Kazuo Kodaira1, Michinobu Nagao2, Masami Yoneyama3, Yasuhiro Goto1, Isao Shiina1, Yutaka Hamatani1, Mamoru Takeyama1, Isao Tanaka1, and Shuji Sakai2
1Department of Radioligical Services, Tokyo Women's Medical University Hospital, Tokyo, Japan, 2Department of Diagnostic imaging & Nuclear Medicine, Tokyo Women's Medical University Hospital, Tokyo, Japan, 3Philips Japan, Tokyo, Japan
Synopsis
Improvement of fat suppression is important for whole heart coronary magnetic
resonance angiography (WHC-MRA), because Unwanted signal arising from fat can
compromise vessel delineation and decrease the diagnostic value of WHC-MRA. Balanced DIXON (bDIXON) can reduce fat suppression
unevenness than SPIR method while maintaining high signal value. However,
it is yet not applied for WHC-MRA. Furthermore, 3D
non-selective bDIXON (3D NSbDIXON) with shortening TR can suppress an increase in
imaging time. We
propose a new sequence of 3D NSbDIXON for WHC-MRA, and
examine the image quality in comparison to the conventional methods.
Introduction
Whole
heart coronary magnetic resonance angiography (WHC-MRA) is a promising noninvasive method for detection of coronary artery
disease1. WHC-MRA requires
effective fat suppression because some coronary arteries are embedded in the epicardial
fat2, 3. WHC-MRA typically
applies SPIR pre-pulse, but it often obscured left circumflex artery (LCX) due
to partial fat suppression failure. DIXON with spoiled gradient-echo
(T1TFE-DIXON) improved the image quality of WHC-MRA compared to SPIR2.
However, a limitation of T1TFE-DIXON at 1.5T is its lower signal-to-noise ratio
(SNR).
3D balanced DIXON (3D bDIXON)
has the possibility to solve these problems. By applying the balanced-TFE readout
instead of T1TFE, it could increase the SNR, but it may suffer from banding
artifacts. To prevent the increasing of banding artifacts, we applied 3D
non-selective excitation pulses, which has already been demonstrated to improve
the quality of 3.0T bSSFP WHC-MRA4, for balanced-TFE DIXON. 3D non-selective
bDIXON (NSbDIXON) enables shortening TR (Fig. 1),
resulting in decreasing the banding artifacts in addition to increasing the TFE
factor during one heartbeat. Furthermore, 3D NSbDIXON has the potential to improve the effect of fat
suppression compared to SPIR while ensuring high SNR thanks to bSSFP readout
compared with conventional DIXON. Therefore, we hypothesized
that 3D NSbDIXON can improve the image quality of WHC-MRA. In
this study, we
investigated the feasibility of 3D NSbDIXON for WHC-MRA by comparing with NSbTFE
and non-selective T1TFE-DIXON (NST1TFE-DIXON).Methods
A
total of six volunteers (6 males;
age range: 26~44)
were examined on a 1.5T MRI (Ingenia CX, Philips Healthcare). The study was
approved by the local IRB, and written informed consent was obtained from all
subjects.
First, we compared the WHC-MRA images among
3D NST1TFE-DIXON, 3D NSbTFE-SPIR, and 3D NSbDIXON. Three sequences were
acquired with navigator and cardiac triggering with the
following parameters: FOV=350×350mm,
voxel-size=1.5×1.5×1.5mm, C-SENSE_factor=7, TR/TE1/TE2/FA for NST1TFE-DIXON=5.3/1.73/3.6/20, for NSbTFE=2.3/1.14/NA/70, and for NSbDIXON=3.3/1.14/2.2/70.
Curved
Planer Reconstructions (CPRs) were performed using Ziostation2 (Ziosoft Co,
Tokyo) and image quality was evaluated by visual score at the 10 points (RCA:
#1, 2, 3, 4 LAD: #5, 6, 7, 8 CX: #11, 13) based on American Heart Association
(AHA) classification. we evaluated them as 4-point grades (grade “4” was
excellent, “1” was severe) by three blinded readers.
