Gaspar Delso1, Paz Garre2, Francisco Alarcón2, Daniel Lorenzatti2, Julián Vega2, Teresa M. Caralt2, Adelina Doltra2, José T. Ortiz-Pérez2, Rosario J. Perea2, Susanna Prat2, Lluís Mont2, Marta Sitges2, and Martin Janich3
1ASL MR, GE Healthcare, Barcelona, Spain, 2Hospital Clínic de Barcelona, Barcelona, Spain, 3GE Healthcare, Munich, Germany
Synopsis
We present the results of a comparative study of the impact
of clinically available MR coil combination methods on the diagnostic value of
3D Delayed Myocardial Enhancement data, with a focus on arrhythmogenic tissue
analysis.
INTRODUCTION:
The use of Delayed Myocardial Enhancement sequences has
become a well-established technique for the determination of tissue viability
in ischemic cardiomyopathies1. Imaging is based on the use
of T1-prepared MR pulse sequences to determine the distribution of a
Gadolinium-based contrast agent in the myocardium. In order to attain the
optimal contrast between viable and pathologic tissue, the acquisition is
performed after allowing the contrast agent to redistribute for approximately
15 minutes.
Due to the diagnostic relevance of subtle uptake patterns, achieving
a good intensity contrast in the myocardium is critical for the success of
these exams. This is limited by the accuracy of the inversion time parameter,
which determines the suppression of healthy myocardium signal, and the overall signal-to-noise
ratio of the image.
Advanced coil combination methods incorporating calibration
information are known to provide improved noise performance, in comparison to
conventional sum-of-squares methods. On the other hand, these methods are susceptible to anatomical misalignment with the calibration series.
In this study, we examine the impact of two
different coil combination approaches on the diagnostic value of Delayed
Myocardial Enhancement images acquired with a three-dimensional pulse sequence
prototype. Particular attention is given to the outcome of automated clinical
metrics of the arrhythmogenic properties of tissue, used to drive therapeutic
decision-making.METHODS:
This study was performed by retrospectively reconstructing 3D-DME data from 15 subjects (12M/3F, age 56±14, weight 78±13kg)
referred for a clinically-indicated contrast-enhanced cardiac MRI examination at
the Clinical Hospital of Barcelona.
Acquisitions were performed on a General Electric Architect
3T scanner using a prototype pulse sequence with the following parameters: 3D SPGR readout; inversion-recovery T1 preparation with TI 283±29ms determined using a CINE scout scan; 40cm field-of-view
R/L; matrix 320x320; ST 2.0-2.4mm; bandwidth 62.5kHz; TE 2.2ms; spatial and chemical saturation; ARC
acceleration; respiratory navigator; ECG triggering with trigger time 616±138ms.
Cartesian 3D image reconstruction was performed offline
using GE’s Orchestra libraries on the raw data of the Delayed Enhancement scan
and those of a coil-matched calibration series, acquired during the same exam. Two
coil channel combination approaches were tested: Standard sum-of-squares; and a
calibration-based method akin to that used for ASSET
acceleration.
The resulting datasets were examined by two board-certified
cardiologists. Image quality was assessed in terms of signal and noise levels in
the background, blood pool and myocardium. They were also compared in terms of
artifacts and morphology of the ventricles and atria, as well as pathological
uptake when present.
Further analysis was performed using the ADAS arrhythmic
substrate identification software (Galgo Medical, SL). The total myocardial
area, border zone and lesion core percentages obtained with both
reconstructions, as well as the overall lesion distribution, were compared in
all cases2.RESULTS:
All cases were successfully exported and reconstructed with
both coil combination methods. Qualitatively, the most obvious difference
between the methods was a reduction of wrap-around artifacts (found in the arms
region with our acquisition settings), sometimes replaced by a slight decrease
of signal intensity in that region. Narrower intensity windows revealed
improved noise performance in the calibration-based method.
This was confirmed by the results of region of interest
analysis: Background average decreased by a factor 0.3±0.3 with
the calibration-based method, myocardium signal decreased by a 0.8±0.2 and
blood signal remained within 1.0±0.0. Wrap-around artifacts decreased in all cases and motion artifacts in 64%. The reconstructions were
found to be morphologically equivalent in all the cases for the left ventricle,
right ventricle, right atrium and left atrium, respectively. Pathological
uptake was identified in 2 cases, and both of them were qualified as
diagnostically equivalent with both reconstruction methods.
The quantitative results provided by ADAS
segmentation were found to differ by less than 5% between the reconstructions,
despite being segmented independently: Total area 0.5±2.2cm2;
Border zone 0.4±0.5cm2, 0.3%±0.4%; Core zone 0.0±0.5cm2, -0.1%±0.9%.
The arrhythmic substrate pattern was qualitatively equivalent in all cases.DISCUSSION:
The results obtained in this evaluation study suggest that,
while both coil combination techniques are suitable for diagnostic 3D-DME imaging, using a calibration-based method offers reduced
noise and artifacts without any noticeable drawbacks. A suitable calibration
scan is certainly required, but can be readily acquired in a few seconds, when
not already available from previous sequences (e.g. those using image-space
acceleration or calibration-based homogeneity correction).
No incidences were found of diagnostically relevant
artifacts caused by anatomical mismatch between the calibration and DME series. While the limited number of cases included in this
preliminary study does not allow to draw general conclusions, the absence of noticeable
artifacts is a relevant finding, considering that only pre-existing calibration
series were used, typically acquired before contrast injection.
Concerning the impact on the dedicated post-processing for
arrhythmogenic tissue characterization, the differences between the
reconstructions are likely to be dominated by the reproducibility of the
segmentation3. These results indicate that
advanced coil combination methods can be implemented without changing the analysis and therapy guidance protocols currently implemented by the
electrophysiology team.CONCLUSION:
The results of this comparative study indicate that the
well-known benefits of calibration-based coil combination extend to
three-dimensional DME imaging. No instances of
artifacts caused by calibration misalignment were encountered. The use of
different coil combination approaches didn’t affect the outcome of arrhythmogenic
pathway analysis. Acknowledgements
No acknowledgement found.References
1. Vogel-Claussen,
J. et al. Delayed Enhancement MR Imaging: Utility in Myocardial
Assessment1. RadioGraphics (2006) doi:10.1148/rg.263055047.
2. Benito, E. M. et al. Left atrial fibrosis
quantification by late gadolinium-enhanced magnetic resonance: a new method to
standardize the thresholds for reproducibility. Europace 19, 1272–1279
(2017).
3. Margulescu, A. D. et al. Reproducibility and accuracy
of late gadolinium enhancement cardiac magnetic resonance measurements for the
detection of left atrial fibrosis in patients undergoing atrial fibrillation
ablation procedures. Europace 21, 724–731 (2019).