Hsin-Jung Yang1, John Stager1, Linda Azab1, Waishing Liu1, Meng Lu1, Yuheng Huang1, Ghazal Yoosefian1, Skyler Selvin1, Richard Handelin1, Yujie Shan1, Fardad Michael Serry1, Yibin Xie1, Anthony Christodoulou1, Xiaoming Bi1, Rohan Dharmakumar1, Debiao Li1, and Hui Han1
1Cedars-Sinai Medical Center, Los Angeles, CA, United States
Synopsis
B0-field inhomogeneity caused by tissue-air
interface has been a long standing challenge
in high field(>=3T) cardiac MRI. Although high-order shimming methods has
been proposed in recent publications using
surface shimming coils, the shimming capability in deep tissue, such
as the heart, is still limited due to the restrained current amplitudes and the
geometry constraints of the receiving loops. In this study, we developed a cardiac
high-order shimming coil based on a novel coil decoupling design to overcome
the aforementioned limitations. We demonstrated the developed coil can
successfully reduce B0 field variation by 50% in the whole-heart in healthy
human volunteers.
Introduction
Main
field (B0) inhomogeneity has been a long standing challenge in high field(>=3T)
cardiac MRI. The tissue-air interface between heart and lung induces strong
local B0 variation, which is proportional to the scanner field strength. This
effect degrades imaging capabilities of many important CMR sequences that are
sensitive to off resonance, such as bSSFP, EPI readouts and fat saturation and
T2 prep pre-pulses. It also diminish the
SNR and spectral benefit from increased field strength and restrict CMR
applications in high field scanners. Although state of the art scanners provide
the capability of the 2nd spherical harmonic shimming, it often fell
short when correcting the high order local field variations. In the past few years, RF coil arrays that
integrate shimming currents into receiving RF loops were developed by us and
other groups to enable high order local B0 shimming [1-3]. Although promising
results has been presented in neurological applications, the shimming
capability in deep tissue, such as the heart, is limited due to the restrained current
amplitudes and the geometry constraints of the receiving loops. In this study,
we use a novel technique we proposed recently, named unified coil (UNIC) [4,5].
UNIC overcomes the aforementioned limitations by decoupling the shimming and
reception loops using a figure-8 configuration and by enabling multiple-turn
shim loops to dramatically increase the shim field strength (Hardware design
details will be discussed in an educational course on Multi-Coil B0 Field Modelling &
Systems, ISMRM 2020). This allows shimming loops to be placed in close
peripheral of the receiving coils and thus enable strong local shimming
capacity. We tested the method in healthy human volunteers at a 3T clinical
scanner. Methods
A
unified coil was constructed with 12 RF receive and 42 shimming channels with 2-turn
shim loops in each shim channel. The circuit design is depicted in figure 1. The
constructed coil was placed on the subject’s chest wall to maximized SNR and shimming efficiency.
Scans were performed in healthy human subjects (N=10) under breath holds and
ECG gating. B0 field maps and bSSFP cine images with extended TR (TR= 6ms) were
acquired to test the shimming efficacy. The field homogeneity and bSSFP image
qualities were measured and compared with and without activating the UNIC
shimming. Images were acquired After localization scans, whole-heart shimming,
and scouting to determine the appropriate center frequency, breath-held,
flow-compensated bSSFP acquisitions were prescribed in multiple planes along
short-axis orientation with following scan parameters: TR = 6ms, flip angle =
50°, imaging resolution = 1.9 x 2.5 x 6 mm3, temporal resolution = 42 ms,
readout bandwidth = 550 Hz;Results
B0
Field and image qualities under baseline and UNIC shimming were compared in
figure 2. Panel A presents the quantitative comparison of the myocardial off-resonance frequency with and without UNIC shimming. Off resonance frequency
from each slice were measured. Both absolute mean and standard deviation of the
myocardial off-resonance frequency were significantly reduced (62.9 ± 57.3 Hz
vs 30.3± 40.5 Hz; p<0.05) after UNIC shimming. A set of representative short-axis images acquired with and without UNIC shimming is shown in panel B. Field
maps and bSSFP cine images acquired at the matching slice position are
presented. Significant field inhomogeneity is demonstrated at the heart-lung
interface before UNIC shim application (black arrows). Banding artifacts at the
posterior wall were observed in cine images and strongly degrade the image
qualities throughout the cardiac cycle. After UNIC shim application, B0 field
became remarkably more homogeneous in the myocardium and the banding artifacts
were eliminated in the bSSFP cine images.Conclusion
To
our best knowledge, this is the first study to perform local cardiac shimming in
a clinical high field scanner. UNIC coils successfully reduce off-resonance
frequency by 50% in the heart at 3T. The field improvement opens up the opportunity
for reliable CMR in high field scanners and can potentially unleash the full
SNR and spectral benefit for CMR applications. Acknowledgements
Thanks
to Fei Han, Bernd Stoeckel, Fraser Robb and Miguel Navarro for
their support. References
1. Han H
et al. MRM 2013;70:241-247.
2.
Truong TK et al. Neuroimage 2014;103:235-240.
3.Stockmann
JP, Witzel T, Keil B, et al. A 32-channel combined RF and B0 shim array for 3T
brain imaging. Magn Reson Med 2016;75:441-451.
4. Hui
Han, John Stager, Hsin-Jung Yang, Sizhe Guo, Zhuoqi Li, and Debiao Li. Unified
Coils (UNIC) for Parallel Imaging and B0 Shimming. 2016 Gordon Research
Conference, In Vivo Magnetic Resonance, 'MRI Inside-Out and Outside-In:
Innovative Technologies, Unmet Needs and New Opportunities'. Proctor Academy,
Andover, NH. July 17-22, 2016.
5. H.
Han, J. Stager, W. Cao, Z. Li, J. Cho, D. Zhou, Y. Wang, D. Li. Unified Coils
(UNIC) for Parallel Imaging and B0 Shimming. European Society for Magnetic
Resonance in Medicine and Biology (ESMRMB). ESMRMB 2016 Congress. 33rd Annual
Scientific Meeting, Vienna/AT, September 29 - October 1, 2016. (Oral
presentation).