Kai Ruppert1, Faraz Amzajerdian1, Yi Xin1, Hooman Hamedani1, Luis Loza1, Tahmina S. Achekzai1, Ryan J. Baron1, Ian F. Duncan1, Harrilla Profka1, Yiwen Qian1, Mehrdad Pourfathi1, Federico Sertic1, Stephen Kadlecek1, and Rahim R. Rizi1
1Radiology, University of Pennsylvania, Philadelhia, PA, United States
Synopsis
Measurements
of the apparent alveolar septal wall thickness (SWT) with hyperpolarized
xenon-129 (HXe) MRI are sensitive to inflammatory or fibrotic pathologies in
the lung parenchyma but are also affected by lung inflation level. Here, we
investigated the dependence of such measurements on the choice of acquisition
parameters in a rabbit model. We found the SWT measurements to be strongly affected
by the imaging parameters: in particular, the number and flip angle of the
applied RF pulses centered at the dissolved-phase resonances. If this bias is
minimized, the previously reported dependence of septal thickness measurements
on lung inflation disappears.
Purpose
Measurements
of apparent alveolar septal wall thickness (SWT) obtained with hyperpolarized
xenon-129 (HXe) MRI are sensitive to inflammatory or fibrotic pathologies in
the lung parenchyma1-8 but are also affected by lung inflation level9
as well as physiological variations throughout the breath hold. Here, we
investigated the dependence of such measurements on the choice of acquisition
parameters in a rabbit model.Methods
Imaging
experiments were performed in sedated New Zealand rabbits (approx. 4 kg). Animals
were ventilated with room air until imaging began, at which point the gas mix
was switched to 20% oxygen and 80% HXe for 5 breaths (6 ml/kg tidal volume),
followed by an 8-s breath-hold at either end inspiration (EI) or end expiration
(EE). All studies were approved by the Institutional Animal Care and Use
Committee.
MR imaging was conducted using
either a 1D-projection gradient-echo sequence with
left-to-right frequency encoding or a chemical shift saturation recovery (CSSR)
spectroscopy sequence. The former employed a non-selective 700-μs Gaussian RF excitation pulse centered 3,530 Hz
downfield from the gas-phase resonance. The following sequence parameters were
used: matrix size 640 or 1920×80; TE 2.6 ms; FOV 220 mm; receiver bandwidth 120
Hz/pixel; flip angle 7-16°, TR 10-50 ms (TR90°,equiv 1.3 - 4 s10).
During the breath hold, the dissolved-phase magnetization was saturated 6 times
every 500 ms with 3 consecutive 3-ms frequency-selective Gaussian RF pulses
centered at 200 ppm and separated by 1.2 ms spoiler gradients. CSSR
measurements were conducted with either 8 or 40 delay times varying between 3 and
500 ms, following the same dissolved-phase saturation scheme. All other
parameters were as described by Qing et al3. Septal wall thickness
was calculated using the analytical uptake model of Patz et al.11, based
on the recovery of the averaged total dissolved-phase signal following each
saturation pulse. For a direct qualitative comparison between EI and EE, the
fitted analytical curves were normalized by the dissolved phase (DP) to gas
phase (GP) ratio at a delay time of 100 ms. All MR studies were performed at 1.5T (Avanto; Siemens) using a custom
xenon-129 transmit/receive birdcage coil (Stark Contrast, Erlangen, Germany).
Enriched xenon gas (87% xenon-129) was polarized using a prototype commercial
system (XeBox-E10, Xemed LLC, Durham, NH).Results and Discussion
Figure 1 illustrates the qualitative differences
in the fitted and normalized DP-GP ratios at EE vs EI for six acquisition
schemes. In a conventional CSSR acquisition with 40 sequentially increasing
delay times (Fig. 1a), a marked qualitative difference between EI and EE
measurements is noticeable, which translates into a significant difference in
the extracted SWTs (8.8 μm at EE, 12.7 μm at EI). However, if the CSSR study is repeated
using only 8 delay times (Fig. 1b), the EE and EI curves are beginning to
converge. Similar trends can be observed in the 1D measurements. In three
acquisitions with similar TR90°,equivs of approximately 1.3 s (Fig.
