Qingdi Wang1, Xiaojing Guo1, Emma Hornsey2, Lucy McKenna2, Leonid Churilov1,3, Mark Brooks1,2, George Matalanis1,2, Jason Chuen1,2, Eric Poon1, Daniel Staeb4, Ning Jin5, Andrew Ooi1, and Ruth P Lim1,2
1The University of Melbourne, Melbourne, Australia, 2Austin Health, Heidelberg, Australia, 3The Florey Institute of Neuroscience and Mental Health, Melbourne, Australia, 4Siemens Healthcare Pty Ltd, Melbourne, Australia, 5Siemens Medical Solutions, Chicago, IL, United States
Synopsis
A
four-dimensional phase-contrast magnetic resonance imaging sequence with
respiratory-controlled adaptive k-space reordering (ReCAR-4DPC) offers
potential benefits of improved scan efficiency and motion robustness. Imaging was performed on 15 volunteers and 2
patients with aortic dissection. Inter-scan repeatability and inter-reader
agreement of flow metrics derived from ReCAR-4DPC was assessed and compared to
metrics derived from 2-dimensional phase contrast imaging (2DPC). There was
almost perfect inter-scan and inter-reader concordance (Lin's concordance
correlation coefficient for all metrics > 0.91 and >0.98 respectively).
Concordance with 2DPC was also high (LCCC all > 0.873). RC-4DFlow is reproducible
and repeatable, with high concordance with 2DPC metrics.
Introduction
Aortic dissection is a potentially fatal
disease and four-dimensional phase contrast imaging (4DPC) helps quantify
aortic haemodynamics for improved monitoring and treatment1. In the thoracic aorta, imaging
times can be long when performed with respiratory navigation. Respiratory-controlled
adaptive k-space reordering 4DPC (ReCAR-4DPC) can improve scan efficiency and
robustness to motion by acquiring the centre of k-space at end-expiration and
outer k-space during inspiration2. We evaluate the inter-reader, intra-reader and
inter-scan agreement of flow metrics within the thoracic aorta derived from ReCAR-4DPC
and compare these to reference standard 2-dimensional phase contrast imaging (2DPC)-derived
metrics. Method
15 healthy volunteers (9F, 6M, range 23-47y,
mean heart rate 71bpm) and 2 patients (2M, 74 and 79y, mean 68bpm) with aortic
dissection were recruited and imaged on a 3T MRI system (MAGNETOM Skyra,
Siemens Healthcare, Erlangen, Germany). All PC imaging was performed following
0.2 mmol/kg injection of gadobutrol (Gadovist, Bayer).
ReCAR-4DPC was performed in the oblique
sagittal orientation using a 4D flow prototype gradient echo sequence with retrospective
ECG-gating: TR/TE 4.85/2.3ms, temporal resolution 38.8 ms, FA 15°, voxel size
2.4 x 2.4 x 2.5mm3 (interpolated), 40 slices, FOV 380 x 285, GRAPPA
2, navigator acceptance window 8mm. ReCAR-4DPC was acquired twice in 14/17
subjects during the same examination. 2DPC breath hold conventional
phase-contrast gradient echo imaging was performed at the ascending aorta (AA),
descending aorta (DA) and diaphragm (DX): TR/TE 4.85/2.66ms, temporal
resolution 38.8ms, FA 20°, 1.4 x 1.4mm2 pixel size, 5mm slice
thickness, GRAPPA 3, TA~14s. 30 cardiac phases were reconstructed for both ReCAR-4DPC
and 2DPC. Velocity encoding (VENC) of 150-200cm/s (subject-dependent) was
selected for both ReCAR-4DPC and 2DPC, with VENC of 50cm/s in one patient with
low aortic velocities. ReCAR-4DPC scan time was recorded.
After training, Reader 1 (R1, no
experience with cardiovascular MRI) and Reader 2 (R2, 14 years’ experience) assessed
all anonymized ReCAR-4DPC datasets for peak systolic velocity (PSV), average
flow (AF) and net forward volume (NFV) using prototype software (4D Flow
Demonstrator, Siemens Healthcare, Erlangen, Germany) at AA, DA and DX, matching
2DPC acquisitions. R1 assessed all initial ReCAR-4DPC acquisitions twice to
assess intra-reader agreement, and R2 assessed both initial and repeat ReCAR-4DPC
acquisitions in 14 subjects for inter-scan repeatability. 2DPC metrics were
measured by R2 on a commercial workstation (Multimodality Workplace, Siemens Healthcare,
Germany), at least 1 month prior to ReCAR-4DPC reads.
