Mariana B. L. Falcão1, Lorenzo Di Sopra1, Liliana Ma2,3, Mario Bacher1,4, Davide Piccini1,5, Jérôme Yerly1,6, Peter Speier4, Tobias Rutz7, Milan Prša7, Michael Markl2,3, Matthias Stuber1,6, and Christopher Roy1
1Department of Diagnostic and Interventional Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland, 2Department of Radiology, Feinberg School of Medicine, Northwestern University, Chicago, IL, United States, 3Department of Biomedical Engineering, Northwestern University, Chicago, IL, United States, 4Siemens Healthcare GmbH, Erlangen, Germany, 5Advanced Clinical Imaging Technology, Siemens Healthcare AG, Lausanne, Switzerland, 6Center for Biomedical Imaging, Lausanne, Switzerland, 7Department of Cardiology, Lausanne University Hospital (CHUV), Lausanne, Switzerland
Synopsis
Conventional
4D flow MRI techniques often have prolonged and unpredictable scan times, due
to the use of respiratory navigation. To address this, a fully self-gated
cardiac and respiratory motion-resolved whole-heart 5D flow protocol with a fixed
scan time was recently developed using a free-running framework. This protocol
extracts cardiac and respiratory signals from periodic readouts (self-gating). In
this study, we explore the use of Pilot Tone signals as an alternative method
for cardiac and respiratory signal extraction to reconstruct 5D flow data and
compare reconstructions to those using the previously established self-gating method
and the conventional 4D flow sequence.
Introduction
Conventional
4D flow MRI techniques often have prolonged and unpredictable scan times, due
to the use of respiratory navigation. As a result, acquiring whole-heart 4D flow
images in a clinically acceptable scan time (<10min) is challenging. To
address this, a fully self-gated cardiac and respiratory motion-resolved whole-heart
5D flow protocol with a fixed scan time was recently developed using a
free-running framework[1,2,3]. This method employs a continuous,
non-ECG triggered 3D radial acquisition with phyllotaxis sampling[4]
and periodic readouts for cardiac and respiratory self-gating (SG)[2].
Pilot Tone signals (PT), acquired in parallel to the MR acquisition, have also been
recently proposed for cardiac and respiratory gating[5,6]. The PT Navigation system
consists of a transmitter that generates an alternating magnetic field with a frequency outside the band
occupied by the MR signal. The field received by the local MR coils is
modulated by motion[5,6]. Both
SG and PT potentially replace the ECG signal and navigator scans in 4D/5D flow imaging.
In this study, we explore the use of PT signals to reconstruct 5D flow data and
compare reconstructions to those using the previously reported SG method[3]
as well as a reference standard 4D flow sequence.Methods
One navigator
gated 4D flow sequence[7] covering the aorta and one prototype
free-running radial whole-heart 5D flow sequence[3] were used for
imaging in 11 healthy adult volunteers (age: 27.9±3.6 years) who provided written informed
consent on a 1.5T MAGNETOM Sola (Siemens
Healthcare, Erlangen, Germany), using a 12-channel body coil with an integrated
PT transmitter. Scan parameters were: field of view (4D: (200-292.8 mm) x (360-366
mm) x (75-137.4 mm), 5D: (220mm)3-(260mm)3); spatial
resolution (4D: 2.4x2.4x2.5 mm3, 5D: (2.3mm)3); temporal
resolution (4D: 38.6-57.9ms, 5D: 50ms); velocity encoding (4D/5D: 150cm/s). Both 4D flow and 5D flow were performed during
free-breathing and the ECG signal was measured during 5D flow scans for
subsequent comparison. The acquisition time of each sequence was also recorded.
For 5D flow,
cardiac and respiratory signals were retrospectively extracted from SG and PT signals
using the post-processing steps outlined in Figure 1. For both SG and PT, the cardiac
gating error was quantified as the standard deviation of the difference between
SG or PT triggers and the recorded ECG, normalized by the mean ECG RR-Interval[2].
