Giorgia Milotta1, Camila Munoz1, Karl Kunze1, Radhouene Neji1, Stefano Figliozzi1, PierGiorgio Masci1, Claudia Prieto1, and Rene Botnar1
1Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom
Synopsis
Grey-blood phase-sensitive inversion-recovery
(PSIR) late gadolinium enhancement (LGE) imaging has shown promising and robust
results for the assessment of myocardium viability. Conventionally 2D grey-blood
LGE images are acquired under several breath-holds in different image
orientations to depict scar extension. However, these approaches achieve
limited spatial resolution and coverage and can be challenging in un-collaborative patients. In this work, we have
proposed a free-breathing 3D whole-heart LGE sequence with water/fat Dixon
encoding and blood nulling which provides grey-blood PSIR images with isotropic
resolution for scar visualization and complementary 3D fat images for
pericardial and myocardial adipose tissue detection in ~6min.
Introduction
Late gadolinium enhancement (LGE) cardiovascular
MR plays an important role in the detection and assessment of myocardial
viability. Particularly, grey-blood phase-sensitive inversion-recovery (PSIR) LGE
imaging has shown promising and robust results due to increased
contrast-to-noise ratio (CNR) between scar, normal myocardium and blood,
overcoming most of the limitations of bright-blood LGE such as endocardial scar
detection1,2. Conventionally, 2D grey-blood LGE images are acquired over multiple breath
holds in different image orientations to visualise and quantify scar extension.
However, this approach is limited by image misalignment, anisotropic resolution
and the need for multiple breath-holds. Furthermore, the presence of myocardial
fat infiltration or epicardial fat may render the distinction between fat and
LGE challenging both in ischemic and non-ischemic cardiomyopathy3. Here we propose a novel free-breathing 3D whole-heart LGE sequence with
water/fat Dixon encoding and magnitude blood-nulling which provides a grey-blood
PSIR volume, for scar visualization and complementary 3D fat images for fat
detection. The proposed approach was evaluated in a T1 phantom and 6 patients
with suspected cardiovascular disease and compared against standard 2D grey-blood
LGE with blood-nulling. Methods
The framework of
the proposed sequence is shown in Fig.1. Two interleaved 3D volumes are acquired
with ECG-triggered two-point bipolar Dixon, spoiled gradient echo readout and 3x
undersampled variable-density Cartesian spiral-like trajectory4,5. The two
datasets are acquired with 1) Inversion Recovery (IR) preparation and 2) no
preparation. Low-resolution 2D image navigators (iNAVs) are acquired prior to imaging
to correct translational respiratory motion, enabling 100% scan efficiency6. Translational
motion correction is performed for each echo, and each undersampled in- and
out-of-phase volume is reconstructed with iterative-SENSE7. A water-fat
separation algorithm is used to generate fat and water images for both datasets8 and a PSIR
reconstruction is performed to obtain the grey-blood LGE volume.
Acquisition: A T1 phantom composed
of 7 vials (including T1s of post-contrast myocardium, blood and scar) filled with
different gadolinium concentrations and six patients (4 male, 63±7yo) were
scanned on a 1.5T scanner (Siemens Magnetom Aera). Acquisition parameters
included: FA=25deg and 5deg for IR-prepared and non-prepared dataset, isotropic
resolution of 2mm3, FOV=320x320x104-128mm3, 14 low flip
angle echoes (FA=3deg) for iNAV acquisition, subject specific mid-diastolic
trigger-delay and acquisition window of 102-122ms, TR/TE1/TE2=6.41/2.38/4.76ms,
bandwidth=602Hz/pixel and total scan time=6±1min42sec. A 2D scout scan with imaging
parameters matching the 3D LGE sequence was performed prior to each patient acquisition
to select the optimal inversion time (TI) to null left ventricular blood
signal. The phantom experiment was carried out to evaluate the reliability of
the scout scan for TI selection and to compare CNR between myocardium, blood
and scar for both blood-nulling and conventional myocardium-nulling approaches.
