Dan Zhu1,2 and Michael Schär2
1Department of Biomedical Engineering, Johns Hopkins University School of Medicine, Baltimore, MD, United States, 2Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, MD, United States
Synopsis
Velocity encoded
MRI has clinical significance in the diagnosis
of coronary artery disease. High spatial and
temporal resolution increase the accuracy and reproducibility and preserve
reliable flow pattern. By using rotated golden angle with k-t space reconstruction
and retrospective cardiac gating, high quality flow CINE could be achieved with
high spatial and temporal resolution during a single breath-hold. However, binning
to cardiac frames may lead to large gaps in k-space. In this work, we propose a
new scheme of rotated golden angle and optimized retrospective gating by shifting
the binning window. Improved image quality is demonstrated with the proposed
methods.
Introduction
Flow velocity measurement using phase contrast (PC)
MRI has been validated for clinical diagnosis of coronary artery disease1. High spatial resolution is required for high
accuracy and reproducibility2, while high temporal resolution better
preserves flow pattern, which is known to be affected by coronary diseases and
provide information on disease state3. Clinical assessed single breath-hold
approaches achieve either merely high spatial4 or high temporal3
resolution. A recently proposed method using golden angle (GA) rotated spiral sampling and k-t sparse parallel imaging
reconstruction (GASSP) has been evaluated on right coronary artery (RCA) with both high temporal and spatial resolution5. However, data binning of
spiral arms using retrospective cardiac gating may result in large k-space gaps
in certain CINE frames that may lead to image artifacts (Figure 1A). In this
study, we reset the rotation of the
spiral arms to the golden angle after each ECG trigger to minimize those k-space gaps and
optimize the position of the binning window to further minimize the gaps.
This approach enables to reject data at the end of
the acquisition for patients who cannot hold their breath for 20s, which is not
possible with segmented GA approaches that are otherwise very efficient in reducing the gap size6.Methods
Proposed
new scheme of rotating angle: In
the traditional scheme5, each spiral arm was applied to both
positive and negative flow encoding and then rotated in golden angle (i.e.
137.508°)7 for the next arm. 500 spiral arm pairs were
acquired in a single breath-hold of 20s.
In
this work, we proposed to use a new scheme, with flow encoding direction toggled
every heartbeat (using ECG detected R-wave trigger). The first arms after each R-wave
was rotated by half of golden angle (i.e. 68.754°) relative to the
first arm of the previous heartbeat. Within each heartbeat, consecutive arms
were rotated by 83.8°
so that arms of either positive of negative flow
encodings are rotated by the GA. Within each heartbeat, consecutive arms were
rotated by 83.8°.
Shifting the
binning window to minimize gap: Data
from each arm
were retrospectively binned into cardiac frames for both velocity encodes
according to ECG (without sliding window). Number of cardiac frames $$$n_{cf}$$$ was calculated by mean heartbeat duration over repetition time, TR = 19.2ms.
The largest spiral arm gap (max-gap)
of each cardiac frame and velocity encoding was determined (Figure 1). Mean-max-gap
was defined as the average of all max-gaps. This process war repeated 20-times:
Each time the retrospective binning window was shifted by TR/20. The shift
leading to the lowest mean-max-gap was used for image reconstruction. This was
performed for both the traditional and proposed angle rotation schemes. For
illustration, the shift leading to the largest mean-max-gap was applied as
well.
Simulation: The compare the gap size, both traditional and new
schemes were computed with simulated heart rate (HR) ranging of 40 to 100 beats
per minute. The simulated HR randomly changes RR interval with a normal
distribution, with mean value of 60/HR (in seconds) and standard deviation of 50ms8. Computation for each HR was repeated 100 times.
