Xin Dong1,2, Olumide Awelewa 1,2, Graham Galloway2, Viral Chikani3, Rob Robergs1, and Arnold Ng4,5
1Queensland University of Technology, Brisbane, Australia, 2Translational Research Institute, Woolloongabba, Australia, 3Endocrinology, Princess Alexandra Hospital, Woolloongabba, Australia, 4Cardiology, Princess Alexandra Hospital, Woolloongabba, Australia, 5The University of New South Wales, Sydney, Australia
Synopsis
VARPRO
is a novel fat water separation sequence that shows promise for myocardial lipid
characterization. Two questions were addressed in this study:
1. Does VARPRO provide the same myocardial
fat-fraction result compared to established magnetic resonance spectroscopy (MRS)?
2. Is the myocardial fat-fraction
obtained by VARPRO reproducible?
The
results suggest the two methods are strongly associated but agreed poorly. The
VARPRO estimation is proportionally biased, however, the bias can be easily
accounted for using a linear equation. The temporal reproducibility of VARPRO
is satisfactory.
VARPRO
could be a potential, alternative modality to MRS in myocardial lipid
quantification.
Introduction
Myocardial
steatosis increases the risk of developing heart failure, even in the absence
of ischaemia1-3.
Monitoring the myocardial fat-fraction could improve the understanding of
cardiac metabolism, and provide a potential risk assessment for patients. Currently,
proton magnetic resonance spectroscopy (MRS) is the only non-invasive technique that provides
an evaluation of the intramyocardial lipid content 4. Yet, the application of MRS is limited clinically
due to technical and practical constraints3, 4. A novel pulse sequence to provide robust
water/fat separation has now been developed based on multi-echo Dixon
acquisitions with iterative variable projection estimation (VARPRO, Work-in-progress#724C,
Siemens). The images are reconstructed using graph cut algorithm5, 6.
VARPRO produces water, fat, and T2*maps simultaneously in a single breath-hold.
It shows promise as a simple and efficient alternative to MRS for myocardial lipid
characterisation. The objectives of this study were: 1, to compare the accuracy of myocardial
lipid evaluation using VARPRO and established MRS; 2, to evaluate the
reproducibility of using VARPRO to characterise myocardial fat-fraction.Methods
Institutional
ethics approval and written informed consent were obtained.
Two
cohorts of participants were recruited to evaluate the accuracy and
reproducibility respectively.Research subjects
Cohort
1 comprised thirty healthy volunteers (n=30; F=9; mean age=37.5±10.5 years;
mean BMI=24.9±3.2 kg/m2); cohort 2 comprised sixteen participants
with type 2 diabetes (n=16; F=6; mean age=57.6±10.6 years; mean BMI=36.2±3.5 kg/m2;
mean HbA1c= 66.2±12.3 mmol/mol). Study protocol
Cohort
1 underwent MRI examination on a 60cm 3.0T MRI system (Prisma, Siemens AG,
Erlangen, Germany) using two 30-channel flex array. VARPRO was acquired in the left
ventricle (LV) short-axis planes in the basal, mid, and apical positions. Respiratory
and ECG-gated proton point resolved spectroscopy (PRESS) was acquired with and without
water suppression. The voxel was placed on the interventricular septum at
end-diastole.
Cohort
2 were examined twice on a 70cm 3.0T MRI system (Skyra, Siemens AG, Erlangen,
Germany), using a 30-channel flex anterior array combined with the spinal
array. VARPRO was acquired after eight hours of fasting. At the time of rescanning,
cohort 2 had mean BMI and HbA1c of 36.7±4.0kg/m2 and 62.5±14.2mmol/mol
respectively. The mean inter-scan interval was 7.4 months.Data collection
Segment
(v2.2 R6190, Medviso, Lund, Sweden) was used for image analysis. Images with
apparent artefacts were rejected. The regions of interest (ROIs) were placed
within manually segmented myocardia with a further 20% erosion in both endo-
and epicardial directions (Figure 1). Basal and mid-cavity fat-fractions were
averaged to achieve VARPRO fat-fraction.
