Xenios Milidonis1, Russell Franks1, Kajol Verma1, Torben Schneider1,2, Javier Sánchez-González3, Sven Plein1, and Amedeo Chiribiri1
1King's College London, London, United Kingdom, 2Philips Healthcare, Guilford, United Kingdom, 3Philips Healthcare, Madrid, Spain
Synopsis
The arterial input function (AIF) describes the
contrast agent input to the myocardium and is required for blood flow
quantification. However, the impact of the AIF sampling location on
quantification by cardiovascular magnetic resonance and the diagnosis of
coronary artery disease (CAD) has yet to be determined. In this study,
perfusion imaging was performed in patients with suspected CAD and blood flow
was quantified for seven different locations. It was found that the sampling location
has a significant impact on blood flow measurements, while the ascending aorta led
to the most accurate prediction of inducible perfusion
abnormalities.
Introduction
Myocardial
blood flow (MBF) quantification by cardiovascular magnetic resonance (CMR)
first-pass perfusion has a high diagnostic accuracy in the identification of
coronary artery disease (CAD).1-3 Quantification requires accurate sampling
of the arterial input function (AIF) describing the contrast agent input to the
myocardium.4 The current clinical reference location is the basal
left ventricular blood pool, however, there is no evidence to support its use
over alternative locations. The aim of this study was to compare adenosine
stress MBF between seven different AIF sampling locations and determine which
location has the highest diagnostic accuracy for the classification of inducible
perfusion abnormalities by visual assessment.Methods
Twenty-five
patients with suspected or known CAD underwent adenosine stress perfusion
imaging on a 3T Philips Achieva system equipped with a 32-channel cardiac
phased-array coil (Philips Healthcare, Best, The Netherlands). All patients
gave written consent and the study was approved by the regional ethics
committee (15/NS/0030). An ECG-triggered saturation recovery spoiled gradient
echo dual-sequence technique was used,5 after injection of a
contrast bolus at 0.075 mmol/kg (Gadovist, Bayer AG, Leverkusen, Germany). Three
high-resolution slices were acquired at the basal, mid and apical short axis
level of the left ventricle (TR 2.2 ms, TE 1.0 ms, saturation delay 100 ms,
flip angle 15°, resolution 2.6x2.6 mm2, slice thickness 10 mm). A
low-resolution slice (sequence parameters as above except saturation delay 23.5
ms and resolution 2.6x5.3 mm2) was acquired in 3-chamber view to
allow sampling of the AIF at seven different locations by an experienced
cardiologist: right ventricle (RV), left atrium (LA), basal left ventricle
(bLV), mid left ventricle (mLV), apical left ventricle (aLV), ascending aorta
(aAo) and descending aorta (dAo). Signal intensity was converted to gadolinium
concentration,6 and then pixel-wise MBF was estimated using Fermi
function-constrained deconvolution.7 Mean stress MBF over all three
myocardial slices was compared between AIF sampling locations using analysis of
variance with post hoc pairwise comparisons with Bonferroni correction. Dynamic
perfusion images were visually examined by two independent level-3 experts to
classify myocardial segments into ischemic or non-ischemic, based on a
16-segment model according to the American Heart Association.8 For
each location, the diagnostic performance of stress MBF against visual
assessment was evaluated using receiver operating characteristic (ROC) curves.
