Yuanbo Yang1, Zhonghao Li2, Yihao Guo1, Yingjie Mei3, Ming Zhao2, Guoxi Xie4, and Yanqiu Feng1
1Guangdong Provincial Key Laboratory of Medical Image Processing & Key Laboratory of Mental Health of the Ministry of Education, School of Biomedical Engineering, Southern Medical University, Guangzhou, China, 2Department of Pathophysiology, Key Lab for Shock and Microcirculation Research of Guangdong, School of Basic Medical Sciences, Southern Medical University, Guangzhou, 510515, China, Guangzhou, China, 3philips healthcare, Guangzhou, China, 4Department of Biomedical Engineering, School of Basic Sciences, Guangzhou Medical University, Guangzhou, China
Synopsis
Carotid atherosclerosis
is a degenerative disease of the arterial wall, which can result in
predisposition to cerebral thrombo-embolic stroke1. The combination
of DANTE and VF-RARE (DANTE-VF-RARE) can suppress flow signal effectively. It
was used for vessel wall imaging of human in the previous studies4.
However, the feasibility of DANTE-VF-RARE sequence remains unclear for imaging
of carotid for apoE-/- mouse. This work aims to develop 3D DANTE-VF-RARE on
7Tesla (T) and investigate its feasibility for vessel wall imaging of apoE-/-
mouse.
Introduction
Carotid
atherosclerosis is a degenerative disease of the arterial wall, which can
result in predisposition to cerebral thrombo-embolic stroke1. Three-dimensional (3D) variable flip-angle Rapid Acquisition
with Relaxation Enhancement (VF-RARE) sequence can achieve near constant echo
amplitudes over long echo trains and has been demonstrated successfully in
vessel wall imaging2,3. Meanwhile, the combination of DANTE and VF-RARE
(DANTE-VF-RARE) can suppress flow signal effectively. It was used for vessel
wall imaging of human in the previous studies4. However, the
feasibility of DANTE-VF-RARE sequence remains unclear for imaging of carotid for
apoE-/- mouse. This work aims to develop 3D DANTE-VF-RARE on 7Tesla (T) and investigate its feasibility for vessel wall imaging of apoE-/-
mouse.
Methods
DANTE Optimization
The simulation program
was implemented in Matlab. The goal of the DANTE optimization is to
maximize the contrast between the vessel wall and blood. In numerical simulations,
the T1/T2 values of the vessel wall and arterial blood were 1628/46 ms and
2200/68 ms, respectively. The flip angle was ranged from
with increment of 1o and the number
of pulse was ranged from 0 to 600. To simplify the numerical simulations, an
approximation signal evolution5 was adopted, so that the blood flow
velocity was not considered.
Experiments
This study was approved by
local institutional review board. Three ApoE-/- mice (8 weeks old) were fed
with a high-fat diet for approximately 32 weeks. Mice underwent MR scan using
Bruker 7T MRI scanner. Figure 1(a) shows a schematic representation of the
DANTE-VF-RARE pulse sequence. The parameters for the DANTE module included:
flip angle :10°, pulse train length :150, inter-pulse duration td: 0.97 ms.
The parameters for the VF-RARE readout included: TR/TE = 1200/22.3 ms, FOV
= 20×20×16 mm3, True resolution=
0.1×0.1×0.2 mm3 and reconstructed to 0.05×0.05×0.1 mm3, bandwidth=500
Hz/pixel, slice partial Fourier factor = 6/8, NA= 3, total scan time= 14 min24s.
The variable-flip angle scheme for the VF-RARE acquisition is illustrated in
Fig. 1(b). Conventional 2D RARE images were acquired as a reference with the
following parameters: TR/TE = 1200/7 ms , FOV = 20×20 mm2 ,in-plane True
resolution= 0.1×0.1 mm2 and reconstructed to 0.05×0.05 mm2; slice thickness =
0.5 mm, bandwidth=500 Hz/pixel; NA= 3, total scan time= 12 min. A pair of parallel saturation bands were placed above
and below the slices group. To confirm structural changes of the carotid lumen
detected by VF-RARE images, A FLASH sequence was performed for bright blood
imaging with following imaging parameters : TR/TE = 15/2.6ms, flip angle = 20◦,
FOV =20×20×16 mm3, True resolution:= 0.1×0.1×0.2 mm3 and reconstructed to
0.05×0.05×0.1 mm3, NA= 3, total scan time= 13min12s.
Histology
Two days after MRI
measurements, histologic sections were made. Carotid artery was sectioned into
5-µm serial sections perpendicular to the vessel direction.
