Timothy J Allen1, Leah C Henze Bancroft2, Roberta M Strigel1,2,3, Ty Cashen4, Orhan Unal2, Frank R Korosec1,2, Ping Wang1, Lloyd Estkowski4, Fred Kelcz2, Amy M Fowler1,2,3, R Marc Lebel4, and James Holmes2
1Medical Physics, University of Wisconsin-Madison, Madison, WI, United States, 2Radiology, University of Wisconsin-Madison, Madison, WI, United States, 3Carbone Cancer Center, University of Wisconsin-Madison, Madison, WI, United States, 4Global MR Applications and Workflow, GE Healthcare, Madison, WI, United States
Synopsis
A SSFSE-based fast acquisition was
used in conjunction with a vendor-supplied, deep learning-based reconstruction to obtain T2-weighted images (Rapid
T2w+DL) for diagnostic breast MRI. Radiologists compared the Rapid T2w+DL images
to standard-of-care T2w FSE images assessing both image quality and utility in
completing a typical clinical task. The fast acquisition provided a reduction in
average scan time from 3:35 (min:sec) to 1:16 (64%). Deep learning was used to
alleviate issues associated with the fast protocol (e.g. blurring and reduced
SNR). Rapid T2w+DL images allowed for completion of the clinical task
with results comparable to those obtained with standard T2w images.
Motivation
T2w imaging is routinely used in diagnostic breast MRI to identify
benign features (e.g. benign cysts) and assist with differential diagnosis as
benign findings are more likely to be T2 hyperintense (i.e. fibroadenomas).
Historically, RARE acquisitions1 including fast spin echo (FSE) have been relatively
long due to the need for limited echo trains to minimize T2 blurring and long
TRs to allow T1 recovery. With increased desire to improve breast exam
efficiency, many clinical practices have modified their protocols to save scan
time, leading to compromised SNR or limited spatial resolution. Deep learning (DL)
provides an opportunity to deviate from the conventional paradigm of trading
off scan time, resolution, and SNR.
We propose using a faster
single-shot FSE (SSFSE) acquisition with a DL reconstruction to obtain T2w
breast images. Inherently, SSFSE produces images with increased spatial blurring
and reduced SNR. However, application of the DL reconstruction is expected to
alleviate these issues and produce images comparable to those obtained with a slower
conventional T2w FSE.Methods
Patients scheduled for standard clinical breast MRI exams were
approached to participate in this IRB-approved, HIPAA-compliant study. Participants
(N=7) received a standard-of-care breast MRI exam on a 3T scanner (SIGNA
Premier or Discovery 750w, GE Healthcare, Waukesha, WI) including a standard clinical
T2w FSE series (echo train: 8, 288x288 matrix, 2.0 mm slices, mean scan time:
3:35 min:sec). A rapid T2w acquisition was performed using SSFSE (288x288
matrix, 2.0 mm slices, mean scan time: 1:16). The SSFSE series were reconstructed
using a vendor-supplied, DL-based
reconstruction prototype (DLRecon) trained to reduce noise, enhance sharpness,
and remove simple artifacts. The combination of SSFSE with DLRecon is referred
to as Rapid T2w+DL.
A board-certified,
breast imaging fellowship-trained radiologist (radiologist A) reviewed the clinical exams and identified
features of interest including malignancies, cysts, and edema. Two additional board-certified radiologists (readers) reviewed
and scored features identified by radiologist A. First, the standard T2w and Rapid T2w+DL images were evaluated side-by-side using a five-point Likert scale to assess the SNR, artifacts, image
sharpness, and overall image quality (Table 1). A score of three or higher
indicated images were of sufficient quality to aid in diagnosis. Radiologists
were then shown contrast-enhanced T1w images with annotated features alongside
both T2w images. Readers rated the feature signal intensity (low=1 to high=5) on
both T2w images. Lastly, readers selected which T2w image they preferred considering
image quality and feature signal intensity task.
In addition to the clinical
patients, normal volunteers (N=3) were imaged with the standard T2w FSE and Rapid
T2w+DL protocol on a 1.5T scanner (SIGNA Artist, GE Healthcare, Waukesha, WI). Images
obtained at 1.5T have reduced SNR and are therefore apt to further benefit from
DLRecon. For 1.5T scanning, the slice thickness was increased to 2.4 mm and the
Rapid T2w+DL protocol used a 320x192 matrix to partially compensate for field
strength SNR differences. 1.5T images were reviewed by the readers and their image quality was scored.Results
Acquisition times for the
standard T2w scans were between 2:46 (min:sec) and 4:29 (mean: 3:35) and for
the Rapid T2w+DL were between 0:52 and 1:45 (mean: 1:16). Images from a typical
volunteer (Figure 1) demonstrate the denoising and sharpening of features provided by DLRecon. As shown in Figure 2, T2w images
and Rapid T2w+DL images from a patient with T2 bright cysts show comparable T2w
contrast. An example of a T1 bright finding is demonstrated in Figure 3.
A summary of results from
the review by the radiologists is shown in Figure 4.The mean brightness of
lymph nodes (N=9) in the standard T2w images was 4.4 compared to 4.5 in the
Rapid T2w+DL. Cysts (N=6) had a mean brightness of 4.2 and 3.7, respectively,
and known malignancies (N=2) 3.5 and 3.3, respectively. Single examples of additional
features were rated equally bright on each T2w image: hematoma, edema, seroma,
nonmass enhancement, and benign background enhancement.
Reader one rated the
overall quality of the Rapid T2w+DL as 3.1 at 3T and 3.3 at 1.5T. They rated
the overall quality of the standard T2w images as 3.7 at 3T and 4.0 at 1.5T (Figure
4). They preferred standard T2w images in eight out of the ten cases. Reader
two rated the overall quality of the Rapid T2w+DL as 4.7 at 3T and 5.0 at 1.5T,
while rating the standard T2w images as 4.0 regardless of field strength. They
preferred the Rapid T2w+DL images in 8 out of 10 cases. Discussion
Time reduction in the Rapid T2w+DL protocol could be highly
beneficial in a clinical environment, especially for sites using an abbreviated
breast MRI protocol. In our study, scores were similar for the feature-based
assessment and image quality in the T2w images and the Rapid T2w+DL images despite
the reduced scan time for the latter. Signal characteristics of the standard
T2w scan were preserved in the Rapid T2w+DL images. Both radiologists agreed
that both T2w protocols produced images of good quality providing initial
evidence that a SSFSE sequence with deep learning could be used as a substitute
for slower FSE scanning. Our preliminary results are encouraging, but further
study is necessary.Acknowledgements
This project was supported by the Departments of Radiology and Medical Physics, University of Wisconsin.References
1. J. Hennig; A. Nauerth; H. Friedburg (1986), "RARE imaging: A fast imaging method for clinical MR", Magn Reson Med (in German), 3 (6), pp. 823–833, doi:10.1002/mrm.1910030602