Intra-reader agreement, inter-reader
agreement, inter-scan repeatability and concordance with 2DPC derived metrics (all
segments combined) was assessed with Lin’s concordance correlation coefficient
(LCCC) and reduced major axis regression, enabling analysis for both fixed and
proportional biases3. Results
ReCAR-4DPC and 2D PC imaging
acquisitions were successful in all subjects, with flow metrics able to be
derived from all subjects (representative images in Figure 1 and Figure 2). Average
ReCAR-4DPC acquisition time for 17 subjects was 9:57 min (range, 5:55–14:33
min), with mean±SD acceptance rate of 72.65±13.4%. Mean±SD of PSV, AF and NFV
derived at each level are provided in Table 1, with true lumen results used for
patients. For the two patients, analysis of the false lumen at DA revealed a mean
PSV of 47.14 ± 7.93 cm/s for ReCAR-4DPC and 65.41 ± 22.44 cm/s for 2DPC.
The reproducibility and 2DPC
comparison analysis for flow metrics are summarized in Table 2. There was excellent
intra-reader agreement (LCCC all ≥0.987) and excellent inter-reader agreement (LCCC all ≥0.979) for PSV, AF and NFV. There was high inter-scan
repeatability for PSV, AF and NFV (LCCC 0.965, 0.907 and 0.981 respectively).
There was near-perfect agreement
with 2DPC for PSV, AF and NFV (Table 2 and Figure 3). LCCC was 0.89 for AF, and 0.87 for both PSV and NFV. Reduced major
axis regression analysis demonstrated no proportional bias (slope close to 1)
and small fixed biases for PSV, AF and NFV, with slightly lower ReCAR-4DPC
values compared to 2DPC. Discussion
Accurate
and highly reproducible results were achieved with ReCAR-4DPC, within
clinically acceptable acquisition times of approximately 10 minutes. Whilst
more rapid techniques are in development2, 4, the employed ReCAR-4DPC prototype is easily implemented and does not
require intensive computational post-processing, lending itself readily to
clinical use. Small differences to 2DPC derived PSV, AF and NFV were observed,
similar to prior experience5. The slightly lower flow metrics observed with ReCAR-4DPC could relate
to lower in-plane spatial resolution leading to inaccuracies in luminal
segmentation, or slight differences in location/plane of AA, DA and DX samples
between the two techniques. Limited results from false lumen analysis were less
accurate, likely due to relatively low false lumen velocities.
Limitations of the study include small subject number, and relatively poorer
in-plane spatial resolution of ReCAR-4DPC, although our parameters conform to
consensus guidelines6. The selected VENC may not have accurately
captured patient haemodynamics, and a VENC scout could allow for improved VENC
selection. A dual-VENC approach may also enable more accurate measurement of
false lumen haemodynamics7.Conclusions
ReCAR-4DPC is an accurate and
relatively fast approach for comprehensively measuring the flow metrics throughout
the thoracic aorta, exhibiting high reproducibility and robust correlation with
conventional 2DPC. Recruitment of a greater number of subjects, including more
patients with aortic dissection for further technique validation, and
complementary CFD simulations8 are planned.Acknowledgements
This work was supported by funding
from the Royal Australian and New Zealand College of Radiologists.References
1. Clough
RE, Waltham M, Giese D, et al. A new imaging method for assessment of aortic
dissection using four-dimensional phase contrast magnetic resonance imaging.
Journal of vascular surgery. 2012 Apr 1;55(4):914-23.
2. Ma LE, Markl M, Chow K, et al. Aortic 4D flow MRI in 2
minutes using compressed sensing, respiratory controlled adaptive k‐space
reordering, and inline reconstruction. Magnetic resonance in medicine. 2019
Jun;81(6):3675-90.
3. Ludbrook J. Linear regression analysis for comparing two
measurers or methods of measurement: but which regression?. Clinical and
Experimental Pharmacology and Physiology. 2010 Jul;37(7):692-9.
4. Bollache E, Barker AJ, Dolan RS, et al. k‐t
accelerated aortic 4D flow MRI in under two minutes: feasibility and impact of
resolution, k‐space sampling patterns, and respiratory navigator gating
on hemodynamic measurements. Magnetic resonance in medicine. 2018
Jan;79(1):195-207.
5. Bollache E, van Ooij P, Powell A, et al. Comparison of 4D
flow and 2D velocity-encoded phase contrast MRI sequences for the evaluation of
aortic hemodynamics. The international journal of cardiovascular imaging. 2016
Oct 1;32(10):1529-41.
6. Dyverfeldt P, Bissell M, Barker AJ, et al. 4D flow
cardiovascular magnetic resonance consensus statement. J Cardiovasc Magn Reson
Imaging 2015; 17: 72.
7. Schnell S, Ansari SA, Wu C, et al. Accelerated Dual-venc 4D
flow MRI for neurovascular applications. J Magn Reson Imaging 2017; 46:
102-114.
8. Stalder AF, Liu Z, Hennig J, et al. Patient specific
hemodynamics: combined 4D flow-sensitive MRI and CFD. InComputational
Biomechanics for Medicine 2011 (pp. 27-38). Springer, New York, NY.