Xingxian Shou1, Yuan Zheng1, Qi Liu1, Masoud Edalati1, Lele Zhao2, Junpu Hu2, Jian Xu1, Weiguo Zhang1, Donel Tani3, and Nael F. Osman3
1MR, UIH America, Inc., Houston, TX, United States, 2MR, United Imaging Healthcare, Shanghai, China, 3Myocardial Solutions, Inc., Morrisville, NC, United States
Synopsis
SENC MRI is an important
technique for myocardium strain quantification. We proposed a modified
SENC EPI sequence to only acquire about half of the total cardiac phases in
order to get higher SNR strain images. The minimum strains can still be
calculated with better SNR from LT/HT images. With the imaging time kept the
same at one heartbeat, the SNR of the images is higher and strain calculation
is more accurate. The timing of the end-systole stage can be determined by
running a CINE scan prior to the SENC protocol.
Introduction
Strain-encoded
(SENC) magnetic resonance imaging is an advanced tagging technique1,2,
which provides both color-coded visual and quantitative assessment of
myocardial strains. While in conventional tagging the tagging planes are applied
in the orthogonal direction of the imaging slice, SENC applies them parallel to
the imaging slice. The tagging planes in the through-plane direction compress
together during myocardium contracting, and conversely become further distanced
during stretching. This causes a shift in the location of the peak spectrum in
k-space, which can be used to determine myocardial strain1,2. With
the high and low tuning frequency encoded in slice selection, images are
acquired during the cardiac phases. So that the acquisition of a single slice can be done as
short as one heartbeat. With pairs of LT(low-tuning) and HT(high-tuning)
images, the strains can be calculated for various regions of the myocardium1,2.
However, the SNR can be very low in these LT or HT images compromising the
ability to view more detailed regions of the myocardium. A modified version of
SENC sequence is proposed to increase SNR by acquiring fewer images, which are selected
around the end-systole during cardiac cycle. Methods
A SENC MRI sequence with single-shot EPI readout was implemented on a 1.5T MRI system
(uMR 560, United Imaging Healthcare, Shanghai, China). Images of a healthy
30-year old female volunteer were acquired with IRB approval. With the fat
saturation and selective tagging pulse module played out immediately following
the end of diastole, single shot EPI was used to acquire tuned image during
cardiac phase. With low and high tuning alternative imaged, a total of 16-20
images can be acquired during the cardiac cycle. With a pair of low and high
tuning images, the strain can be calculated which corresponds to that cardiac
phase. Not only we acquired the full phases of cardiac cycle, in comparison, we
also acquired about half of the phases around end-systole stage (trigger delay
time was determined by a CINE scan). Imaging parameters were: slice thickness = 10
mm, FOV = 360×180 mm2 with pixel 4.00×3.75 mm2 ,TR = 49 ms, TE = 23.9 ms,
tagging distance = 3.92 mm, bandwidth = 2000 Hz/pixel. Single-shot EPI with
ramped or constant flip angles was used to image each cardiac phase. The ramped
flip angle considers T1 effect3 to maintain constant myocardial
signal intensity during cardiac phases. Results
The modified SENC
sequence generated more SNR compared to its regular version. In figure 2, we
show the images at trigger time about 226 ms with three experiments: (a) FA = 30,
regular SENC (b) FA = 30, fewer phases (c) FA = 60, fewer phases. The SNR was
estimated as the mean value of ROI divided by the noise variation in the
background, also with a correction factor $$$\sqrt{\frac{2}{4-\pi}} =1.53$$$ because of the Rayleigh
distribution of background noise in magnitude images4. $$$m_{ROI}$$$ is the mean value of selected ROI, while $$$s_{noise}$$$ is the standard deviation of the background noise. $$SNR=\frac{S_{mean}}{\sigma_{stdv}}=\frac{m_{ROI}}{\sqrt{\frac{2}{4-\pi}}s_{noise}}$$ With fewer
acquired cardiac phases, the SNR is higher compared to full phases. Also, with
increased flip angle as shown in (c), the SNR increased as expected. With
shortened RF train, higher flip angle increases the SNR in EPI acquisition. This
helped us in the pixel by pixel calculation of the strain values, which gives
us more accurate strain values for the heart. Figure 3 shows all the images from modified
SENC sequence with different RF flip angles: (a) FA = 30 (b) FA = 60 (c) FA =
ramped to 75 (d) FA = ramped to 90. The ramped RF train demonstrates that the
signal is maintained more constant during the cardiac phases, compared to the
constant RF pulse trains. The images can be assessed using the
software provided my Myocardial Solution Inc(MSI), Morrisville, NC, USA. The DICOM images are
imported by MSI software to do post-processing, and the strain images are
generated with peak strains calculated. Figure 4 shows short-axis heart strain
images and the longitudinal strain values for left ventricle (LV). Figure 5
shows the 4-chamber heart strain images and the circumferential strain values
for LV. All these calculated strain values are expected for a healthy young volunteer. Discussion and Conclusion
We
have proposed the modified SENC sequence for higher SNR in the myocardium
strain quantification. With phantom and volunteer scans, we have demonstrated
that the proposed method can generate higher SNR in the SENC images acquisition,
and a direct comparison between the regular full cardiac phases acquisition and
the selected cardiac phases acquisition around end-systole shows that our
method generates more accurate strain calculation. Although we get fewer
cardiac phases scanned and there are fewer detailed strains value regarding
different phases, the most important one which is the peak strains can still be
calculated with more accuracy. Acknowledgements
No acknowledgement found.References
1.
Osman F., Sampath, S., Atalar, E., Prince,
J., Imaging Longitudinal Cardiac Strain on Short-Axis Images Using
Strain-Encoded MRI, Magn Reson Med, 2001; 46:324-334.
2.
Pan, L., Stuber, M., Kraitchman, D., Fritzges,
D., Gilson, W. and Osman N., Real-Time Imaging of Regional Myocardial Function
Using Fast-SENC, Magn Reson Med. 2006; 55:386-395.
3.
Fischer, S., McKinnon, G., Scheidegger, M.,
Prins, W., Meier, D., Boesiger, P., True Myocardial Motion Tracking, Magn Reson
Med. 1994; 31:401-413.
4.
Dietrich, O.,
Raya, J., Reeder, S., Reiser, M., & Schoenberg, S., Measurement of
signal-to-noise ratios in MR images: Influence of multichannel coils, parallel
imaging, and reconstruction filters. J of Magn Reson Imaging 2007; 26(2):
375–385