El-Sayed H Ibrahim1, V Emre Arpinar1, L Tugan Muftuler1, Andrew Nencka1, and Kevin Koch1
1Medical College of Wisconsin, Milwaukee, WI, United States
Synopsis
Cardiac functional MRI has been established
in clinical practice on 1.5T and 3T scanners; nevertheless, the capabilities of
ultra-high field (UHF) MRI have not been fully exploited, which are expected to
significantly improve image quality and provide information not obtainable at
lower-field MRI. Despite its advantages, UHF MRI is challenging due to a number
of technical issues, especially B1 and B0
inhomogeneities. In this study, we provide preliminary results of using a
multi-channel transceiver modular coil and a dielectric pad towards optimizing UHF
cardiac functional MRI by improving image quality and minimizing imaging artifacts.
Introduction
Cardiac functional MRI has been
established in clinical practice on 1.5T and 3T scanners. Nevertheless, the
capabilities of ultra-high field (UHF) MRI have not been fully exploited, which
are expected to significantly improve image quality and provide more
information that allow for better understanding of cardiac function in both
normal and different pathophysiological conditions.1 UHF MRI
incorporates several advantages as well as some limitations. As signal-to-noise
ratio (SNR) is proportional to magnetic field strength, 7T MRI provides ~2.3
times SNR compared to 3T imaging. Enhanced SNR can be traded for improved image
quality, higher temporal or spatial resolution, or shorter scan time.
Despite its advantages, UHF MRI is
challenging due to a number of technical issues.1 First, the applied
radiofrequency (RF) fields at such high frequencies result in short wavelength (~12
cm) of the RF fields inside the tissue and increased dielectric effects that
can induce destructive and constructive interferences and result in distorted
RF fields and non-uniform effective flip angle distribution with observed
shading and signal drop-off. Imaging at UHF also results in increased B0
inhomogeneity due to magnetic susceptibility effects.
In this study, we provide preliminary
results of using a multi-channel transceiver modular coil and a dielectric pad towards
optimizing MRI cardiac functional imaging at 7T and improving image quality
while alleviating artifacts associated with UHF imaging. Methods
Ten healthy subjects were scanned on a 7T GE MRI scanner using a 32-channel
transceiver coil. The modular coil array consists of 8 independent blocks,
where each block contains 4 transceiver elements, whose phase settings were
optimized based on simulations for a multi-oblique plane mimicking a standard
cardiac view.2
The effect of the imaging flip angle on image quality was assessed
in vivo using a gradient-echo cine sequence. Image acquisition was repeated
with flip angles ranging from 1° to 120°. Forward simulation was conducted on
the resulting images to generate an estimate of the B1 transmission
frequency field in the imaged slice using actual imaging flip-angle (AIF) B1
mapping.3
Both short-axis (SAX) and long-axis (LAX) cardiac cine images were
acquired in the scanned subjects using optimal imaging parameters: fast spoiled
gradient echo sequence, TR=8ms, TE=4ms, flip angle=60°, matrix=256x256,
FOV=380x380 mm2, slice thickness=8mm, acquisition
bandwidth=244Hz/pixel, #averages =1, #cardiac phases =25, and SAR=0.095W/kg. Images
with higher spatial and temporal resolutions were also acquired to assess
capabilities of UHF cardiac functional imaging.
The effect of using a dielectric pad to improve B1
homogeneity was investigated in six subjects (three normal-weight and three
over-weight) with the pad placed at different positions close to the imaged
region-of-interest.Results
With proper settings and imaging parameters optimization, we were
able to achieve sufficient RF penetration and B1 homogeneity across
the heart, which allowed for successfully scanning all subjects with decent
image quality. The ECG signal was reliable in all in vivo scans and image
acquisition was correctly triggered with the R-wave, despite elevated T-wave in
most scans.
Figure-1 shows results from an in-vivo scan to estimate the B1
field. The figure shows variable signal intensity and regions of signal loss
across the acquired slice, where the signal intensity profile changes with the
imaging flip angle. The estimated B1 and proton density distribution
maps reflect the observed signal intensity inhomogeneity in the imaged slice.
The resulting R2 map shows excellent data fitting, as shown for
certain pixels located in the imaged slices.
Adding a dielectric pad close to the imaged slice could help
improve B1 homogeneity (Figure-2). Moving the pad around the thorax
area affects the presence and location of the signal nulling regions, where in
most cases placing the pad anteriorly improved signal homogeneity in
normal-weight subjects, while subjects with larger body habitus showed adequate
signal homogeneity even without using the dielectric pad.
Exploiting the capabilities of UHF imaging allowed for achieving high-spatial
resolution of 0.75x0.75x2 mm3, which is ~16 times better than
conventional cine imaging at 1.5T (Figure-3). Such high spatial resolution
allows for visualizing detailed anatomical features without sacrificing SNR as
in low-field imaging. Alternatively, higher temporal resolution of ~20 ms (50
cardiac phases) could be achieved to reveal subtle tissue motion pattern. Discussion
While
the majority of UHF MRI studies have focused on brain and orthopedic
applications, recent improvements in multi-channel coil development would open
the door for cardiac imaging to benefit from UHF capabilities.2,4
The
transceiver modular coil used in this study allowed for obtaining high-quality
in vivo images with high resolution and adequate myocardium-to-blood contrast,
while minimizing B1 non-uniformity and shading effects.
It
has been previously shown that actual flip angle varies by as much as 50% from
the nominal flip angle across the heart during high-field imaging.5 Adjusting
the imaging flip angle and adding a dielectric pad to the imaged
region-of-interest, especially when imaging subjects with small body habitus, could
help improve B1 homogeneity and reduce signal nulling resulting from
standing-wave effects. Conclusion
In conclusion, through proper scan settings and imaging parameter
optimization, 7T cardiac MRI would allow for improved cardiac functional imaging.
This is expected to open the door for more cardiac applications of UHF MRI and
potential adoption in clinical practice in the near future. Acknowledgements
Funding from Daniel
M. Soref Charitable Trust, MCW, USA.References
- Niendorf
T et al. NMR Biomed 2016;29(9):1173-97.
- Graessl
A et al. Magn Reson Med 2014;72(1):276-90.
- Yarnykh
VL et al. Magn Reson Med 2007;57(1):192-200.
- Thalhammer
C et al. J Magn Reson Imaging 2012;36(4):847-57.
- Suttie
JJ et al. NMR Biomed 2012;25(1):27-34.