After
signal extraction, the SG and PT 5D flow datasets were binned into 4
respiratory and 17-23 cardiac frames, depending on the subject’s heart rate, and
reconstructed using XD-GRASP[8]. For the 4D flow datasets and the
two 5D flow datasets (with SG and PT), flow curves were calculated in two
segments of the ascending aorta (AAo1, AAo2), one segment of the descending
aorta (Dao), and one segment of the aortic arch (Arch) using Siemens 4D Flow
v2.4. Additionally, the net flow volume and peak flow were computed and compared
using the Wilcoxon signed rank test.Results
The acquisition time
of the 5D flow sequence (7min 57s±15s) was shorter (p=0.08) and less variant than that
of the 4D flow protocol (10min 16s±3min 55s) across all subjects (Figure 2). The
mean and standard deviation of the cardiac gating error across volunteers were 2.9±1.3%
for SG and 3.3±1.4% for PT (Figure 2, p=0.08). Figure 3 shows flow curves for
the four aortic segments in four subjects, as well as flow streamlines
immediately after peak systole for 4D flow and 5D flow reconstructed with self-gated
and Pilot Tone gated signals. A flow animation for a full cardiac cycle using
the three reconstructions is depicted in Figure 4. The 3D hemodynamics of the
flow streamlines across the same subject were consistent, although some reduced
flow values were observed on both 5D flow reconstructions. Regarding the net flow
volume (Figure 5), there were no significant differences between 4D flow and either
SG or PT 5D flow. Nevertheless, compared to the 4D flow measurements, the peak flow
was significantly decreased for both SG and PT 5D flow for the lower segment of
the ascending aorta, the descending aorta and the aortic arch (p<0.05). Discussion
5D flow
scans provided whole-heart coverage in a predictable and shorter scan time,
whereas 4D flow suffered from variable efficiency due to heart rate and variable
navigator efficiency secondary to variability in breathing patterns. The cardiac
gating error was consistent between SG and PT and within the range of
previously reported values for SG[2], demonstrating the feasibility
of PT as an alternative method for cardiac and respiratory gating in 5D flow imaging.
While underestimation of the peak flow has been reported before[3]
and may be caused by temporal undersampling and regularization, further
investigation is required. Still, the temporal evolution of the flow curves and
agreement in net flow volumes suggests 5D flow using SG or PT is a promising
alternative to 4D flow, providing whole-heart coverage with matching temporal
and spatial resolution predictably in less than 10min.Conclusions
Pilot Tone
signals provide a valuable alternative to self-gating for ECG- and
navigator-free cardiac and respiratory motion-resolved 5D flow, while being
completely independent of the acquisition. As a result, the feasibility of
using Pilot Tone for signal extraction may open new opportunities for improving
flow acquisitions, including reduced scan times, and improved k-space sampling.Acknowledgements
No acknowledgement found.References
[1] S. Coppo et al. Free-running 4D whole-heart self-navigated golden
angle MRI: Initial results. Magnetic Resonance in Medicine 2015, 74(5),
1306–1316.
[2] L. Di
Sopra et al., An Automated Approach to Fully Self-Gated Free-Running Cardiac
and Respiratory Motion-Resolved 5D Whole-Heart MR Imaging. Magn Reson Med 2019,
In Press
[3] L. Ma
et al. 5D flow MRI: A free-running, fully self-gated, radial imaging framework
for cardiac and respiratory motion-resolved assessment of 3D blood flow
dynamics, Proc. Intl. Soc. Mag. Reson. Med. 27 (2019), 0090.
[4] D. Piccini et al. Spiral Phyllotaxis: The Natural Way to Construct a 3D Radial
Trajectory in MRI. Magnetic Resonance in Medicine 2011, 66, 1049–1056.
[5] P.
Speier et al. PT-Nav: A Novel Respiratory Navigation Method for Continuous Acquisition
Based on Modulation of a Pilot Tone in the MR-Receiver, Proc. ESMRMB 129:97-98,
2015.
[6] M.
Bacher et al. Retrospective Evaluation of Pilot Tone Based Cardiac Trigger
Quality In A Volunteer Cohort, Book of Abstracts ESMRMB 2017 30:360- 361.
[7] P.
Dyverfeldt et al. 4D flow cardiovascular magnetic resonance consensus
statement. Journal of Cardiovascular Magnetic Resonance 2015, 17(1), 1–19.
[8] L.
Feng et al. 5D whole-heart sparse MRI. Magn Reson Med 2017, 79(2), 826–838.