Quentin Lebret1,2,3, Pierre Bour1,2,3, Valéry Ozenne1,2,3, and Bruno Quesson1,2,3
1IHU LIRYC, Fondation Bordeaux Université, Pessac, France, 2U1045 CRCTB, Université de Bordeaux, Bordeaux, France, 3INSERM, CRCTB, U1045, Bordeaux, France
Synopsis
Using Wave-CAIPI-like techniques to accelerate Balanced Steady-state
Free Precession (bSSFP) sequences is a new parallel imaging technique that
reduces the noise levels on the reconstructed image compared to conventional
parallel imaging. On the heart, this combination of sequence has not been
tested. The experiments show that the movements of the thorax are not an issue,
as long as a respiratory and ECG triggering are employed. A retrospective acceleration of 5.4
on a sheep heart image with a 1.4mm isotropic resolution has been achieved.
INTRODUCTION
3D acquisitions in the
heart are time consuming due to the necessity to trigger the acquisition on
the heart rate and compensate the respiratory motion. In
clinical settings a trade-off between acquisition duration and spatial coverage is
often compulsory. Parallel-imaging techniques allow reducing the
acquisition time while producing good quality images by taking
benefits of array receivers1. More recently, Wave-CAIPI acquisitions
initially proposed for whole brain imaging have shown the added value of adding
sinusoidal gradients during readout to better control aliasing2. Su and
al have shown the effectiveness of using a modified Wave-CAIPI technique
conjointly with a bSSFP sequence (Wave-bSSFP) in reducing the g-factor when high
acceleration factors are employed3. While this holds true for the
brain, the feasibility of such a technique on the heart remains to be
evaluated. First, coil coverage surrounding the heart remains less optimal as
compared to whole-brain coil arrays. In addition, issues might be related to cardiac
and respiratory motions, together with the presence of structures showing
different magnetic susceptibilities inside the thorax. In this work, we implemented a
wave-bSSFP sequence with T2-preparation module for in vivo 3D rapid cardiac imaging.METHODS
- Pulse Sequence: We implemented at 1.5T (MAGNETOM Aera,
Siemens Healthineers) a wave encoded b-SSFP sequence with a 40 ms T2-preparation
module. Wave gradients
were truncated (as proposed by Su and al) to maintain the balanced gradient condition3.
Wave gradients intensity was set up to 4mT/m. The acquisitions were ECG gated
in cardiac diastole and a crossed pair navigator echo was used to track and
compensate respiratory motion (4 mm acceptance window). A flip angle ramp up over 10 TR was used to reach the steady state. Trajectories for the PSF estimation
were measured using 1D calibration scans prior to image acquisition4. Trajectories calibration consisted in acquiring two 2 mm slices in the y direction and two 2 mm slices in the z
direction. During these acquisition, only the gradient to characterize were played. The phase on the MRI signal was analyzed to retrieve the trajectory. The chronogram
of the sequence is shown in Figure 1.
-
In Vivo Experiment: Fully sampled
acquisitions were performed in the thorax of an anesthetized sheep (protocol
approved by ethic committee) with the following parameters: TE=2ms; TR=350ms; Flip Angle = 90°; 50
segments acquired per cardiac beat; FOV=300x300x112mm3 and matrix
size=208x208x80 resulting in an isotropic spatial resolution of 1.4 mm.
Acquisition were oversampled 8 folds on the readout direction. We used 40 coil elements (body and spine
multi-array coils). Calibrations and imaging data were streamed inline to the Gadgetron framework5 and
a MATLAB (MATLAB 9.5, The Mathworks, Inc) script using BART toolbox6 was run for image reconstruction.
-
Offline Image Reconstruction: The fully sampled in-vivo acquisition was reconstructed
to create a reference 3D volume. The effect of the wave gradient was
de-convoluted from the fully sampled data using the PSF formalism developed by
Bilgic and al2: $$\small\begin{equation}wave[x,y,z] = F_x^{-1}(PSF[k_x,y,z]\times F_x.m[x,y,z])\end{equation}$$ where $$$\small\begin{equation}m[x,y,z]\end{equation}$$$ is the underlying magnetisation, $$$\small\begin{equation}wave[x,y,z]\end{equation}$$$ is the image convoluted by the wave gradients and $$$\small\begin{equation}F_x\end{equation}$$$ is the Fourier transform operator on the x axis. The
corresponding images served as a gold standard.
