Anna Padée1, Lorenzo Di Sopra2, John Heerfordt2,3, Jérôme Yerly2,4, Marco Merlo1, Tobias Kober2,3,5, Davide Piccini2,3,5, Matthias Stuber2,4, Christopher Roy2, and Jonas Richiardi2,3
1Laboratory for Psychiatric Neuroscience and Psychotherapy, University of Fribourg, Fribourg, Switzerland, 2Department of Radiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland, 3Advanced Clinical Imaging Technology, Siemens Healthcare, Lausanne, Switzerland, 4Center for Biomedical Imaging, Lausanne, Switzerland, 5LTS5, Ecole Polytechnique Fédérale de Lausanne, Lausanne, Switzerland
Synopsis
ECG recording during whole heart MR imaging requires extra setup time
and is not always reliable. Recently, self-gated (SG) acquisition
techniques have been developed, which extract the cardiac signal from
k-space center amplitude modulations. Here, we investigated whether
different single- (Empirical Mode Decomposition (EMD), Ensemble EMD
(EEMD)) and multi-coil (PCA, ICA with novel stabilisation)
decomposition methods yielded SG triggers with minimal variability to
ECG R-wave location. EEMD yielded median variability similar to PCA.
Compared to PCA, our proposed stabilised ICA approach yielded SG
triggers with 66% lower variability (median over 5 subjects),
although not for all subjects.
INTRODUCTION:
Motion-resolved imaging1,2 of the heart depends on
reliable physiological gating signals. ECG-triggering is commonly
used for synchronising the MR acquisition to the cardiac cycle.
Recently, self-gating (SG) methods have shown promise for simplifying
the setup by instead extracting the cardiac motion information
directly from the acquired imaging data3. Approaches for
extracting self-gating signals have mainly rely on principal
component analysis (PCA)1,2,4, or independent component
analysis (ICA)1. However, PCA and ICA assume linearity and
stationarity of the analysed time-series, which may not be realistic
in cardiac imaging due to e.g. non-sinus rhythm, and ICA runs can
yield very different components5. Therefore, we
hypothesize that decomposition methods with less restrictive
stationarity assumptions or more stability may result in improved
cardiac SG signal extraction. Given varying coil distances to the
myocardium, it is also uncertain whether all coils are needed to
recover an SG signal, and therefore we explored both univariate
methods (single-coil) and multivariate methods (multi-coils).METHODS:
Datasets:
Single
breath-hold (scan time=28s) 3D cine datasets were acquired in N=5
healthy volunteers on a 1.5T clinical MRI scanner (MAGNETOM Aera,
Siemens Healthcare) using a free-running 3D golden angle radial
bSSFP6,7 sequence. An ECG was simultaneously recorded for
analysis purposes.
The
signals of interest were obtained by selecting the absolute value of
the central coefficient of periodically acquired k-space readouts (K0
modulation).
Decomposition approaches:
We
tested single-coil and multi-coil decompositions. Single-coil
decomposition were performed on all coils sequentially.
Empirical
mode decomposition (EMD), a single-coil method, is particularly
suited for non-stationary signals, but can be affected by mode
mixing. Thus, we also used ensemble EMD (EEMD), which improves over
EMD by running the decomposition multiple times with added white
noise, turning EMD into an adaptive dyadic filter bank. Averaging
over the runs makes the method less sensitive to the perturbations to
data8. For
multi-coil decompositions, we used the ICASSO principle5 with FastICA9
implementation to obtain 12 components over 200 runs of ICA. We then clustered all components into
12 clusters (Figure 3) using hierarchical clustering with a Ward
variance minimization algorithm.
We propose two automated approaches
to picking a decomposition from the multiple ICA runs.
In
the centrotype ICA approach, we estimated the sources by
averaging all components within each cluster. This centrotype was
then used to compute “pseudo sources” from the original signals,
with no independence guarantee.
In the proposed second approach, stabilised ICA (StabICA), we
selected the most reliable decomposition based on a cluster quality
index Iq: the difference between average
within-cluster(Cm) similarities σi,j and average
between-cluster(C-m) similarities: $$ I_q(C_m) =\frac{1}{|C_m|^2}\sum_{i,j\in C_m} \sigma _{i,j} -\frac{1}{|C_m||C_{-m}|}\sum_{i\in C_m}\sum_{j\in C_{-m}} \sigma _{i,j}\text{; where }C_{-m} = C - C_m $$
These
methods were compared to PCA as a baseline.
