Grant S Roberts1, Kevin M Johnson1,2, Steven R Kecskemeti2, Ozioma Okonkwo3, Sarah Lose3, Laura Eisenmenger2, and Oliver Wieben1,2
1Medical Physics, University of Wisconsin - Madison, Madison, WI, United States, 2Radiology, University of Wisconsin - Madison, Madison, WI, United States, 3Medicine, University of Wisconsin - Madison, Madison, WI, United States
Synopsis
Pulse wave velocity (PWV) is a biomarker that indirectly relates
to arterial stiffness, an early indicator of cardiovascular disease. Breath-hold
phase contrast (PC) MRI can be used to assess PWV in the aorta, however in
certain populations, breath-holds may be difficult. We present a method to
measure aortic PWV using a free-breathing radially-undersampled PC sequence. Initial
results show that the free-breathing PC-derived PWV measures are comparable to
the measures obtained from breath-hold Cartesian PC scans. Larger cohorts are
warranted to verify these findings.
Introduction
Pulse wave velocity (PWV) is the rate at which pulse
pressure propagates through a vessel. PWV is clinically important because it indirectly
measures arterial stiffness, which is an early manifestation of vessel wall
remodeling from cardiovascular disease occurring well before luminal narrowing1. Developing biomarkers to
detect early vascular disease is clinically desirable for potential therapeutic
intervention. 2D cine phase-contrast (PC) MRI has been successfully used to non-invasively
determine aortic PWV by measuring temporal shifts (Δt) in flow waveforms between imaging planes and calculating the aortic centerline distance (Δx) between the
planes2, hence PWV=Δx/Δt. When
imaging the aorta, respiratory motion is avoided by performing breath-hold (BH)
exams (~10-15 seconds). However, in certain patient populations, BHs may be
difficult. Radially-undersampled MR sequences offer flexible tradeoffs for spatial/temporal resolution and in retrospective physiological gating, as well as decreased motion artifacts
due to oversampling of central k-space. The aim of this work is to assess the
feasibility of a free-breathing (FB) 2D radially-undersampled PC sequence by comparing
PWV measures to a standard Cartesian 2D PC BH sequence.Methods
Six healthy subjects (3F/3M, mean age=33y) were scanned with
both a radially-undersampled FB PCVIPR3 sequence and a Cartesian BH
product sequence at 3T (Signa Premier, GE Healthcare, Waukesha, WI) using a
30-channel anterior AIR™ coil and 60-channel posterior array. For both
sequences, two axial planes were prescribed in the aorta: one in the aortic arch and the
second in the abdominal aorta (Figure 1). Additionally, ungated free-breathing FIESTA images
were acquired to manually draw centerlines for distance calculations. 2D PC
Cartesian scans were acquired with prospective peripheral pulse-oximeter (PG) gating :
scan time=10-13s (breath-hold); TR=5ms; TE=3ms; flip=25˚;Venc=1500cm/s;
cardiac frames=40; temporal res.=41ms. 2D radial PCVIPR scans were acquired with
retrospective PG and retrospective respiratory gating: scan time=4:54;
projections=20,000; TR=7ms; TE=4ms; flip=25˚; VENC=1500; golden-angle sampling;
reconstructed frames=40; temporal res.=22-39ms (depending on heart-rate). To determine the minimal number
of projections needed, the radial PCVIPR scans were subsampled to 15000, 10000, 5000, 2500, and
1250 projections (corresponding to 3:41, 2:27, 1:14, 0:37, 0:18, and 0:09s scan
times).