The raw data were then
retrospectively subsampled and reconstructed to simulate accelerated
acquisitions. For this, a Poisson undersampling mask was applied in ky-kz
dimensions. Coils Sensitivities Maps were estimated using ESPIRIT1 on autocalibration scans (ACS kernel size = 24x24
pixels) or on low-resolution Wave-bSSFP images de-convoluted by low resolution
PSFs (kernel size=24x24 pixels)7. The images were afterwards
reconstructed using general ESPIRIT model proposed in BART toolbox.
RESULTS
During acquisition, the fully sampled dataset were acquired over 15 minutes, reconstructed online using our
proposed Gadgetron/BART algorithm and transferred back to the scanner for immediate image visualization. The proposed sequence and reconstruction pipeline offer
flexibility to tune parameters during the experiment.
Figure 2a and b show
the theoretical and
experimental trajectories resulting from a wave gradients played along ky or kz
directions respectively. Strong drifts are observed between experimental and
theoretical trajectories mainly due to eddy currents and gradient
imperfections. Using the theoretical
trajectories ghosts are clearly visible (Figure 2c) as we move away from the scanner
isocenter. With the proposed calibration scan, ghosts disappeared in the reconstructed image (Figure
2c).
Figure 3 shows images
after 3.4 and 5.4 acceleration factors. Estimation of coil sensitivities using
low-resolution deconvoluted data or ACS lines results in less than 0.2% of
signal difference.DISCUSSION
The results validate
the use of a Wave-bSSFP type sequence on the heart. In a first approach, data
subsampling was performed offline to offer flexibility in image
reconstruction.
Images show acceptable
RMSE compared to the fully sampled data at both 3.9 and 5.4 acceleration
factors. Data with higher accelerations factors are currently being investigated
upon. Incorporating Compressed Sensing in the SENSE reconstruction model should
result in an improvement of the image quality.CONCLUSION
This work presents the
first usage of a Wave-CAIPI type sequence on the heart using a bSSFP sequence. Similar image qualities were obtained after subsampling the full k-space by a 3.9 and 5.4 factor showing that 3D acquisition of the heart with a 1.4 isotropic spatial resolution is feasible much faster with good image quality.Acknowledgements
This work received financial support from the French National Investments for the Future Programs: ANR-10-IAHU-04 (IHU Liryc)References
1. Uecker M, Lai P,
Murphy MJ, et al. ESPIRiT-an eigenvalue approach to autocalibrating parallel MRI: Where
SENSE meets GRAPPA. Magnetic Resonance in Medicine. 2014;71(3):990-1001.
2.
Bilgic B, Gagoski
BA, Cauley SF, et al. Wave-CAIPI for highly accelerated 3D imaging. Magnetic Resonance in
Medicine. 2015;73(6):2152-2162.
3.
Su S, Liang D, Qiu
Z, et al. Accelerating Three-Dimension Balanced Steady-State Free Precession
Imaging with Modified Wave-CAIPI Technique. Annual Meeting ISMRM, Montreal 2019, In Proc. Intl.
Soc. Mag. Reson. Med. 27
4.
Zhang Y, Hetherington HP, Stokely EM, Mason GF, Twieg DB. A novelk-space
trajectory measurement technique. Magnetic Resonance in Medicine.
1998;39(6):999-1004.
5.
Hansen MS, Sørensen TS. Gadgetron: An open source framework for medical
image reconstruction: Gadgetron. Magn Reson Med. 2013;69(6):1768-1776.
6.
Uecker M,
Ong F, Tamir J, Bahri D, Virtue P, Cheng J, Zhang T, and Lustig M. Berkeley Advanced Reconstruction Toolbox. Annual Meeting ISMRM, Toronto 2015, In Proc. Intl. Soc.
Mag. Reson. Med. 23:2486
7.
Chen F, Zhang T, Cheng JY, Shi X, Pauly JM, Vasanawala SS.
Autocalibrating motion-corrected wave-encoding for highly accelerated
free-breathing abdominal MRI: Wave-Encoding for Free-Breathing Abdominal MRI. Magn
Reson Med. 2017;78(5):1757-1766.