Giulio Ferrazzi1, Sarah McElroy1, Radhouene Neji1,2, Karl Kunze1,2, Muhummad Sohaib Nazir1, Peter Speier3, Daniel Stäb4, Christoph Forman3, Reza Razavi1, Amedeo Chiribiri1, and Sébastien Roujol1
1School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom, 2MR Research Collaborations, Siemens Healthcare Limited, Frimley, United Kingdom, 3Cardiovascular MR predevelopment, Siemens Healthcare GmbH, Erlangen, Germany, 4MR Research Collaborations, Siemens Healthcare Limited, Melbourne, Australia
Synopsis
The clinical need of covering multiple
slices in myocardial perfusion constrains the saturation delay time to be
short. However, this may be suboptimal in terms of myocardial/defect tissue contrast and blood/myocardium
signal ratios. Moreover, it is not possible to acquire the same slice twice, and all slices are at different cardiac phases.
In this study, we
develop a perfusion sequence which provides dual phase and dual contrast data
using simultaneous multi-slice at two different saturation delay times.
Simulations and in-vivo data acquired in patients demonstrated a 150% increase
of myocardial/defect contrast, and decreased blood/myocardium signal ratio by
60 to 80%.
Introduction
Conventional myocardial perfusion
protocols employ short saturation delay times (TS; 100-120ms) for the
acquisition of multiple slices covering base, mid and apex (Figure1a). However, the use of a longer TS may be beneficial to improve detection of
perfusion defects by providing increased myocardial/defect (MD) contrast, and a
reduced dark rim artefact due to lower blood/myocardium (BM) signal ratio. Simultaneous multi-slice1 (SMS) or multiband (MB)
imaging with CAIPIRINHA encoding2 is an acceleration technique which enables
simultaneous imaging of multiple slices. This technique has been successfully
extended to bSSFP imaging3 using the GC-LOLA correction and a multiband factor
of 24. It has also been successfully employed in cardiac perfusion to double
spatial coverage5. In this study, we sought to develop a dual
contrast/dual phase myocardial perfusion sequence by extending SMS-bSSFP GC-LOLA
to multiband 3.Methods
Sequence description
The proposed prototype sequence consists
of two saturation recovery blocks acquired in each heartbeat with different
saturation times (block1: long TS (LTS)=300ms; block2: short TS (STS)=130ms)
(Figure1b). Three slices (base, mid, apex) are imaged using SMS-bSSFP with
GC-LOLA correction and a multiband factor of 3 (see next section). T-GRAPPA
acceleration (R=7)6 and phase oversampling (300%) are employed for an
effective in-plane acceleration of 2.3. Images were reconstructed with an inline
prototype non-linear iterative reconstruction algorithm with spatio/temporal L1
regularisation7. Note that in our framework slice separation is achieved
along the phase encoding direction where the (shifted) slices are
reconstructed on the oversampled field of view8.
SMS-bSSFP with MB3 GC LOLA
Three triple-band pulses with CAIPIRINHA
RF phase increments of -120°, 0° and 120° for slices 1, 2 and 3 were generated as the complex
summation of a native single band (SB) pulse. This achieves shifts in image
space of -FOV/3, 0 and FOV/3 for slices 1, 2 and 3, which results in lower
g-factor amplification at reconstruction2. However, because each band is
subject to an independent phase cycling scheme, the frequency response of the
bSSFP signal is also shifted. For slices 1, 2 and 3, these shifts
equal to -1/3, 0 and 1/3 of the native bSSFP passband interval. The GC-LOLA
framework addresses this undesirable effect by i) applying an additional slice unbalancing gradient within each TR
interval to align the frequency response of each band and ii) adding an additional GC-LOLA phase cycling term to center the
frequency response of the (now aligned) bands onto the water peak4.
Simulation
Numerical simulations were employed to
study the relationship between TS, MD contrast and BM ratio (Figure2). The
following sequence parameters were employed: flip angle α=45°, TR=2.56ms, start-up pulses=6,
number of bSSFP readout pulses=70. For MD contrast (Figure2a), myocardial T1
(T1m) and T2 (T2m) times were set to 250ms and 44ms, respectively. A range of simulated
myocardial defect T1/T2 times (T1d/T2d) were evaluated from 400/46ms to 1200/50ms.
For the BM signal ratios (Figure2b), both peak blood (T1m/T2m=1200ms/50ms) and
peak myocardium (T1m/T2m=250ms/44ms) conditions were simulated. Blood T1/T2
(T1b/T2b) simulated range was 28.5/26.5ms to 168.8/108.7ms.