For quantitative comparison, SNR
and contrast-to-noise ratio (CNR) were measured. The SNR was assessed in the aorta
and epicardial fat. To allow quantitative SNR
measurements, we used a noise measurement method proposed
by Zwanenburg et al5. Each
sequence was repeated with exactly the same receiver gain, but without any RF and gradient pulses. The reconstructed images showed only noise,
including the noise added due to the Compressed-SENSE (C-SENSE) reconstruction. The standard deviation of a region of
interest of the
corresponding area in the noise image was used as
metric for the noise. SNRaorta and SNRepicardial fat were
then calculated as follows:
SNRaorta = SI (aorta) / SDnoise (aorta)
SNRepicardial fat = SI
(epicardial fat) / SDnoise (epicardial fat)
where
SI (aorta) and SI
(epicardial fat) are the mean average signal intensity of the aorta and epicardial fat respectively, and the
corresponding SDnoise is the standard
deviation at the same location on the noise images.
Subsequently, we measured the
CNR for comparing image contrast quantitatively. The CNR was
estimated for aorta and epicardial fat (CNRaorta-epicardial fat) and aorta and myocardium (CNRaorta-myocardium). The CNRaorta-epicardial fat and CNRaorta-myocardium were calculated by the
following equations:
CNRaorta-epicardial
fat = [SI (aorta) - SI (epicardial fat)] / 0.5
[SDnoise (aorta) + SDnoise (epicardial fat)]
CNRaorta-myocardium =
[SI (aorta) - SI (myocardium)] / 0.5 [SDnoise (aorta) + SDnoise (myocardium)].
The SNR and CNR were assessed by wilcoxon
signed-rank test.Results
Figure 2
and 3 shows the representative images using NST1TFE-DIXON, NSsbTFE and NSbDIXON.
The results of visual score are shown in Figure 4. 3D NSbDIXON was significantly higher
than NST1TFE DIXON and NSsbTFE
regarding RCA, LAD, CX.
Figure 5 shows the SNR and the
CNR comparison among three sequences. For the SNRaorta, NSbTFE and NSbDIXON showed significantly
higher value compared to NST1TFE-DIXON. In the SNRepicardial fat, NSbDIXON and NST1TFE-DIXON
showed significantly lower value compared to NSbTFE. Furthermore,
NSbDIXON showed significantly lower value compared to NST1TFE-DIXON. In the CNRaorta-epicardial fat, NSbDIXON showed significantly higher value compared to NSbTFE and NST1TFE-DIXON. In the CNRaorta-myocardium, NSbDIXON and NSbTFE showed significantly higher value compared to NST1TFE-DIXON.Discussion
From the results of SNR,
it is considered that bDIXON provides higher SNR similar to conventional bSSFP
compared with conventional DIXON. In addition, DIXON had a higher fat
suppression effect than SPIR, and even a slight fat signal could be suppressed.
Therefore, bDIXON has a lower epicardial fat signal value than bTFE, and bDIXON
is considered to have an improved CNRaorta-epicardial fat over bTFE.Conclusion
WHC-MRA derived from 3D
NSbDIXON has a higher blood signal than that of T1TFE-DIXON and a better
contrast between arteries and background fat tissues than that of conventional
bTFE-SPIR. This might improve the visualization of entire coronary arteries.Acknowledgements
No
acknowledgements found.References
(1) Teresa
Correia et al. Technical note: Accelerated nonrigid motion-compensated isotropic 3D coronary MR angiography. Med Phys. 2018 Jan; 45(1): 214-222.
(2) Maryam
Nezafat et al. Coronary
MR angiography at 3T: fat suppression versus water-fat separation. Magn Reson Mater Phy. 2016; 29: 733-738.
(3) Peter
Bornert et al. Water/Fat-Resolved Whole-Heart Dixon Coronary MRA: An Initial
Comparison. Magn Reson Med. 2014; 71: 156-163.
(4) Isao Shiina, Kazuo Kodaira, et al. Whole-heart coronary MRA using Non-Selective
balanced SSFP sequence at 3T: comparison of image quality. Proceedings
of the 27th Annual Meeting of ISMRM (2019).
(5) Zwanenburg et al. MR Angiography of the cerebral perforating arteries
with magnetization prepared anatomical reference at 7T: comparison with
time-of-flight. J Magn Reson Imaging. 2008; 28: 1519–1526.