into MRI room.
All
MR experiments were performed on a Siemens 3T scanner (Skyra, Siemens
Medical Solutions, Erlangen, Germany) with a bird-cage 129Xe lung coil (RAPID
Biomedical GmbH, Wuerzburg, Germany). The resonance frequency of 129Xe was set
to 34.09 MHz. A 2D-GRE sequence was used for 129Xe gas imaging. For phantom
imaging, the parameters were as follows: flip angle = 7o,
repetition time = 15 ms, echo time = 3.24 ms, field of view = 400*400 mm2,
matrix size = 96*96, single slice, slice thickness = 25 mm, bandwidth = 370
Hz/Pixel, 4 averages and 4 repetitions. For human lung imaging, the parameters
were flip angle = 7o, repetition time = 20 ms, echo time = 7.81 ms,
field of view = 500*500 mm2, matrix size = 64*64, single slice, slice thickness
= 180 mm, bandwidth = 50 Hz/Pixel, and 6 averages.Results and discussions
Figure
3 shows the changes in the magnitude (Figure 3 first row) and phase (Figure 3 second row) images in a 129Xe
gas phantom. Figure 4 demonstrates the
corresponding image intensity and variance of the magnitude and phase images. The
image intensities of the magnitude images decreased following each repetition, while
that of phase images was nearly unchanged, supporting that the phase images
were less affected by depolarization induced by the RF excitation.
Figure 5 shows the magnitude
and phase images of a human lung in the axial plane. There was an increasing
signal intensity in the magnitude image from anterior to posterior (Figure 5A), which might associate with gravity-dependent
gradient in the human lung tissue density from anterior to posterior in the
supine position3. However, there was no such a change in the phase image (Figure 5B). The phase image was
computed by the arctangent of the ratio between the real and imaginary parts.
Since the real and imaginary signals originate from the same source, they have
similar signal and noise distributions which might effectively cancel both the gradient
and the noise. Because of such a cancellation, the phase image can work in
on/off mode to show HP-Xe in a low concentration.
The signal loss
in the anterior part of lung in the magnitude image (red arrows in Figure 5A) could be also caused by the
depolarization of HP-Xe. Phase image has the capability in imaging low-level
HP-Xe in the anterior part of lung (red arrows in Figure 5B). Conclusions
Phase image is less sensitive to the
depolarization, and can work in on/off mode to show whether or not there is
HP-Xe in lung. Measuring the phase signal has the potential for improved
sensitivity for HP-Xe imaging.Acknowledgements
We would like to acknowledge the supports
of the Australian Research Council (Grant LE130100035), the National Health and
Medical Research Council (Grant APP606944), and CASS foundation for the
polarizer and the Australian National Imaging Facility Fellowship.References
[1] Dregely I, Ruset IC, Wiggins G, Mareyam
A, Mugler JP 3rd, Altes TA, Meyer C, Ruppert K, Wald LL, Hersman FW. 32-channel
phased-array receive with asymmetric birdcage transmit coil for hyperpolarized
xenon-129 lung imaging. Magn Reson Med. 2013; 70(2):576–583.1
[2] Doganay O, Matin TN, Mcintyre A, Burns B, Schulte RF, Gleeson FV, Bulte D.
Fast dynamic ventilation MRI of hyperpolarized 129 Xe using spiral imaging. Magn Reson Med. 2018
May;79(5):2597-2606.
[3] Mugler JP, Altes TA. Hyperpolarized
129Xe MRI of the human lung. J Magn Reson Imaging. 2013; 37(2):313–331.