In-vivo evaluation
The sequence was evaluated in 5 patients
referred for contrast CMR at 1.5T (MAGNETOM Aera, Siemens Healthcare, Erlangen,
Germany) using an 18-element body coil and a 32-channel spine coil. All data was acquired using the following parameters: FOV=360x360mm2, slice thickness=10mm,
resolution=2.3x2.3mm2, TR=2.56ms, TE=1.09ms, flip angle α=45°, readout bandwidth=1008Hz/Px, start-up pulses=6,
number of bSSFP readout pulses=70,
readout duration=179ms, total acquisition time within a heartbeat=608ms. The
slice gap was adjusted for each patient to cover base, mid and apical regions. Each
patient underwent late gadolinium enhancement (LGE) imaging (one patient LGE
positive). A contrast dose of 0.075 (4 cases)/0.150 (1 case) mmol/kg was
injected. Patients performed an exhale breath-hold during first pass perfusion.
Data analysis
LTS and STS data were compared as
follows: blood pool and left ventricular myocardium were segmented at baseline,
peak blood, and peak myocardial enhancement. Contrast between peak and baseline
myocardium (used as a surrogate for defect) is reported. BM ratios at peak
blood and peak myocardium were calculated.Results
Simulations showed that MD contrast is
maximized for a TS range of 300-500ms (Figure2a). Figure2b suggests that the
BM ratio decreases with increasing TS. Across all patients (example in Figure3),
LTS images led to higher peak vs baseline myocardium contrast (158±21%, p<0.01)
as well as decreased BM ratio at peak blood (62±13%) and peak myocardium (79±12%)
(Figure4a-c). In the LGE-positive patient (Figure5), myocardial/scar contrast
increased by 158% at peak myocardial enhancement.Discussion
In this study, a dual contrast/phase
myocardial perfusion sequence with extended spatial coverage and increased
myocardial/defect contrast plus decreased blood/myocardium signal ratio was implemented.
In the present framework,
all slices are acquired twice at two different cardiac phases, with the
potential of triggering the LTS acquisitions at systole which could ultimately
facilitate the analysis of apical slices. Further
studies in a larger patient population with coronary artery diseases are warranted
to explore the benefits of the present technique. The robustness of the
sequence to respiratory motion should also be explored. Conclusion
The proposed perfusion sequence enables dual phase imaging with two different saturation delay times, allowing a 150% increase of myocardial/defect contrast and decreased blood/myocardium signal ratio by 60 to 80% for the LTS images when compared to conventional STS images.Acknowledgements
This work was supported by the EPSRC grant (EP/R010935/1) the
Wellcome EPSRC Centre for Medical Engineering at Kings College London
(WT 203148/Z/16/Z) and by the National Institute for Health Research
(NIHR) Biomedical Research Centre based at Guy’s and St Thomas’ NHS
Foundation Trust and King’s College London. The views expressed are
those of the authors and not necessarily those of the NHS, the NIHR or
the Department of Health.References
[1] Larkman DJ, Hajnal JV, Herlihy AH, Coutts GA, Young IR, Ehnholm
G. Use of multicoil arrays for separation of signal from multiple slices
simultaneously excited. J Magn Reson Imaging. 2001;13:313 317.
[2]
Breuer FA, Blaimer M, Heidemann RM, Mueller MF, Griswold MA, Jakob PM.
Controlled aliasing in parallel imaging results in higher acceleration
(CAIPIRINHA) for multi slice imaging. Magn Reson Med. 2005;53:684 691.
[3].
Stäb D, Ritter CO, Breuer FA, Weng AM, Hahn D, Köstler H. CAIPIRINHA
accelerated SSFP imaging. Magn Reson Med.2011;65:157‐164.10.
[4]
Stäb D, Speier P. Gradient-controlled local Larmor adjustment (GC-LOLA)
for simultaneous multislice bSSFP imaging with improved banding
behavior. Magn Reson Med. 2019 Jan;81(1):129 139.
[5] Nazir SB,
Neji R, Speier P, Reid FDA, Staeb D, Schmidt M, Forman C, Razavi R,
Plein S, Ismail TF, Chiribiri A, Roujol S. Simultaneous Multi Slice
(SMS) SSFP first-pass myocardial perfusion MRI with iterative
reconstruction at 1.5T. J Cardiovasc Magn Reson. 2018;20(84).
[6]
Breuer FA, Kellman P, Griswold MA, Jakob PM. Dynamic autocalibrated
parallel imaging using temporal GRAPPA (TGRAPPA). Magn Reson Med.
2005;53:981 985.
[7] Liu J, Rapin J, Chang T, Lefebvre A, Zenge
M, Mueller E, Nadar MS. Dynamic cardiac MRI reconstruction with weighted
redundant Haar wavelets. ISMRM. 2012; #4249.
[8] Stäb D, Speier
P, Reiter T, Klink T, Neubauer H, Bley TA, Wech T, Weng AM, Kstler H.
Restating MS-CAIPIRINHA as an In-plane Acceleration Problem: An
Efficient Method for Integrating High Coverage Cardiac Perfusion MRI
into Clinical Workflow. ISMRM. 2015; #2686.
[9] Weigel M.
Extended phase graphs: Dephasing, RF pulses, and echoes - Pure and
simple. J. Magn. Reson. Imaging 2015;41(2):266 295.