In-vivo Measurements: Flow data was acquired in the popliteal artery of 4
subjects (age 26-58, 2 females) and in both right coronary artery (RCA) and
left anterior descending artery (LAD) of 3 subjects (age 24, 40, and 56, 2
female) in a 3T Philips Scanner (Achieva, Philips Healthcare) using a
32-channel cardiac array. ECG signal was collected during each scan. Data with
both the traditional scheme (GA-old) and the proposed scheme (GA-new) were
acquired. Both were conducted with 2D gradient echo and two-sided velocity
encoding in slice direction. Parameters: Slice Thickness = 8mm, field of view
350mm, and voxel size (0.8mm)2, TR/TE=19-20/2.7ms, scan time 20s. Imaging
plane was planned orthogonally to target arteries on scout images.
The binned data were reconstructed
with phase sensitive k-t sparse reconstruction that jointly reconstructs images
from both velocity encoding.
Data analysis: Flow-encoded images were combined with geometric
mean to generated flow-compensated images, which is then utilized in deblurring9
and segmentation process. Segmentation of coronary arteries was semi-automatically drawn on
Medis platform with Qflow 8.1 application (Leiden, Netherlands).RESULTS AND DISCUSSION
Fig 2 a) shows the decrease in mean-max-gaps could be of all
simulated heart rate. Mean-max-gap of in-vivo cases demonstrated in Fig 2
b) indicates that new GA schemes significantly reduce mean-max-gap.
Fig 3 a, c) demonstrate a large gap in binned data
with the best shifted binning window using old scheme leading to patchy
artifacts. By using new scheme, the gap is largely reduced and the image
quality is greatly improved Fig 3 b, d).
Flow curve shown in Fig 3 e) indicate that the flow curve with new
scheme and the best shifted binning window has higher flow peak and smaller
fluctuation.
Animated Gif showing of the magnitude and through-plane
velocity shown in Fig 4 demonstrate high spatial and temporal resolution and
improved image quality with the new GA scheme.
RCA flow curve of old and new schemes shown in Fig
5 agree well.
CONCLUSION
With the proposed new scheme of rotating angle and
shifted binning window, the k-space gaps are reduced and image quality is more
stable.
Acknowledgements
No acknowledgement found.References
1. Ofili EO, Labovitz AJ, Kern MJ. Coronary
flow velocity dynamics in normal and diseased arteries. Am J Cardiol.
1993;71:3D–9D
2. Johnson K, Sharma P,
Oshinski J. Coronary artery flow measurement using navigator echo gated phase
contrast magnetic resonance velocity mapping at 3.0 T. J Biomech 2008; 41:
595–602.
3. Keegan J, Raphael CE,
Parker K, et al. Validation of high temporal resolution spiral phase velocity
mapping of temporal patterns of left and right coronary artery blood flow
against Doppler guidewire. J Cardiovasc Magn Reson 2015;17:85.
4. Brandts A, Roes SD,
Doornbos J, Weiss RG, de Roos A, Stuber M, Westenberg JJM. Right coronary
artery flow velocity and volume assessment with spiral K-space sampled
breathhold velocity-encoded MRI at 3 tesla: Accuracy and reproducibility. J
Magn Reson Imaging. 2010;31:1215–1223.
5. Zhu D, Bonanno G, Weiss
RG, et al. Phase contrast coronary blood
velocity mapping with both high temporal and spatial resolution using Golden
Angle rotated Spiral k-t Sparse Parallel imaging (GASSP). In Proceedings of the
27th Annual Meeting of the ISMRM, Montreal, Canada, 2019:1591.
6. Han F, Zhou Z, Rapacchi S, Nguyen KL,Finn JP, Hu P. Segmented golden ratio radial reordering with variable temporal resolution for dynamic cardiac MRI. Magn Reson Med 2016;76:94–103.
7. Winkelmann S,
Schaeffter T, Koehler T, Eggers H, Doessel O. An optimal radial profile order
based on the Golden Ratio for time-resolved MRI. IEEE Trans Med Imaging. 2007;26:68–76.
8.
Nunan
D. et al. A Quantitative Stematic Review of Normal Values for Short-Term Heart
Rate Variability in Healthy Adults. Pace 33(11):1407-17, 2010
9. Noll DC, Pauly JM,
Meyer CH, Nishimura DG, Macovski A. Deblurring for non-2D Fourier transform
magnetic resonance imaging. Magn Reson Med. 1992;25:319–333.