Automatic SG component selection:
For
each dataset, the subject-specific heartrate was extracted using FFT
as the frequency with maximum power in the 0.5-2Hz band. This
frequency was later used to automatically identify the components of
interest, by selecting the component with the highest power spectral
density in the ±0.3Hz band around
heart rate.
Evaluation metrics:
The
timing differences between peaks of the SG signal and triggers
obtained from ECG R-waves were computed for each of the around 25
heartbeats per dataset.
The distance
between SG peaks varies, due to delay between the electrical activity
and heart motion, ECG filtering and sign indeterminacy of ICA. Thus,
we used standard deviation of the distribution of timing differences
as the evaluation measure.
The
distributions of standard deviations were compared subjectwise
between methods using the robust Bonett-Seier
test of scale for paired samples10.RESULTS:
Most
methods were able to recover a component representing the cardiac
cycle. Specifically, StabICA,
Centrotypes ICA, PCA, and EEMD recovered all triggers in 4 out of 5
subjects, while EMD recovered all triggers only in 2 cases. No method
recovered all triggers for subject 4 (see Figure 3).
The median standard deviation of the differences between SG triggers
and ECG triggers across the 5 datasets were: StabICA: 13.2 ms,
centrotypes: 12.5 ms, PCA: 37.8 ms,
EMD: 61.3ms and EEMD: 36.9 ms. Figure 2 shows standard deviations for
the various decomposition of one dataset.
Two subjects had significantly lower variability with StabICA than
PCA (p=0.005, p<10-16), one had lower variability but
no significant difference (p=0.25), and two had higher variability
with StabICA. Figure 3 shows results for all subjects.
Qualitative
examination of ECG, PCA, and StabICA reconstruction for one subject
showed no visible quality difference (Figure 5).DISCUSSION/CONCLUSION:
Stabilized
ICA and Centrotypes ICA showed lower variability than other
decomposition methods. This supports our hypothesis that fewer
assumptions might provide more “ECG-like” cardiac self-gating, as
gating based on the ECG signal itself has no such assumption.
However, the ECG as a reference standard is not ideal due to missed
triggers or falsely identified triggers caused
by MHD effects. More generally, our
proposed approaches could improve over vector gating, which requires
manual selection of the ECG vector and constant monitoring for missed
triggers.
Image quality on an example subject was not qualitatively different
between SG decomposition methods, and future work includes expanding
the analysis to more subjects including patients with arrhythmia,
varying sequences, as well as solving sign indeterminacy of
decompositions to further improve consistency.Acknowledgements
No acknowledgement found.References
1.Di Sopra
L, et al. An automated approach to fully self‐gated
free‐running cardiac and respiratory motion‐resolved 5D
whole‐heart MRI. Magn. Reson. Med. 2019;82:2118–2132
2. Feng L, et
al. 5D whole-heart sparse MRI. Magn. Reson. Med.
2018;79:826–838 doi: 10.1002/mrm.26745.
3. Larson
A.C, et al.
Self-Gated Cardiac Cine MRI Magnetic Resonance in
Medicine 51:93–102 (2004)
4. Pang J, et
al. ECG and navigator-free four-dimensional whole-heart coronary
MRA for simultaneous visualization of cardiac anatomy and function.
Magn. Reson. Med. 2014;72:1208–1217 doi: 10.1002/mrm.25450.
5. Himberg
J, Hyvärinen A.
Icasso: software for investigating the
reliability of ICA estimates by clustering and visualization.
In Proc. 2003 IEEE Workshop on Neural
Networks for Signal Processing (NNSP2003),
pp. 259-268, Toulouse, France, 2003
6. Piccini
D, et al.
Four-dimensional whole-heart cine
imaging with isotropic resolution in one single breath-hold: initial
results. Proc. ISMRM 2019.
7. Coppo
S,
et
al. Free‐running
4D whole‐heart self‐navigated golden angle MRI: Initial results.
MRM 2014
8. Zhaohua W,
Norden E. Ensemble Empirical Mode Decomposition: a Noise-Assisted
Data Analysis Method Advances in Adaptive Data Analysis, Vol.
1, No. 1 (2009)
9. Hyvärinen
A, Oja E. Independent Component Analysis: Algorithms and
Applications, Neural Networks, 13(4-5):411-430, 2000
10. Bonett D.G, Seier E. Statistical
inference for a ratio of dispersions using paired samples.
Journal of Educational and Behavioral Statistics, 28, 21-30, 2003