1c-e), the DP-GP ratio curves at EE and EI are converging as the excitation
flip angle decreases. Finally, if the TR90°,equiv is increased to 4 s
(Fig. 1f), the EE and EI uptake curves and the associated gas exchange
parameters become essentially indistinguishable. The key difference between the
six performed measurements is the total number of DP saturations (in the CSSR
studies) or the excitation flip angle (in the 1D studies). High flip angle RF
pulses applied to the DP magnetization result in a regional depolarization of
the GP magnetization due to gas exchange. This depolarization is therefore weighted
by the spatially variable gas exchange efficiency in the lung, which is
particularly large at EE. Thus, by preferentially destroying the GP
magnetization in fast exchanging lung volumes, the MR acquisition itself
imprints a time-varying bias onto the measured DP-GP ratio that is lung
inflation-dependent. If this bias is minimized by using fewer RF saturations
and/or decreasing the employed flip angles, however, the observed gas uptake
curves become qualitatively identical and the dependence of the SWT on lung
inflation vanishes. Sampling the entire uptake curve after each DP saturation,
as performed for the 1D acquisition, proved to be particular robust.Conclusion
We demonstrated the dependence of the observed
pulmonary xenon gas uptake on the MR acquisition parameters. Measurements that
employ high flip angles selectively destroy the GP magnetization in lung
volumes with high gas exchange. If this effect is minimized, the previously reported
SWT dependence on lung inflation disappears.Acknowledgements
Supported by NIH grants
R01 EB015767, R01 HL129805, S10 OD018203 and R01 CA193050. References
[1] Ruppert K et al. NMR
of hyperpolarized 129Xe in the canine chest: spectral dynamics during a
breath-hold. NMR Biomed 2000;13:220-228. [2]
Butler JP et al. Measuring surface-area-to-volume ratios in soft porous
materials using laser-polarized xenon interphase exchange nuclear magnetic
resonance. J Phys Condens Matter 2002;14:L297-L304. [3] Qing et al. Assessment of lung function in asthma and COPD
using hyperpolarized 129Xe chemical shift saturation recovery spectroscopy and
dissolved-phase MRI. NMR in Biomed 2014;27(12):1490-1501. [4] Zhong et al. Simultaneous assessment of both lung morphometry
and gas exchange function within a single breath‐hold by
hyperpolarized 129Xe MRI. NMR in Biomed 2017; 30(8). [5] Doganay et al. Quantification of regional early stage gas
exchange changes using hyperpolarized 129Xe MRI in a rat model of
radiation-induced lung injury. Med Phys 2016;43(5):2410-2420. [6] Kern et al. Regional investigation
of lung function and microstructure 129Xe chemical shift saturation
recovery parameters by localized and dissolved‐phase imaging: A
reproducibility study. MRM 2019; 81(1):13-24. [7] Kern et al. Mapping of regional lung microstructural parameters
using 129 Xe dissolved‐phase MRI in healthy volunteers hyperpolarized and
patients with chronic obstructive pulmonary disease. MRM 2018; 81(4):2360-2373.
[8] Zanette et al. Physiological gas
exchange mapping of hyperpolarized 129Xe using spiral-IDEAL and MOXE
in a model of regional radiation-induced lung injury. Med Phys
2018;45(2):803-816. [9] Stewart et
al. Reproducibility of quantitative indices of lung function and microstructure
from 129 Xe chemical shift saturation recovery (CSSR) MR spectroscopy. MRM
2017; 77(6):2107-2113. [10] Ruppert
et al. Assessment of flip angle-TR equivalence for standardized dissolved-phase
imaging of the lung with hyperpolarized 129Xe MRI. MRM 2019; 81(3):1784-1794. [11] Patz et al. Diffusion of
hyperpolarized 129Xe in the lung: a simplified model of 129Xe septal uptake and
experimental results. New J Physics 2011;13:015009.