Joshua McAteer1, James Harkin2, Olivier Mougin1, Christoph Arthofer2, Paul Glover1, and Penny Gowland1
1Physics, University of Nottingham, Nottingham, United Kingdom, 2Medicine, University of Nottingham, Nottingham, United Kingdom
Synopsis
We investigated accelerating two paradigms of in vivo diaphragm imaging using a 0.5T
upright scanner. These two paradigms, high resolution and low resolution, were
optimised to deliver minimal artefacts and maximal information about the
diaphragm surface. These scans delivered sufficient image quality for the
tissue boundary of the diaphragm to be located. This will allow for future work
to study the diaphragm in the upright position of participants with COPD who
are only able to hold their breath for ~5 sec.
Introduction
Disorders of the diaphragm can have serious effects on health, but
people with serious respiratory conditions such as COPD can have severe problems
lying supine. Therefore it is necessary to study diaphragmatic function in a
seated position limiting the use of conventional MRI3. Our Paramed
0.5T MR Open scanner allows for participant positions including,
supine and sitting. However, the permanent magnet design and reduced gradient
performance necessitated by the open geometry, causes scan times to be longer
than on conventional scanners. This can limit the use of breath-hold imaging
particular in patients with respiratory diseases (for instance in severely
incapacitated participants with COPD, the breath-hold time is restricted to a
few seconds). In order to reduce the scan time to acceptable levels the field
of view or image resolution may be reduced, however, this prevents the
resolution of important structures due to wrap-around artefacts or poor PSF. Here we have implemented compressed sensing on an open MRI
scanner to access short imaging times for imaging respiratory function in short
breath-holds.Method
In compressed sensing, only a subset of the phase encode lines are sampled. In order to reconstruct the missing
points, assumptions are made about the MRI images1. Initially, a 3D GRE sequence (256x200x22) was targeted with a
TR/TE=15.4/8msec in order to image the diaphragm. This scan would take approximately
68sec, which is not possible within a breath-hold. An accel factor of 3
and a fully-sampled scan in the same time were acquired (~22 sec), as well as a
highly accelerated sequence at a lower resolution suitable for COPD patients with
severally reduced breathing capacity (5.1 sec).
Small datasets such as these are typically less sparse, and
so the maximum acceleration factors cannot be as high as for larger datasets.
Digital phantoms were used to design a sampling pattern, these have
different sparsity than the target anatomy (fig1). The sampling pattern was optimised to remove ghosting
artefacts, retain tissue boundaries, and to maximise SSIM and VIF. Variable density, elliptical Poisson disk sampling
was used, with additional points added-in uniformly, at random. This sampling
pattern was then checked by retrospectively undersampling fully sampled data
taken of the lung and diaphragm. The reconstruction was performed in BART2.
Three regularises were used: L1 sparsity in image and wavelet space, and TV. The regularisation level was optimised for each
dataset. The required sampling patterns were implemented on the
scanner via the underlying TechMag Redstone spectrometer. Phantom and in vivo
undersampled data was then acquired with 5 dummy lines before
each scan to allow the magnetization to reach a steady-state.Results
Fig2 shows results of the sampling
and reconstruction applied to an ACR phantom. Some ‘noise-like’ artefacts are
present, but key resolution is retained. This scan takes 22.6sec, which is within a breath-hold for healthy volunteers.
This sequence
was then applied in vivo on a
healthy volunteer (Fig3). Fig4 shows results from a fully sampled scan acquired
in the same time. The boundary of the diaphragm is more clearly defined in the
accelerated image, without the wrapping artefacts present in some slices of the
fully sampled image (an affected slice is outlined in red).
Fig5 shows the highly accelerated scan acquired in 5sec, at lower
resolution and with fewer slices. Key tissue boundaries are still preserved, and
could still be automatically detected (panel in red).Discussion
This initial work demonstrates the
usefulness of the MR Open system with compressed sensing for diaphragm imaging,
even in cases where participants are not able to complete a typical length
breath-hold. The ability to optimise the sampling and reconstruction process in
order to extract the key information from an acquisition is a useful tool for
future research.
With
few Kz points the likelihood of
ghosting in this direction is higher, the artefacts created by the sampling are
not ‘noise-like’, and so are not removed by the compressed sensing
reconstruction. However, the alternative is either increasing the voxel size,
or reducing the field of view. The optimal choice depends on the specific case,
however, in this scenario the fold-over artefacts in the fully sampled scan
obscure anatomy.
Several methods exist for accelerating MRI scans, however, the
receive coil available only has one channel preventing the use of SENSE/GRAPPA
acceleration.Conclusion
In future work, the sampling will be further optimized for the definition of the surface of the diaphragm in the images, and k-space filters
will be used to correct the approach to steady-state instead of using dummy
scans. We intend to image COPD participants in order to measure properties
of the diaphragm, in order to understand the effects of respiratory conditions
in specific individuals, where further acceleration may be required individually. In addition, to better quantify the respiratory
mechanics, we will be able to image at different points in the breathing cycle
and in different positions. Similar acceleration methods could also be applied to studies
involving participants who may not be able to remain still for long periods,
such as children, or increase motion robustness for imaging participants with tremors.
We are also interested in applying similar acceleration techniques to hyperpolarised
xenon imaging of the lungs of healthy volunteers.Acknowledgements
Medical Research Council
Engineering and Physicsal Science Research Council
Oxford-Nottingham Biomedical Imaging Doctoral Training Centre.
References
1. Lustig M,
Donoho D, Pauly JM. Sparse MRI: The application of compressed sensing for rapid
MR imaging. 1. Magn Reson Med. 2007 Dec;58(6):1182-95.
2. Uecker
M, Ong F, Tamir JI, et al. Berkeley advanced recon-struction toolbox. Proc Intl
Soc Mag Reson Med Toronto.2015;23:2486.
3. Arthofer C, Safavi S, Cooper A, Harkin J, Alenazi S, Gowland P, Boltan C,
Hall I. Impact assessment
of posture and breath-hold state on diaphragm shape. Magn Reson Med conference
paper. 2019