Sujata M. Shanbhag1, Rajiv M. Ramasawmy1, W. Patricia Bandettini1, Christine M. Mancini1, Delaney R. McGuirt1, Jennifer L. Henry1, Margaret M. Lowery1, Marcus Y. Chen1, and Adrienne Campbell-Washburn1
1NATIONAL INSTITUTES OF HEALTH/NHLBI, BETHESDA, MD, United States, Bethesda, MD, United States
Synopsis
Low-field MRI equipped with high-performance technology and modern imaging methods offers
the potential to make technically demanding cardiac imaging more accessible. This
study demonstrates that a low-field (0.55T) system can provide accurate phase-contrast
flow measurements in both healthy volunteers and patients who were clinically
referred for assessment of intracardiac shunt and valvular disease. Comparative
studies between contemporary high-performance 0.55T and 1.5T systems generated
well-correlated quantitative flow measurements (r>0.87).
Introduction
High-performance
low-field imaging technology is attractive for cardiovascular MRI due to the lower costs, favorable imaging properties and improved device safety [1, 2]. We have developed a 0.55T system with
retained hardware performance and modern imaging methods, which may be well
suited for the demands of rapid cardiac imaging [3]. In cardiovascular MRI, 2D phase-contrast flow measurements are routinely used for the clinical assessment of valvular
heart disease, vascular flow assessment and intracardiac shunts [4]. In this study, we evaluate the
quantification of pulmonary and aortic flow in normal volunteers and in
patients referred for clinical cardiac MRI evaluation and compare flow
quantification to conventional imaging performed at 1.5 Tesla.Methods
Eighteen subjects
underwent paired examinations using a 1.5T system (MAGNETOM Aera, Siemens
Healthcare, Erlangen, Germany) and prototype 0.55T system (MAGNETOM Aera,
Siemens Healthcare, Erlangen, Germany). Both healthy subjects and clinically-referred
patients were included. Clinical
indications for MRI exams included assessment of valvular regurgitation and intracardiac
shunts. The study was approved by our local Institutional Review Board and
informed consent was obtained on all subjects and volunteers. Flow was measured
using retrospectively gated, free-breathing gradient-recalled echo (GRE)
phase-contrast imaging. Imaging parameters are provided in Table 1. Bandwidth
was reduced and the flip angle was increased to compensate for lower SNR at
0.55T. Phase-contrast using a single directional through-plane velocity-encoded
gradient was performed for the assessment of the total pulmonary flow output (Qp)
through the main pulmonary artery (MPA) and the total systemic flow output (Qs)
through the ascending aorta (Ao). The ratio of pulmonic-to-aortic flow was
calculated to assess for the presence of a hemodynamically significant
left-to-right intracardiac shunt (Qp/Qs >1.5). Correlation coefficient and Bland-Altman
analyses were performed to determine the level of agreement between paired
measurements. The coefficient of variation was also calculated to determine the
reproducibility of each set of flow measurements.Results
Eighteen subjects (mean age 43 ± 17 years, 11
males) were enrolled in this comparison.
Ten of the 18 were healthy subjects, and the remaining eight were
clinical patients referred for evaluation of a valvular or shunt issue. Figure 1 illustrates the comparison of aortic and pulmonic flow assessment in a healthy
volunteer. Figure 2 provides comparison
images of a patient with aortic valve insufficiency, where a diastolic flow jet
is visible. When all subjects (pathology and healthy volunteers) were evaluated
together, Qp and Qs measured at 0.55T demonstrated high correlation with the
measurements performed at 1.5T (R = 0.89 and R = 0.87, respectively; p<0.0001 for both).
Qp/Qs also demonstrated high correlation for the total group (R = 0.89,
p<0.001) (Figure 3). Bland-Altman
analysis (Figure 4) showed reasonable agreement for flow parameters (Qp and Qs)
between 1.5T and 0.55T MRI systems. The coefficient of variation for flow
parameters was 7.1% for Qp, 10.5% for Qs, and 5% for Qp/Qs. Discussion
In our study, phase-contrast
flow assessment using a high-performance 0.55T system was comparable to a
conventional 1.5T system. Previous studies have compared flow measurements
using 0.35T and 3T in healthy volunteers [5], whereas our study included both
healthy volunteers and clinical patients with valvular regurgitation and
suspected intracardiac shunts, to assess the diagnostic capabilities of the
0.55T MRI system. Paired flow measurements showed a high degree of correlation
and close agreement for both aortic and pulmonic flow measurements. No phase
non-linearities caused by concomitant gradients were observed in phase-contrast
flow measurements at low field. Image quality was preserved, and the coefficient of
variation was within the range between 5-10.5%, demonstrating the suitability
of a low-field system for diagnostic cardiovascular MRI. Acknowledgements
We would like to acknowledge the assistance
of Siemens Healthcare in the modification of the MRI system for operation at
0.55T under an existing cooperative research agreement (CRADA) between NHLBI
and Siemens Healthcare.References
1. Rashid, S., et al., Cardiac balanced steady-state free
precession MRI at 0.35 T: a comparison study with 1.5 T. Quant Imaging Med
Surg, 2018. 8(7): p. 627-636.
2. Simonetti,
O.P. and R. Ahmad, Low-Field Cardiac
Magnetic Resonance Imaging: A Compelling Case for Cardiac Magnetic Resonance's
Future. Circ Cardiovasc Imaging, 2017. 10(6).
3. Campbell-Washburn,
A.E., et al., Opportunities in
Interventional and Diagnostic Imaging by Using High-performance
Low-Field-Strength MRI. Radiology, 2019: p. 190452.
4. Nayak,
K.S., et al., Cardiovascular magnetic
resonance phase-contrast imaging. J Cardiovasc Magn Reson, 2015. 17: p. 71.
5. Varghese,
J., et al., Cine and flow imaging at 0.35
T and Comparison to 3T. SCMR, 2019: p. P238.