Image Analysis
ImageJ was used to measure lumen signal ( Slumen ), noise level(
σ
) and wall signal(Swall). Swall was measured as
the mean signal intensity of a one-pixel-width circumferential path traced
manually through the middle of the arterial wall6. To allow for fair
comparison of CNR values with respect to imaging parameter differences, CNR
efficiency ( CNReff ) was calculated by the relation7:
CNReff=CNR÷(VOXEL×√TASLICE )
Where
VOXEL is the imaging voxel volume and was expressed in cubic millimeters, TASLICE is the scan time
for each slice (expressed in minutes).
SPSS was used
for statistical analysis. A two-tailed paired Student’s t test was performed
between 3D DANTE-VF-RARE and 2D RARE images for SNRlumen, SNRlumen , CNRwall_lumen , CNReff , LA(carotid lumen area), WA(carotid wall area);
Moreover, the intra-class correlation coefficient (ICC) and Bland-Altman plots
were performed to assess the agreement in WA and LA measurements obtained from
the 3D DANTE-VF-RARE and 2D RARE . A p value of less than 0.05 was considered
statistically significant. Data were presented as means
standard
deviations. Result
Figure 3c shows that the contrast is anticipate to peak earlier but
in a lower steady state for higher flip angles, as compared with that for lower
flip angles. Figs. 3 shows that sufficient contrast can be obtained using a
DANTE pulse with flip angle of 10° and the train length of approximately 150. 3D
DANTE-VF-RARE and 2D RARE images from a representative ApoE-/- mouse are shown in Figs. 3 and 4. DANTE-VF-RARE can generate a
comparable image quality while achieving 2.5 fold improvement in slice
resolution over 2D RARE (0.2mm vs 0.5mm). Quantitative measurements and
statistical analyses were listed in Table 1. There were no significant difference
in CNRwall_lumen (25.70±4.96 vs 21.99±5.34, P=0.071) between the 3D and 2D acquisitions. However, the 3D
protocol offered a significantly improved CNReff (30266.98 ± 5844.60 vs
5115.11±1242.83, P<0.001). For a comparison of morphology, the carotid LA
and WA obtained from DANTE-VF-RARE and 2D RARE were in a good agreement based
on the ICC analysis (LA: ICC=0.983; WA: ICC =0.996) and the Bland-Altman plots(Figs.5).Discussion and Conclusion
In this work, the results
shows that the combination of DANTE and VF-RARE can achieve high spatial resolution carotid
artery imaging at 7T. 3D DANTE-VF-RARE is able to generate a comparable
while achieving 2.5 fold improvement in slice
resolution over 2D RARE (0.2mm vs 0.5mm). 3D DANTE-VF_RARE allows for reliable
visualization and quantitative evaluation of the arteriosclerotic lesions
within the carotid artery of an apoE-/- mouse model at 7T.Acknowledgements
No acknowledgement found.References
1.Ingall T . Stroke-incidence, mortality, morbidity and risk[J]. Journal
of insurance medicine (New York, N.Y.),2004, 36(2):143-152.
2. Fan Z , Zhang Z , Chung Y C ,
et al. Carotid arterial wall MRI at 3T using 3D variable-flip-angle turbo
spin-echo (TSE) with flow-sensitive dephasing (FSD)[J]. Journal of Magnetic
Resonance Imaging Jmri, 2010, 31(3):645-654.
3.Qiao Y , Steinman D A , Qin Q , et al. Intracranial arterial wall
imaging using three-dimensional high isotropic resolution black blood MRI at
3.0 Tesla[J]. Journal of magnetic resonance imaging : JMRI, 2011, 34(1):22-30.
4.Xie Y , Yang Q
, Xie G , et al. Improved black-blood imaging using DANTE-SPACE for combined
carotid and intracranial vessel wall evaluation[J]. Magnetic Resonance in
Medicine, 2015, 17(Suppl 1):2286-2294.
5.Li L, Miller KL, Jezzard P. DANTE-prepared pulse trains: a novel
approach to motion-sensitized and motion-suppressed quantitative magnetic
resonance imaging. Magn Reson Med 2012;68: 1423–1438
6.Koktzoglou I ,
Li D . Diffusion-Prepared Segmented Steady-State Free Precession: Application
to 3D Black-Blood Cardiovascular Magnetic Resonance of the Thoracic Aorta and
Carotid Artery Walls[J]. Journal of Cardiovascular Magnetic Resonance, 2007,
9(1):33-42.
7.Zhang Z, Fan Z,
Carroll TJ, et al. Three-dimensional T2-weighted MRI of the human femoral
arterial vessel wall at 3.0 Tesla. Invest Radiol 2009;44:619–626