Myocardial
spectra were analysed using Java-based magnetic resonance user interface
(jMRUI, V6.0, Leuven, Belgium). Advanced method for accurate, robust and
efficient spectral (AMARES) fitting of MRS data was used to quantify the
amplitude of water and lipid signals. Intramyocardial lipid was recorded as the
sum of methyl and methylene resonances at 0.9 and 1.3 ppm (Figure 2).Statistical analysis
Paired t-test and Pearson correlation tests were used to assess the relationship
between the fat-fractions obtained by the two methods. A sub-group of cohort 1
(n=20) were randomly selected to develop a linear adjustment equation, and this
was validated in the remaining 10 participants. The agreement between adjusted
VARPRO and MRS fat-fraction was tested using Bland and Altman method7 which was also used to assess
temporal reproducibility in the repeated
VARPRO measurements on cohort 2.Results
There
was a strong linear association (r=0.897, P<0.001) between myocardial lipid
concentration obtained by VARPRO and MRS. However, the numerical values of the
two methods are significantly different (p<0.001), suggesting poor agreement;
therefore, VARPRO values were adjusted according to the linear regression equation
obtained from a derivation set ($$$MRS=-0.78+0.32*VARPRO$$$) (Figure 3). Adjusted VARPRO was compared to
MRS in a validation set, with good concordance (11% mean variation) (Figure 4).
Our results also showed VARPRO exhibits good temporal reproducibility with the
coefficient of repeatability of 3.18 (Figure 5). Discussion
Using
VARPRO to quantify fat-fraction is achieved by directly measuring mean signal
intensity on water-only and fat-only images. Thus, the accuracy is susceptible
to thermal noise and other image artefacts, which could explain why lipid
concentrations determined by VARPRO are consistently higher than those measured
using MRS. However, this bias is non-differential, and can easily be adjusted using
linear models. Moreover, the high reproducibility of VARPRO makes it an ideal
method for longitudinal studies. Finally, VARPRO does not require lengthy scan
times or special technical expertise. This makes the investigation of metabolic
heart disease less burdensome clinically.
The
research has some limitations. Firstly, the overall sample size is small. Secondly,
as the accuracy-test was conducted on healthy volunteers, the full spectrum of
myocardial steatosis have not been studied. Thirdly, the test re-test
variability was sampled, on average, 7.4 months apart. Although there was no
significant changes in BMI or HbA1c, interval change might still be a confounder. Conclusions
VARPRO could be a potential, alternative modality to MRS in myocardial
lipid quantification. It is less susceptible to technical disadvantages.
Patient tolerance is markedly better compared to MRS. VARPRO can be easily incorporated
into the routine clinical protocol, making a non-invasive assessment of cardiac
metabolism possible. Further studies on larger and diverse groups of participants could
help evaluate the performance of VARPRO, covering the full spectrum (moderate to severe) of cardiac steatosis. Acknowledgements
No acknowledgement found.References
1. Ng
AC, Delgado V, Djaberi R, Schuijf JD, Boogers MJ, Auger D, et al. Multimodality
imaging in diabetic heart disease. Curr Probl Cardiol. 2011;36(1):9-47.
2. Graner M, Pentikainen MO, Nyman K,
Siren R, Lundbom J, Hakkarainen A. Cardiac steatosis in patients with dilated
cardiomyopathy. Heart. 2014;100.
3. Faller KM, Lygate CA, Neubauer S,
Schneider JE. 1 H-MR spectroscopy for analysis of cardiac lipid and creatine
metabolism. Heart failure reviews. 2013;18(5):657-68.
4. Weiss K, Martini N, Boesiger P,
Kozerke S. Cardiac proton spectroscopy using large coil arrays. NMR in
Biomedicine. 2013;26(3):276-84.
5. Hernando D, Kellman P, Haldar J,
Liang ZP. Robust water/fat separation in the presence of large field
inhomogeneities using a graph cut algorithm. Magnetic Resonance in Medicine: An
Official Journal of the International Society for Magnetic Resonance in
Medicine. 2010;63(1):79-90.
6. Hernando D, Haldar J, Sutton B, Ma
J, Kellman P, Liang ZP. Joint estimation of water/fat images and field
inhomogeneity map. Magnetic Resonance in Medicine: An Official Journal of the
International Society for Magnetic Resonance in Medicine. 2008;59(3):571-80.
7. Bland
JM, Altman DG. Statistical methods for assessing agreement between two methods
of clinical measurement. Lancet. 1986;1.