Faraz Amzajerdian 1,2, Stephen J Kadlecek1, Hooman Hamedani1,2, Yi Xin1,2, Ryan Baron1, Ian Duncan1, Maurizio Cereda1,3, Kai Ruppert1, and Rahim R. Rizi1
1Radiology, University of Pennsylvania, Philadelphia, PA, United States, 2Bioengineering, University of Pennsylvania, Philadelphia, PA, United States, 3Anesthesiology and Critical Care, University of Pennsylvania, Philadelphia, PA, United States
Synopsis
Xenon
Polarization Transfer Contrast (XTC) imaging is useful for quantifying the
exchange rates between gas-phase and dissolved-phase xenon, particularly in
emphysematic lungs which experience reduced local tissue and blood volumes.
However, this technique does not account for local gas volumes, which can
provide important context for interpreting these exchange rates. In this work,
a method of incorporating both gas exchange rates and local gas volumes is
described.
Introduction
Hyperpolarized
xenon-129 (HXe) is a powerful agent for evaluating lung function due to its
solubility in lung parenchyma and affinity for hemoglobin in red blood cells
(RBCs), analogous to trans-membrane oxygen exchange. While measurements of HXe
RBC binding are often acquired by imaging the dissolved-phase components of HXe
and normalizing to the gas-phase signal, such a scheme is unable to determine
the rate of exchange between the two compartments. Alternatively, gas exchange
can be imaged directly by observing the loss of gas phase signal after
selective saturation of the dissolved components; however, this approach yields
only a fractional exchange rate that fails to account for local gas volumes.
However, by combining a measure of gas exchange per time per unit gas volume
with voxel-wise measurements of gas volume, a robust, bidirectional
airspace-RBC exchange rate, q̇ (units of volume per time), can be defined. Methods
Imaging was performed, with
Institutional Review Board (IRB) approval, on one healthy volunteer and three
COPD subjects, including one bilateral lung transplant recipient. A 1.5T
scanner (Magnetom Avanto, Siemens) with a flexible 8-channel xenon-129 chest
coil (Stark Contrast, Germany) was used for all studies, and a prototype
commercial system (XeBox-E10, Xemed LLC, NH) was used to polarize 87% enriched
xenon-129.
A map of the exchange from gas to
RBC (per time per unit gas volume) was generated using the Xenon Polarization
Transfer Contrast (XTC) technique1. Three gas-phase images were
acquired via a 2D multi-slice GRE during a single 12 s breath-hold after
inhalation of pure HXe at a prescribed tidal volume (TV) of 15% of total lung
capacity (TLC). Imaging parameters included: flip angles of 4°, 9° and 13°,
respectively, TR/TE = 5.1/2.4 ms, matrix size = 36x48x6, 25 mm slice thickness
(20% inter-slice gap), and FOV of 300-340 mm. 100 frequency-selective inversion
pulses (Gaussian) with pulse width of 8 ms and TR = 30 ms were centered at 217
ppm (RBC resonance) for contrast preparation and repeated at -217 ppm to
correct for T1 decay and off-resonance excitation effects. In order to maximize
SNR, the contrast preparation scheme was applied between the second pair of
images, S2 and S3, while the first pair of images, S1
and S2, were used for the T1 correction. Images were normalized by
their respective flip angles prior to analysis.
The voxel-wise gas volumes (GV)
at end-inspiration were derived from fractional ventilation (FV) maps acquired
from a multi-breath imaging protocol2. This imaging was performed in
the same session and with the same slice positioning, resolution, and
orientation as the XTC imaging. The FV imaging series consisted of 6 breaths at
TV composed of 35% HXe diluted with nitrogen and 21% oxygen. Additional
parameters included a flip angle of 6°, TR/TE = 7.0/3.3 ms. A global measure of
GV, which was defined as TV plus the functional residual capacity (FRC), was
estimated from the calculated FV and the volume of delivered gas. A compartment
model was then used to iteratively divide the GV measurement down to voxel
scale.
$$q̇ = \frac{1}{TR}(1-(\frac{S_3S_1}{S_2^2})^\frac{1}{N}) *(GV)$$
Finally, q̇
was calculated by multiplying the GV by the fraction of HXe depolarized per
inversion pulse divided by the time allowed for exchange. This is shown in the equation above, where N and TR are the number and repetition time
of the inversion pulses, respectively. Results
Figure 1
shows the q̇ maps for a healthy subject, two COPD
III subjects (A, B), and one bilateral lung transplant recipient with COPD (C).
The healthy subject has a more homogenous distribution of exchange compared to
the patchy, heterogenous distribution in the COPD patients. This finding is
further supported by the histograms in figure 2, which illustrate these
different distributions of gas exchange. The skewness for the healthy subject as
well as subjects A, B, and C were 1.5, 3.5, 2.5, and 2.8, respectively. Table 1
lists the percent predicted FEV1, along with the DLCO and DLCO/VA, for each
subject. Decreases in these PFT values compared to the healthy control indicate
increased disease severity, which is mirrored by decreases in whole-lung q̇
measurements. The gas exchange rate for the entire lungs of the healthy subject
(1107.72 mL/s) was ~2.5 fold higher than in COPD subject A (451.83 mL/s) and
~8.5 fold higher than subjects B (131.08 mL/s) and C (125.01 mL/s). We note
that ‘gas exchange’ in this context refers to the rate at which inhaled gas
traverses the alveolar membrane and binds to hemoglobin, and vice versa; because
it describes an exchange process, this rate greatly exceeds that of gas uptake.
For this reason, and because of differing solubility and diffusion rates, these
values are not directly comparable to DLCO, VO2, or other standard metrics. Nonetheless,
we expect similar trends with loss of function and disease.Conclusion
By
incorporating ventilation in the form of gas volumes, the proposed
quantification method provides a more robust measure of gas exchange rates.Acknowledgements
No acknowledgement found.References
[1] Ruppert, Kai, et al (2000). Probing
lung physiology with xenon polarization transfer contrast (XTC). MRM 44(3),
349-357.
[2]
Hamedani, Hooman, et al (2016). Regional fractional ventilation by using
multibreath wash-in 3He MR imaging. Radiology 279(3), 917-924.