Fang Dong1 and Jing An1
1Siemens Magnetic Resonance Ltd., Shenzhen, China
Synopsis
T2-weighted MR imaging is a non-invasive way for vasculitis
diagnosis. SPACE sequence has been successfully used for T1-weighted aortic
wall imaging. While its application in T2w aortic wall imaging
is hampered by the long scanning time, blood suppression performance
and physiological movement. In this study, three black blood methods with
optimized protocols have been tested and evaluated. Diffusion-Prepared SPACE
sequence with diastolic acquisition was proved to be a promising way for T2w
aortic wall imaging within a clinically acceptable acquisition time.
Introduction
Vasculitis, the
inflammation of blood vessels, is rare [1], yet can affect people of all ages
and sometimes can be very serious. T2-weighted
(T2W) MR imaging provided a non-invasive way to detect and diagnose the vasculitis.
The conventional way for T2W aortic wall imaging is T2W DIR-prepared 2D fast spin-echo [2].
However, it is susceptible to partial volume effects and compromises aortic
coverage when sections are positioned perpendicular to the direction of aortic
blood flow.
High resolution, Isotropic 3D fast spin echo sequence with
variable flip angle (SPACE) is advantageous in aortic wall imaging. It
has been successfully used for T1-weighted (T1W) aortic wall imaging in healthy
volunteers and patients with cardiovascular or atherosclerotic disease with
high isotropic resolution [3-6]. However, it is not widely used for T2W aortic
wall imaging because of the long scanning time. Other
challenges like insufficient blood suppression, motion from heartbeat and respiration
are also critical for aortic wall imaging.
In this study, we aim to implement three blood suppression methods
and evaluate the performance in healthy volunteers and optimize the protocols to
achieve good image quality within a clinically acceptable scan time.Methods
SPACE sequence has its intrinsic
black blood effect, however, blood suppression is less effective in the
presence of slow or complex flow, and flow artifacts have been observed even in
healthy volunteers with normal aorta geometry [2]. Three methods have been
provided to further suppress the blood signal: (1) Diffusion-Prepared pulse (2)
DANTE
preparation pulse (3) Data acquisition during systole to enhance its
intrinsic black blood effect.
2D navigator-triggered prospective acquisition correction
(2D-PACE) [7] technique and cardiac triggering have been used to alleviate the aortic
motion. Acquisition was triggered every respiratory cycle. Therefore,
the real TR was equal to one respiratory cycle (3-4s). Pulse sequence diagram
of SPACE sequence with combined 2D PACE and ECG trigger is shown in Fig. 1. The
other basic parameters of T2w SPACE sequence were as follows: TE = 106ms; echo
train length = 135; FOV = 300mm×400mm; matrix size = 240×320;
bandwidth = 820 Hz/pixel; 56 sagittal slices with 1.3mm slice thickness; spectrally
selective fat saturation was used, parallel imaging (GRAPPA) acceleration
factor = 2, average = 1.8; The total acquisition time was around 5.5min (5+75
respiratory cycle); 1.3mm isotropic resolution was achieved. Results
In compliance
with IRB requirements, one healthy adult volunteer was scanned on 1.5T MRI
system (MAGNETOM Amira, Siemens Healthcare, Erlangen, Germany). Four measurements
have been performed as the following: (1) Diastolic acquisition without black
blood preparation pulse. (2) Diastolic acquisition with DANTE preparation pulse.
(3) Diastolic acquisition with diffusion-prepared pulse. (4) Systolic
acquisition without black blood preparation pulse, fast flow during systole can
enhance SPACE intrinsic black blood effect. A comparison of sample
images results from SPACE with different black blood techniques is
shown in Fig. 2.
During diastole, the blood flow is slow and complex flow exists in
aortic arch. Therefore the intrinsic black blood effect is not enough. Flow
artifacts can be observed in aortic arch in Fig. 2(a). Fig. 2(b) shows the
addition of DANTE removed flow artifacts partially, while the wall signal is
also decreased significantly. Fig. 2(c) shows the addition of diffusion-prepared
pulse reduced the flow artifacts substantially and the wall signal is mostly
remained. Fig. 2(d) shows the data acquisition during systole can remove the
flow artifacts completely. However, the wall signal is also influenced by the
pulsation of the aorta.
Diffusion-Prepared SPACE
with diastolic acquisition was found to be most promising for T2W whole aortic
wall imaging. Images in sagittal, coronal and
transversal plane are displayed in Fig. 3.Conclusion
This study optimizes a
high resolution, 3D MRI for T2w Aortic wall imaging in a clinically acceptable
scan time. The diffusion preparation module substantially reduces flow
artifacts that are otherwise noted on SPACE in health volunteer. Diffusion-Prepared
SPACE shows promise as a non-invasive tool for vasculitis diagnosis.Acknowledgements
References
1.
Gonzalez-Gay MA, Garcia-Porrua C: Epidemiology of the
vasculitides. Rheum Dis Clin North Am 2001, 27:729–749.
2.
Ioannis K, Debiao Li. Diffusion-Prepared Segmented Steady-State
Free Precession: Application to 3D Black-Blood Cardiovascular Magnetic
Resonance of the Thoracic Aorta and Carotid Artery Walls. Journal of
Cardiovascular Magnetic Resonance (2007) 9, 33–42.
3.
Mihai G, Chung YC, Merchant A, Simonetti OP, Rajagopalan S.
T1-weighted-SPACE dark blood whole body magnetic resonance angiography
(DB-WBMRA): initial experience. J Magn Reson Imaging. 2010; 31:502–509.
4.
Mihai G, Varghese J, Lu B, Zhu H, Simonetti OP, Rajagopalan S.
Reproducibility of thoracic and abdominal aortic wall measurements with
three-dimensional, variable flip angle (SPACE) MRI. J Magn Reson Imaging. 2013
5.
Wong SK, Mobolaji-Iawal M, Arama L, Cambe J, Biso S, Alie N, Fayad
ZA, Mani V. Atherosclerosis imaging using 3D black blood TSE SPACE vs 2D TSE.
World J Radiol. 2014; 6:192–202.
6. Chengcheng
Z, Henrik H, Farshid F. Isotropic 3D Black Blood MRI of Abdominal Aortic
Aneurysm Wall and Intraluminal Thrombus. Magn Reson Imaging. 2016 January ; 34(1): 18–25.
7.
Asbach P et al. Magnetic resonance cholangiopancreatography using
a free-breathing T2-weighted turbo spin-echo sequence with navigator triggered
prospective acquisition correction. Magn Reson Imaging 2005; 23: 939 –945.