The proposed 3D grey-blood LGE sequence was compared to standard 2D grey-blood
LGE in-vivo in terms of scar detection and image quality. Scar detection analysis
was performed using the 17-segment ventricular model, whereas image quality
score was performed using a 4-point Likert scale (1: non-diagnostic, 4:
excellent diagnostic quality). 2D LGE acquisition parameters included: bSSFP
readout with FA=45deg, resolution=1.4x1.4mm2, slice thickness=8mm, and
12 second breath-hold/slice. One slice in the two-chamber, three-chamber and
four-chamber orientations, and 13-15 slices in the short-axis orientation were
acquired in a total acquisition time of ~ 3min30sec excluding breath-hold
pauses and commands. 0.15 mml/Kg of gadobutrol was used for patient scans and
2D and 3D grey-blood LGE was performed ~10min and 20-25min after contrast
injection respectively.Results
Phantom: Good nulling of each phantom vial (Fig.2A)
was achieved by selecting the correct TI on the scout scan, particularly a
TI=240ms and TI=395ms nulled the vial corresponding to post-contrast blood and
myocardium respectively (Fig.2B). CNRmyoc-blood, CNRscar-myoc and
CNRscar-blood for blood-
and myocardium-nulling are shown in Fig.2C: a higher CNRscar-blood
is obtained with the grey-blood approach, consistent with previously reports in
the literature2.
Patients:
A motion-compensated 3D grey-blood LGE volume and co-registered fat volume are
shown for one patient P1 in Fig.3. The 3D nature of the acquisition allowed to
reformat the acquired datasets in different orientations. Good visualization of
scar was observed in all reformatted views with good scar-myocardium and
scar-blood contrast. The proposed 3D grey-blood LGE is qualitatively compared
to standard 2D grey-blood LGE in Fig.4 for three representative patients. Good
scar depiction and delineation is observed with the proposed 3D sequence, comparable
to standard 2D approach for patients P2 and P4, whereas no scar was detected in
patient P3. Furthermore, the contrast between myocardium, blood and scar
obtained with the proposed 3D grey-blood LGE is comparable to that obtained
with standard 2D LGE. The 2D grey-blood LGE acquisition generally resulted in a
better scar sharpness and delineation due to higher in-plane image resolution (1.4mm
vs 2mm), nevertheless excellent agreement in scar detection and comparable
image quality scores were obtained between the 2D and 3D LGE approaches (Fig.5).
Conclusion
The proposed 3D motion-compensated
grey-blood LGE sequence enables scar visualization and complementary 3D fat
images in ~6min. Good CNR between myocardium, scar and blood was obtained in
phantom acquisition. Good agreement in scar visualization and detection was
observed between the proposed approach and standard 2D grey-blood LGE imaging in
patients. Future work will investigate the increase in spatial resolution to
improve scar visualization and validation in a larger cohort of patients. Acknowledgements
This work was supported by EPSRC (EP/L015226/1, EP/P001009/1, EP/P007619
and EP/P032311/1) and the Wellcome EPSRC Centre for Medical Engineering
(NS/A000049/1).References
1. Kellman, P. et al. Dark blood
late enhancement imaging. J. Cardiovasc. Magn. Reson. 18, 77 (2016).
2. Holtackers, R. J., Chiribiri, A.,
Schneider, T., Higgins, D. M. & Botnar, R. M. Dark-blood late gadolinium
enhancement without additional magnetization preparation. J. Cardiovasc.
Magn. Reson. 19, 64 (2017).
3. Kellman, P. & Arai, A. E. Cardiac
imaging techniques for physicians: late enhancement. J. Magn. Reson. Imaging
36, 529–542 (2012).
4. Prieto, C. et al. Highly
efficient respiratory motion compensated free-breathing coronary MRA using
golden-step Cartesian acquisition. J. Magn. Reson. Imaging 41, (2015).
5. Bustin, A. et al. Five-minute
whole-heart coronary MRA with sub-millimeter isotropic resolution, 100%
respiratory scan efficiency, and 3D-PROST reconstruction. Magn. Reson. Med.
0, (2018).
6. Henningsson, M. et al.
Whole-heart coronary MR angiography with 2D self-navigated image
reconstruction. Magn. Reson. Med. 67, (2012).
7. Pruessmann, K. P., Weiger, M.,
Börnert, P. & Boesiger, P. Advances in sensitivity encoding with arbitrary
k-space trajectories. Magn. Reson. Med. 46, 638–651 (2001).
8. Liu, J., Peters, D. C. &
Drangova, M. Method of B0 mapping with magnitude-based correction for bipolar
two-point Dixon cardiac MRI. Magn. Reson. Med. 78, 1862–1869 (2016).