Optimal sensitivity and specificity values were estimated using the Youden
index and the area under the ROC curves (AUC) was compared using the DeLong
method. Statistical significance was set at p
< 0.05.Results
The
sampled AIF curves had measured peak gadolinium concentrations ranging from
3.23 ± 1.25 mmol/L for the mLV to 6.18 ± 2.77 mmol/L for the RV. Stress MBF ranged
from 1.94 ± 0.61 mL/g/min for the RV to 2.94 ± 1.36 mL/g/min for the aLV and was
significantly different across AIF sampling locations (p = 0.002; Figure 1a). Pairwise comparisons revealed that mean stress
MBF for the clinical reference bLV (2.64 ± 0.56 mL/g/min) differed
significantly with the RV (p =
0.001). ROC analysis indicated that segmental stress MBF based on the aAo led
to a better prediction of inducible perfusion abnormalities than all other
locations (AUC 0.90 [95% confidence interval (CI) 0.87-0.93], sensitivity 80%,
specificity 85%, p < 0.001; AUC
compared against all other locations p
< 0.01; Figure 1b). Segmental stress MBF for the bLV had a high sensitivity and
a moderate specificity (AUC 0.81 [95% CI 0.77-0.85], sensitivity 91%,
specificity 63%, p < 0.001), while
the worst performing stress MBF was for the mLV (AUC 0.71 [95% CI 0.66-0.75],
sensitivity 79%, specificity 52%, p
< 0.001; AUC compared against all other locations p < 0.01). Analysis was based on 400 myocardial segments across
all 25 patients, of which 100 (25%) were visually classified as ischemic. Figure
2 shows representative AIF curves for all seven sampling locations for a normal
subject and Figure 3 demonstrates the impact on corresponding pixel-wise stress
perfusion maps for the clinical reference location bLV and the aAo.Discussion
This
is the first study to reveal that the AIF sampling location significantly
affects absolute stress MBF estimates by CMR perfusion. Lower peak gadolinium
concentrations in the AIF typically lead to higher MBF. Global alterations in
MBF can render existing thresholds of ischemia invalid and therefore affect the
clinical interpretation of perfusion maps and decision-making in the clinic. Indeed,
ROC curve analysis suggests that different AIF sampling locations lead to
different estimates of diagnostic accuracy. Segmental stress MBF for the aAo is
diagnostically superior to estimates for the current clinical reference bLV and
all other locations, most likely due to its close proximity to the coronary
arteries. Use of this location for AIF sampling and MBF quantification is
supported by the indicator-dilution theory,4 and is made possible by
contemporary dual-sequence techniques for CMR perfusion that allow decoupling
of the AIF and myocardial sampling locations.Conclusion
This
study demonstrates that sampling the AIF in the aAo for myocardial perfusion
quantification by CMR offers superior diagnostic accuracy than the bLV.
However, future studies are required for its validation against reference techniques
for CAD detection, such as invasive fractional flow reserve or positron emission
tomography perfusion.Acknowledgements
XM,
RF and AC were funded by the European Metrology Programme for Innovation and
Research (EMPIR) project 15HLT05 PerfusImaging, which is co-funded by the
European Union's Horizon 2020 research and innovation programme and the EMPIR
Participating States. XM, RF and SP were funded by the
British Heart Foundation [TG/18/2/33768, PG/18/71/34009, CH/16/2/32089].
Further support was received by the Department of Health (DoH) through the
National Institute for Health Research (NIHR) comprehensive Biomedical Research
Centre and Cardiovascular MedTech Co-operative awarded to Guy’s and St Thomas’
NHS Foundation Trust in partnership with King’s College London, and the
Wellcome/EPSRC Centre for Medical Engineering [WT 203148/Z/16/Z]. TS and JSG
are employees of Koninklijke Philips N.V. The views expressed are those of the
authors and not necessarily those of the DoH, the NIHR, the NHS, the Wellcome
Trust, the EPSRC or Koninklijke Philips N.V.References
-
Villa ADM, Corsinovi L, Ntalas I et al. Importance of
operator training and rest perfusion on the diagnostic accuracy of stress
perfusion cardiovascular magnetic resonance. J Cardiovasc Magn Reson.
2018;20(1):74.
- Mordini FE, Haddad T, Hsu LY et al. Diagnostic
accuracy of stress perfusion CMR in comparison with quantitative coronary
angiography: fully quantitative, semiquantitative, and qualitative assessment.
JACC Cardiovasc Imaging. 2014;7(1):14-22.
- Patel AR, Antkowiak PF, Nandalur KR et al. Assessment
of advanced coronary artery disease: advantages of quantitative cardiac
magnetic resonance perfusion analysis. J Am Coll Cardiol. 2010;56(7):561-569.
- Zierler KL. Theoretical Basis of Indicator-Dilution
Methods for Measuring Flow and Volume. Circulation Research.
1962;10(3):393-407.
- Sanchez-Gonzalez J, Fernandez-Jimenez R, Nothnagel ND
et al. Optimization of dual-saturation single bolus acquisition for
quantitative cardiac perfusion and myocardial blood flow maps. J Cardiovasc
Magn Reson. 2015;17:21.
- Henderson E, Sykes J, Drost D et al. Simultaneous MRI
measurement of blood flow, blood volume, and capillary permeability in mammary
tumors using two different contrast agents. J Magn Reson Imaging.
2000;12(6):991-1003.
- Jerosch-Herold M, Wilke N, Stillman AE. Magnetic
resonance quantification of the myocardial perfusion reserve with a Fermi
function model for constrained deconvolution. Med Phys. 1998;25(1):73-84.
- Cerqueira MD, Weissman NJ, Dilsizian V et al.
Standardized myocardial segmentation and nomenclature for tomographic imaging
of the heart. A statement for healthcare professionals from the Cardiac Imaging
Committee of the Council on Clinical Cardiology of the American Heart
Association. Circulation. 2002;105(4):539-542.