Hongchuang Li1, Ming Zhang1, Haidong Li1, Xiuchao Zhao1, Qiuchen Rao1, Xiaoling Liu1, Yeqing Han1, Xianping Sun1, Xinrui Wang2, Chaohui Ye1, Xin Lou2, and Xin Zhou1
1Wuhan Institute of Physics and Mathematics, National Center for Magnetic Resonance in Wuhan, Wuhan Institute of Physics and Mathematics, Innovation Academy of Precision Measurement Science and Technology, Chinese Academy of Sciences-Wuhan National Laboratory for Optoelectronics, Wuhan, China, 2Department of Radiology, Chinese PLA General Hospital, Beijing, P. R. China, Beijing, China
Synopsis
Lung compliance is commonly used in
clinical diagnosis of pulmonary fibrosis and emphysema, and it is generally
measured using pulmonary function tests (PFTs) in clinic. In this study, the
imaging methods including both hyperpolarized 129Xe MRI and micro-CT
were used to quantify lung compliance in rats, and the significant correlation was
found between these two imaging methods. Our study also demonstrates the
feasibility of measuring lung compliance using hyperpolarized 129Xe
MRI, which would contribute to fully understanding the lung function changes
caused by the pulmonary diseases.
Introduction
Lung compliance is generally driven by Compliance = ∆V/∆P,where ∆P is the change of pressure of the air way and ∆V is the change of volume of the air space in
the lung. Micro-CT has been reported to be used for measuring lung compliance
in rats1. Hyperpolarized
129Xe MRI is a non-invasively new method used in evaluating the
function and microstructure of the lung2, 3, and
it can obtain the ventilation map and absolute ventilated lung volumes. In the
previous study, hyperpolarized gas MRI has been used to measure the absolute
ventilated lung volumes in rats, and the results agree well with those by micro-CT4, 5.
However, the feasibility of hyperpolarized 129Xe MRI in measuring
lung compliance and its correlations with micro-CT are still unclear. In this study,
129Xe MRI and micro-CT were used to quantify lung compliance in
rats, and the correlation was also analyzed. Methods
Six Sprague Dawley rats were imaged using
hyperpolarized 129Xe MRI and micro-CT scan in sequence. Before the
imaging experiments, the rats were anesthetized with pentobarbital sodium
solution (1%) and endotracheal intubated using a 14-gauge tube catheter. Then,
they were quickly transferred to a home-built MRI‐compatible ventilator and
ventilated with a mixture of 96.6% O2 and 3.5% isoflurane for the
hyperpolarized 129Xe MRI. After that,the rats were transferred to another home-built micro-CT compatible
ventilator for micro-CT scanning. All the MR images were acquired using a 2D
Flash sequence on a 7 T animal MRI scanner (Bruker Bio Spec 70/20 USR) and a
home-built double-tuned birdcage RF coil was used. The corresponding imaging
parameters were as follows: field of view = 5×5 cm;
matrix size = 64×64; TR/TE =160.7/2.2 ms; number of
slices = 24. The MRI images were acquired during the breath-hold, and repeated
10 times for the breath-hold pressure in the air-space ranged from 0 to 15 cm H2O
by setting the inspiration time of 100, 150, 200, 250, 300, 350, 400, 450, 500
and 550 ms, respectively. Micro-CT (Bruker Skyscan1176) images were also
acquired using similar ventilation strategy, and four different pressures ranged
from 0 to 15 cm H2O were used. Results
Figure 1 showed the measured lung volume using 129Xe
MRI and micro-CT, and the volumes measured by hyperpolarized 129Xe
are obviously greater than those measured by micro-CT for all of the six rats with
the same pressure. However, the P-V curve shows consistent trend and the measured
lung compliance correlated well using both methods, as shown in the figure 2.Discussion and Conclusion
Lung compliance was measured using hyperpolarized 129Xe MRI
and micro-CT in rats and good correlation was found between the two imaging modalities.
The difference of measured ventilation volume using 129Xe MRI and
micro-CT was mostly because the different spatial resolution of the images and the
different segment technique were used.
In this study, the feasibility of hyperpolarized
129Xe MRI in measuring lung compliance was demonstrated, and the
measured lung compliance correlated well with that measured using micro-CT. Our
results indicate hyperpolarized 129Xe MRI has the potential in
evaluating the lung compliance and may be helpful in diagnosing lung diseases
that related to changing lung compliance.Acknowledgements
This work was supported by National Key R&D Program of China (2018YFA0704000), National Natural Science Foundation of China (81625011, 91859206, 21921004, 81601491), Key Research Program of Frontier Sciences, CAS (QYZDY-SSW-SLH018) and Hubei Provincial Natural Science Foundation of China (2017CFA013, 2018ACA143).References
- Ford
NL, McCaig L, Jeklin A, Lewis JF, Veldhuizen RA, Holdsworth DW, Drangova M. A
respiratory-gated micro-CT comparison of respiratory patterns in free-breathing
and mechanically ventilated rats. Physiol
Rep. 2017;5(2).
-
Li H, Zhang Z, Zhao X,
Han Y, Sun X, Ye C, Zhou X. Quantitative evaluation of pulmonary gas- exchange
function using hyperpolarized (129) Xe CEST MRS and MRI. NMR Biomed. 2018;31(9):e3961.
- Li H, Zhang Z, Zhao X,
Sun X, Ye C, Zhou X. Quantitative evaluation of radiation-induced lung injury
with hyperpolarized xenon magnetic resonance. Magn Reson Med. 2016;76(2):408-416.
-
Fox MS, Ouriadov A,
Santyr GE. Comparison of hyperpolarized (3)He and (129)Xe MRI for the
measurement of absolute ventilated lung volume in rats. Magn Reson Med. 2014;71(3):1130-1136.
- M Reza AS, Lam WW,
Ouriadov AV, Holdsworth DW, Santyr GE, %J Nmr in Biomedicine. Comparison of
hyperpolarized (3)He MRI rat lung volume measurement with micro-computed
tomography. 2010;23(4):359-367.