A customized GUI in MATLAB (Mathworks, Natick, MA) was
developed in order to process the PC data (Figure 2). From the first PC plane
in the aortic arch, both ascending and descending aorta were transected (Figure
1A) thus providing two ROI data measurements. Circular ROIs were drawn around each
vessel and flow waveforms were constructed over all time frames. The
waveforms were further smoothed with a Gaussian filter (width 7 pixels) to
decrease velocity noise, as shown in Figure 4. Centerlines were calculated from
the anatomical FIESTA images by manually placing seed points in the aorta over
multiple image slices and fitting the points to a 3D b-spline (Figure 2B). Centerlines
and ROI locations were kept consistent over each subprojection group. Time-to-foot
(TTF)4, time-to-upstroke (TTU)5, time-to-point (TTPoint)6, and cross-correlation
(Xcorr)7 methods were used to
calculate time shifts in flow waveforms. Measured time shifts for each method
were plotted against measured centerline distances between planes and linear
regression was used to fit the 3 data points (Figure 1, bottom right) where the
inverse of the fitted slope is the PWV. The slopes of each regression were
averaged to create a composite PWV measure for each individual. PWVs for each
radial sub-sampling dataset were compared to the PWV measures from the BH
Cartesian scans using a paired Student’s t-test, with p<0.05 signifying
statistical significance.
Data
was checked to ensure the entire waveform was visible and not clipped by the pulse-oximeter lag.Results
Four of the six subjects were successfully scanned with 2
radial and 2 Cartesian PC scans in the aortic arch and abdominal aorta. One subject
was scanned with only 5,000 projections; the other subject was scanned with
only 1 plane through the aortic arch, both due to scanner acquisition errors.
Table 1 shows composite PWV measures obtained from linear regression of the 3 ROI
data points. None of the subsampled projection groups were significantly
different from the Cartesian PWV, however, there was an observed
bias between the Cartesian and radial acquisitions in some subjects (for
example, Subject 3) even when compared to heavily oversampled radial datasets.
Image quality was greatly diminished below 5,000 projections, however, the
velocity waveforms were not severely affected by subsampling, particularly
after Gaussian smoothing of the waveform. This is evidenced by the consistency of PWV
values with decreasing number of projections. Discussion
We demonstrated the feasibility of a free-breathing (FB) radial
PC sequence for PWV estimations. FB PWV assessment can be useful in severely
diseased or older populations who cannot hold their breath when evaluating global or regional vessel
stiffness and cardiovascular health. The observed bias in some subjects
requires further investigation. We believe it is due to a higher acquired
temporal resolution for the radial sequence and will conduct additional
measurements with Cartesian PC with higher temporal resolution.Conclusion
Based on preliminary results, pulse wave velocity measures
from the FB 2D radial sequence was comparable to that of the BH 2D Cartesian
sequence. Further studies are warranted to (1) increase statistical power and
(2) to investigate the observed bias between the Cartesian BH and the FB radial
acquisitions.Acknowledgements
We gratefully thank GE for their continued MR
research support.References
1. Cavalcante,
J. L., Lima, J. A. C., Redheuil, A. & Al-Mallah, M. H. Aortic stiffness:
Current understanding and future directions. J Am Coll Cardiol 57,
1511–1522 (2011).
2. Wentland, A. L., Grist, T.
M. & Wieben, O. Review of MRI-based measurements of pulse wave velocity: a
biomarker of arterial stiffness. Cardiovasc Diagn Ther 4, 193–206
(2014).
3. Gu, T. et al. PC
VIPR: A high-speed 3D phase-contrast method for flow quantification and
high-resolution angiography. AJNR Am J Neuroradiol 26, 743–749
(2005).
4. Mohiaddin, R. H., Firmin,
D. N. & Longmore, D. B. Age-related changes of human aortic flow wave
velocity measured noninvasively by magnetic resonance imaging. J Appl Physiol
74, 492–497 (1993).
5. Dogui, A. et al.
Measurement of aortic arch pulse wave velocity in cardiovascular MR: Comparison
of transit time estimators and description of a new approach. J Magn Reson
Imaging 33, 1321–1329 (2011).
6. Groenink, M. et al.
Biophysical properties of the normal-sized aorta in patients with Marfan
Syndrome: evaluation with MR flow mapping. Radiology 219, 535–540
(2011).
7. Fielden, S. et al. A
new method for the determination of aortic pulse wave velocity using
cross-correlation on 2D PCMR velocity data. J Magn Reson Imaging 27,
1382–1387 (2008).